Medical device recalls: October 2011
The FDA posted 141 medical device recall records in October 2011, from 71 recall events by 61 companies. 141 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Beckman Coulter Inc. · 3 events, 11 records
- GE Healthcare, LLC · 3 events, 4 records
- Siemens Healthcare Diagnostics, Inc. · 3 events, 3 records
- Philips Healthcare Inc. · 2 events, 4 records
- Terumo Cardiovascular Systems Corporation · 2 events, 3 records
- B Braun Medical, Inc · 2 events, 2 records
- Invivo Corporation · 2 events, 2 records
- The Anspach Effort, Inc. · 1 event, 13 records
FDA root cause
- Nonconforming Material/Component · 31
- Software design · 23
- Process control · 18
- Employee error · 15
- Incorrect or no expiration date · 12
- Device Design · 11
- Material/Component Contamination · 6
- Equipment maintenance · 5
Not yet classified (141)
The FDA has posted these records but has not published a class for them yet.
- Z-0139-2012 · Picis Inc. · posted Oct 31, 2011 · Terminated
Periop/Critical Care Manager EHR Software Application displayed one patient's demographic and health information in place of another patient that was in current view.
- Z-0140-2012 · Cardinal Health · posted Oct 31, 2011 · Terminated
Some of the components contained in the Presource kits are being recalled because Cardinal Health has determined that the component manufacturer(s) does not have all necessary FDA registrations or filings needed for this product.
- Z-0141-2012 · Cardinal Health · posted Oct 31, 2011 · Terminated
Some of the components contained in the Presource kits are being recalled because Cardinal Health has determined that the component manufacturer(s) does not have all necessary FDA registrations or filings needed for this product.
- Z-0142-2012 · Terumo Cardiovascular Systems Corporation · posted Oct 31, 2011 · Terminated
Terumo CVS identified a defective circuit board, which could potentially cause the CDI monitor to not function after a power loss. Once power is lost, the monitor will not power up again, even if AC or battery power are restored.
- Z-0133-2012 · GE Healthcare, LLC · posted Oct 28, 2011 · Terminated
A potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails.
- Z-0134-2012 · Baxter Healthcare Corp. · posted Oct 28, 2011 · Terminated
The Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test.
- Z-0135-2012 · Sierra Scientific Instruments Inc · posted Oct 28, 2011 · Terminated
The recall was initiated after Given Imaging received a test report during a Canadian licensing process where the AR ManoShield exhibited pinholes and did not pass the test, the Field Safety Corrective Action (FSCA) Committee was convened. As a result of…
- Z-0136-2012 · Stryker Communications Corp · posted Oct 28, 2011 · Terminated
Installation records for the SwitchPoint Infinity, Booms, Lights, and Flat Panel systems are incomplete or inadequate.
- Z-0137-2012 · Stryker Communications Corp · posted Oct 28, 2011 · Terminated
Installation records for the SwitchPoint Infinity, Booms, Lights, and Flat Panel systems are incomplete or inadequate.
- Z-0138-2012 · Stryker Communications Corp · posted Oct 28, 2011 · Terminated
Installation records for the SwitchPoint Infinity, Booms, Lights, and Flat Panel systems are incomplete or inadequate.
- Z-0073-2012 · Mizuho Orthopedic Systems Inc · posted Oct 27, 2011 · Terminated
Reports of injury related incidents while using the Mizuho OSI Modular Table System.
- Z-0113-2012 · Invivo Corporation · posted Oct 27, 2011 · Terminated
Philips Invivo has determined that it is possible for the system's alarm limits and other device settings to spontaneously change without user intervention. This problem is associated with intermittent and frequent loss of communication between the…
- Z-0116-2012 · Philips Healthcare Inc. · posted Oct 27, 2011 · Terminated
Front plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
- Z-0117-2012 · Philips Healthcare Inc. · posted Oct 27, 2011 · Terminated
Front plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
- Z-0118-2012 · Philips Healthcare Inc. · posted Oct 27, 2011 · Terminated
Front plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
- Z-0119-2012 · Lifeline First Aid LLC · posted Oct 27, 2011 · Terminated
The first aid kits are recalled because the kits contain Iodine Prep Pad single use wipes which were recalled by Triad.
