VNS Therapy Demipulse Model 103 Generator Cyberonics, Inc.… recallZ-0005-2012 · Cyberonics, Inc · initiated May 6, 2011
Terminated All recalls posted in October 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cyberonics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a patient's Model 103 Generator.
FDA root cause category: Software design (manufacturing process)
What the company is doing
The firm, Cyberonics Inc., initiated containment actions on March 17, 2011 to prevent additional affected devices from being shipped. The firm initiated contact with implanting facilities in the United States on May 6, 2011 and international facilities and distributors on May 10, 2011 to remove product from the field that has not yet been implanted. Additionally, a "Field Safety Alert" letter was sent to all affected medical professionals. The letter identifies the affected product, the issue, and the action(s) to be taken by the physicians. Consignees are instructed to complete and return an attached Effectiveness Check Form. If assistance is needed physicians may contact Clinical Technical Support at 866-882-8804 or via email at [email protected]
Recall details
- Recall number
- Z-0005-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 6, 2011
- Posted by the FDA
- Oct 4, 2011
- Recalling firm
- Cyberonics, Inc
- Quantity
- 92 units
- Product code
- LYJ
- FDA submissions
- P970003
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of AL, AZ, CA, CO, FL, GA, ID, IL, KS, OH, AR, MA, MI, MN, MO, NC, NH, NJ, NM, NY, OH, OR, PA, TN, TX, UT, WA, WI and WV. Products were also distributed to WASHINGTON DC and Puerto Rico and the countries of: Austria, Belgium, Finland, Germany, Ireland, Italy, Netherlands, Norway, Spain, Sweden, Switzerland, and United Kingdom.
Product description
VNS Therapy Demipulse Model 103 Generator Cyberonics, Inc. Houston, Texas, Model 103 Generators are provided in a sterile single use packages as part of the VNS Therapy System. Product labeling may be found at http://us.cyberonics.com/en/vns-therapy-for-epilepsy/healthcare-professionals/vns-therapy/manuals-page. Product Usage: The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in adults and adolescents over 12 years of age with partial onset seizures, which are refractory to antiepileptic medications and for the adjunctive long-term treatment of chronic or recurrent depression for patients 18 years of age or older who are experiencing a major depressive episode and have not had an adequate response to four or more adequate antidepressant treatments.
Code information (models, lots, serial numbers)
Serial #'s: 2181, 3427, 3902, 4082, 4083, 4116, 4129, 4132, 4158, 4177, 4185, 4204, 4235, 4304, 4351, 4495, 4591, 4610, 4621, 4716, 4728, 4737, 4823, 4834, 4855, 4904, 5105, 5165, 5220, 6039, 6046, 6121, 6124, 6234, 6343, 6513, 6552, 6625, 6674, 6680, 6682, 6700, 6715, 7662, 7738, 7875, 8008, 8163,8820, 8971, 9010, 9171, 9213, 9215, 9217, 9222, 9241, 9260, 9283, 9312, 9333, 9398, 9442, 9451, 9547, 9552, 12590, 12672, 18644, 21949, 22061, 22177, 22455, 22814, 23267, 23736, 25079, 25125, 25545, 25888, 26212, 26254, 26346, 26513, 26527, 26847, 26912, 26932, 27012, 27026, 27127, and 32232.
Other devices under the same FDA submission
23 devices cleared under P970003 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VNS Therapy® SenTiva™ Model 1000 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted For Epilepsy3 recalls188 deaths3,378 reports
- VNS Therapy® AspireSR® Model 106 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted For Epilepsy1 recall151 deaths2,934 reports
- VNS Therapy® PerenniaFLEX® Model 304 Lead 304-20Livanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)1 death1,614 reports
- VNS Therapy® Demipulse® Model 103 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)24 deaths371 reports
- VNS Therapy® AspireHC® Model 105 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)1 recall17 deaths307 reports
- VNS Therapy® Demipulse Duo® Model 104 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)20 deaths266 reports
- VNS Therapy® PerenniaDURA® Model 303 Lead 303-20Livanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)1 recall181 reports
- VNS Therapy® Symmetry™ Model 8103 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)8 deaths223 reports
- VNS Therapy® Model 302 Lead 302-20Livanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)161 reports
- VNS Therapy® SenTiva Duo™ Model 1000-D GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)2 recalls4 deaths28 reports
- VNS Therapy® PerenniaFLEX® Model 304 Lead 304-30Livanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)108 reports
- VNS Therapy® Pulse Model 102 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)6 deaths81 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 59139).
- Z-0006-2012 · TerminatedModel 104 VNS Therapy Demipulse Duo Generator Cyberonics, Inc. Houston, Texas, Model 104 Generators are provided in a sterile single use packages as part of the VNS Therapy System.…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page