- Z-0120-2012 · Lifeline First Aid LLC · posted Oct 27, 2011 · Terminated
The first aid kits are recalled because the kits contain Iodine Prep Pad single use wipes which were recalled by Triad.
- Z-0121-2012 · Lifeline First Aid LLC · posted Oct 27, 2011 · Terminated
The first aid kits are recalled because the kits contain Iodine Prep Pad single use wipes which were recalled by Triad.
- Z-0122-2012 · Lifeline First Aid LLC · posted Oct 27, 2011 · Terminated
The first aid kits are recalled because the kits contain Iodine Prep Pad single use wipes which were recalled by Triad.
- Z-0123-2012 · Lifeline First Aid LLC · posted Oct 27, 2011 · Terminated
The first aid kits are recalled because the kits contain Iodine Prep Pad single use wipes which were recalled by Triad.
- Z-0124-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0125-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0126-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0127-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0128-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0129-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0130-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0131-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0132-2012 · Beckman Coulter Inc. · posted Oct 27, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
- Z-0074-2012 · CareFusion 203, Inc. · posted Oct 26, 2011 · Terminated
CareFusion has identified potential risks associated with the EnVe ventilator manufactured between December 2010 and May 2011. There are 3 issues that can result in ventilation delivery to the patient being interrupted. CareFusion will be contacting…
- Z-0107-2012 · GE Healthcare, LLC · posted Oct 26, 2011 · Terminated
GE Healthcare has recently become aware that an exposed high voltage gradient wire internal to the MSK 1.5T Extreme system equipment may cause electrical shock, which may impact service personnel safety. This issue does not impact patient or operator safety.
- Z-0108-2012 · Salter Laboratories, Division of Regulatory Affairs · posted Oct 26, 2011 · Terminated
A small number of oxygen mask components were found to have 100% obstructed airways rendering them incapable of delivering oxygen to patients undergoing respiration therapy.
- Z-0109-2012 · Salter Laboratories, Division of Regulatory Affairs · posted Oct 26, 2011 · Terminated
A small number of oxygen mask components were found to have 100% obstructed airways rendering them incapable of delivering oxygen to patients undergoing respiration therapy.
- Z-0110-2012 · Salter Laboratories, Division of Regulatory Affairs · posted Oct 26, 2011 · Terminated
A small number of oxygen mask components were found to have 100% obstructed airways rendering them incapable of delivering oxygen to patients undergoing respiration therapy.
- Z-0111-2012 · Salter Laboratories, Division of Regulatory Affairs · posted Oct 26, 2011 · Terminated
A small number of oxygen mask components were found to have 100% obstructed airways rendering them incapable of delivering oxygen to patients undergoing respiration therapy.
- Mask adult, elastic strap style, Percent-O2 Lock air entrainment system, w/7' tube (Linde) - 50/cs ↗Z-0112-2012 · Salter Laboratories, Division of Regulatory Affairs · posted Oct 26, 2011 · Terminated
A small number of oxygen mask components were found to have 100% obstructed airways rendering them incapable of delivering oxygen to patients undergoing respiration therapy.
- Z-0114-2012 · GE Healthcare, LLC · posted Oct 26, 2011 · Terminated
GE Healthcare has become aware of inaccurate measurement on GE magnified images provided by Senographe DS and Essential Mammography systems when reviewed on IHE compliant workstations. This issue may impact patient safety.
- Z-0115-2012 · GE Healthcare, LLC · posted Oct 26, 2011 · Terminated
GE Healthcare has become aware of inaccurate measurement on GE magnified images provided by Senographe DS and Essential Mammography systems when reviewed on IHE compliant workstations. This issue may impact patient safety.
- Z-0089-2012 · CORPAK MedSystems · posted Oct 25, 2011 · Terminated
Corpak is removing the Corflo 300 Enteral Pump from the market due to the possibility that with fluid ingress or other damage, the keypad may not function, and the possibility of motor mount failure.
- Z-0106-2012 · Invivo Corporation · posted Oct 25, 2011 · Terminated
Philips Invivo Corporation of Orlando, FL is recalling their Expression MRI Patient Monitoring System. During transport of the system's cart with a docked Display Controller Unit (DCU), a front caster (wheel) may become loose and fall off causing the cart…
- Z-0080-2012 · Victus, Inc. · posted Oct 24, 2011 · Terminated
Victus, Inc. Miami, FL is recalling Vipat Brand Elastomeric Infusion Pump, Model H0020-100, Lot 100533. The product may have a higher flow rate than specified. The use of this product may lead to over-administration of drug solutions to patients.
- Z-0081-2012 · Progressive Medical Inc · posted Oct 24, 2011 · Terminated
The product may have a higher flow rate than specified resulting in over administration of drug solutions to the patients.
- Z-0105-2012 · Carl Zeiss Meditec, Inc. · posted Oct 24, 2011 · Terminated
An incorrect nut and bolt set was used in the axis three (3) spring assembly of the OPMI VARIO 700 surgical microscope distributed to three US locations.
- Z-0082-2012 · Smith & Nephew, Inc. Endoscopy Division · posted Oct 21, 2011 · Terminated
HD1200 camera heads may cause image issues
- Z-0083-2012 · Data Innovations, Inc. · posted Oct 21, 2011 · Terminated
The "previous result" retrieved from the Specimen Management Database may not be the most recent "previous result".
- Z-0084-2012 · Go Medical Industries Pty., Ltd. · posted Oct 21, 2011 · Terminated
O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling the Triad Povidone Iodine Swabsticks…
- Z-0085-2012 · Go Medical Industries Pty., Ltd. · posted Oct 21, 2011 · Terminated
O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling the Triad Povidone Iodine Swabsticks…
- Z-0086-2012 · Go Medical Industries Pty., Ltd. · posted Oct 21, 2011 · Terminated
O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling the Triad Povidone Iodine Swabsticks…
- Z-0087-2012 · Go Medical Industries Pty., Ltd. · posted Oct 21, 2011 · Terminated
O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling the Triad Povidone Iodine Swabsticks…
- Z-0088-2012 · Go Medical Industries Pty., Ltd. · posted Oct 21, 2011 · Terminated
O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling the Triad Povidone Iodine Swabsticks…
- Z-0090-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0091-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0092-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0093-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0094-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0095-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0096-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0097-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0098-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0099-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0100-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0101-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0102-2012 · The Anspach Effort, Inc. · posted Oct 21, 2011 · Terminated
The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Z-0103-2012 · Hamilton Co · posted Oct 21, 2011 · Terminated
Certain lots of syringes have caused an increase in error message "DIS541", which indicates "improper dispensation".
- Z-0104-2012 · Synthes USA (HQ), Inc. · posted Oct 21, 2011 · Terminated
LABELING CORRECTION - Following review of labeling and promotional materials associated with the Zero-P Implant system, it was found that the materials reference a study where PEEK-CR is utilized as a cervical interbody spacer. PEEK-CR is not currently…
- Z-0075-2012 · Stryker Howmedica Osteonics Corp. · posted Oct 20, 2011 · Terminated
Stryker Orthopaedics has become aware that a size 3 Accolade stem was in a box labeled as a size 5 Accolade stem.
- Z-0076-2012 · Biomet, Inc. · posted Oct 20, 2011 · Terminated
Biomet has initiated this action following an investigation which identified that the blueprint calls for engraved measurements on both sides on the Microplasty Taperloc 7.5MM Resection Guide. The above lots, as supplied to Biomet by a vendor, are engraved…
- Z-0077-2012 · Medtronic Perfusion Systems · posted Oct 20, 2011 · Terminated
The DFU for the Medtronic Pressure Display Box, model 66000, states that, for display readings of -100 to +300 mmHg there will be an accuracy of +/-5 mmHg, and for display readings of +301 to +500 mmHg there will be an accuracy of +/-10 mmHg. Through…
- Z-0078-2012 · Medtronic Perfusion Systems · posted Oct 20, 2011 · Terminated
Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the…
- Z-0079-2012 · Medtronic Perfusion Systems · posted Oct 20, 2011 · Terminated
Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the…
- Z-0068-2012 · Kapp Surgical Instrument Inc · posted Oct 19, 2011 · Terminated
A portion of the labeled expiration date is missing. The expiration date on approximately half of the pouch labels distributed are missing the last digit of the year of expiration. However, the shelf box containing the pouches is labeled correctly.
- Z-0069-2012 · Siemens Healthcare Diagnostics, Inc. · posted Oct 19, 2011 · Terminated
Firm has confirmed customer complaints for low recovery of QC and patient samples.
- Z-0070-2012 · Lemaitre Vascular, Inc. · posted Oct 19, 2011 · Terminated
Extended Expiration on the outer box label 2016-05, the correct expiration date is on the tray label: 2014-05
- Z-0071-2012 · Cardiocommand Inc. · posted Oct 19, 2011 · Terminated
CardioCommand, Inc. is recalling their TAPSCOPE 550 esophageal pacing/recording catheters because there maybe electrical discontinuity, which can result in the inability to reliably pace and/or record ECG.
- Z-0072-2012 · Cardiocommand Inc. · posted Oct 19, 2011 · Terminated
CardioCommand, Inc. is recalling their TAPSCOPE 550 esophageal pacing/recording catheters because there maybe electrical discontinuity, which can result in the inability to reliably pace and/or record ECG.
- Z-0051-2012 · B Braun Medical, Inc · posted Oct 18, 2011 · Terminated
B Braun Outlook ES Safety Infusion System "Run" Light Emitting Diodes (LEDs) pumps stop infusing and a backup alarm sounds, but the "Run" LEDs advance as if the pumps were infusing.
- Z-0057-2012 · Sonosite, Inc. · posted Oct 18, 2011 · Terminated
This is a software anomaly. In some circumstances, if the Systolic/Diastolic (S/D) ratio value is interpreted directly from the display, the clinician could potentially use an incorrect value of the S/D ratio in his/her assessment.
- Z-0058-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0059-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0060-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0061-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0062-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0063-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0064-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0065-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0066-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0067-2012 · Wright Medical Technology Inc · posted Oct 18, 2011 · Terminated
The product is labeled with an expiration date without adequate justification.
- Z-0050-2012 · Beckman Coulter Inc. · posted Oct 17, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed that certain Access Total hCG reagent packs show decreased stability from the date of manufacture, which may lead to increased imprecision.
- Z-0052-2012 · Impac Medical Systems Inc · posted Oct 17, 2011 · Terminated
Varian reported a problem related to certain collimator rotation values which are incorrect on TrueBeam compared to planned values in Mosaiq.
- Z-0053-2012 · Terumo Cardiovascular Systems Corporation · posted Oct 17, 2011 · Terminated
Terumo Cardiovascular Systems (Terumo CVS) recently discovered three instances of metal fragments inside the stainless steel tube of the Sarns" Adult Rigid Intracardiac Sucker.
- Z-0054-2012 · Terumo Cardiovascular Systems Corporation · posted Oct 17, 2011 · Terminated
Terumo Cardiovascular Systems (Terumo CVS) recently discovered three instances of metal fragments inside the stainless steel tube of the Sarns" Adult Rigid Intracardiac Sucker.
- Z-0044-2012 · CooperVision Inc. · posted Oct 14, 2011 · Terminated
This recall was initiated because of the unintended presence of a residue (silicone oil) on lenses from certain lots of Avaira Toric contact lenses. The company received complaints of temporary hazy or blurry vision and discomfort, most likely caused by…
- Z-0048-2012 · Abbott Molecular · posted Oct 14, 2011 · Terminated
The Abbott m2000rt System software has software bug which can affect customers who manually enter a Test Order onto the m2000rt and sorts a column on the calibration or control grids of the Test Order screen by double clicking the column header. This…
- Z-0049-2012 · ArthroCare Corporation · posted Oct 14, 2011 · Terminated
ArthroCare initiated the recall of one lot of the MiniMagnum Knotless Implant Fixation Device where the incorrect diver block was packed with the implant.
- Z-0045-2012 · Interventional Spine Inc · posted Oct 13, 2011 · Terminated
This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result…
- Z-0046-2012 · Interventional Spine Inc · posted Oct 13, 2011 · Terminated
This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result…
- Z-0047-2012 · Interventional Spine Inc · posted Oct 13, 2011 · Terminated
This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result…
- Z-0055-2012 · Angiodynamics, INC · posted Oct 13, 2011 · Terminated
The product may contain an incorrect size catheter and introducer.
- Z-0056-2012 · Angiodynamics, INC · posted Oct 13, 2011 · Terminated
The product may contain an incorrect size catheter and introducer.
- Z-0037-2012 · Boston Scientific CRM Corp · posted Oct 12, 2011 · Terminated
Boston Scientific has determined that under certain conditions, LATITUDE Paceart Integration (LPI) software may not properly transfer data resulting in the incorrect summary of programmed parameter values in the ICD Summary tab display and Current…
- Z-0038-2012 · Siemens Healthcare Diagnostics, Inc. · posted Oct 12, 2011 · Terminated
Internal testing has confirmed that recovery of levels 2 and 3 of LOCl Cardiac Troponin I Calibrator is lower than expected bottle value. Following calibration with LOCl Cardiac Troponin I Calibrator, lot 1AD045 an upward shift in patient and QC results…
- Z-0039-2012 · Richard-Allan Scientific Company · posted Oct 12, 2011 · Terminated
The firm has received two complaints of finger lacerations in operators of this device.
- Z-0040-2012 · Worldwide Medical Technologies, LLC · posted Oct 12, 2011 · Terminated
Length of the cannula hub is longer and may not permit a fit with the Applicator
- Z-0041-2012 · Edan Instruments, Inc. · posted Oct 12, 2011 · Terminated
The 3 leads of the cable, RL, LA and LL were labeled in wrong sequence. The left arm (LA) is mislabeled as right leg (RL). The left leg (LL) is mislabeled as left arm (LA). The right leg (RL) is mislabeled as left leg (LL). There have been no reports of…
- Z-0042-2012 · Exalenz Bioscience Ltd · posted Oct 12, 2011 · Terminated
The product is in violation of the FD&C Act in that it was illegally marketed without proper clearance (no 510(k) submitted).
- Z-0043-2012 · EMD Chemicals Inc. · posted Oct 12, 2011 · Terminated
Stains may yield lower then expected absorbance levels resulting in under stained hematology specimens.
- Z-0033-2012 · Isopure Corp · posted Oct 11, 2011 · Terminated
Software problem causes transfer line between mixing and storage tank to not be rinsed after disinfection, leaving residual chlorine in the line.
- Z-0034-2012 · X Spine Systems Inc · posted Oct 11, 2011 · Terminated
The firm learned that the Torque Driver Handles were not actuating at a proper torque. It was determined that the handles were speed determinant and could output a higher torque if the handles were turned at a very slow speed.
- Z-0035-2012 · Accuray Incorporated · posted Oct 11, 2011 · Terminated
A report and investigation determined that the set screws, which were part of the retrofit of a previous recall were incorrectly engaged which allowed rotation and subsequent descent of the assembly.
- Z-0036-2012 · Accuray Incorporated · posted Oct 11, 2011 · Terminated
A report and investigation determined that the set screws, which were part of the retrofit of a previous recall were incorrectly engaged at one facility which allowed rotation and subsequent descent of the assembly.
- Z-0017-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0018-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0019-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0020-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0021-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0022-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0023-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0024-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0025-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0026-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0027-2012 · Bio-Rad Laboratories, Inc. · posted Oct 7, 2011 · Terminated
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a…
- Z-0028-2012 · Beckman Coulter Inc. · posted Oct 7, 2011 · Terminated
The recall was initiated because Beckman Coulter has confirmed an issue with the Coulter PrePlus 2. The reagent level sensing is impaired and will give no warning when a reagent vial is low or out of reagent. Due to the size difference in the vertical…
- Z-0029-2012 · Olympus America Inc. · posted Oct 7, 2011 · Terminated
Firm has received complaints from users of the ability to wipe material from the needle's stylet and the presence of material in patient cytology samples obtained by the needle following endobronchial ultrasound fine needle aspiration of a lesion. Olympus…
- Z-0030-2012 · Tcoag Us, Incorporated · posted Oct 7, 2011 · Terminated
There is a performance issue with Minutex D-Dimer Kit which may cause false negative results for patient samples containing levels of greater than 250ng/mL D-Dimer.
- Z-0031-2012 · Siemens Healthcare Diagnostics, Inc. · posted Oct 7, 2011 · Terminated
Siemens Healthcare Diagnostics has confirmed a low frequency defect in the molding of the Cuvettes that may result in Cuvette Ring and Cuvette Loader jams. It has been determined that the issue is related to a molding defect with the Cuvette flange or…
- Z-0032-2012 · Philips Healthcare Inc. · posted Oct 7, 2011 · Terminated
Clinical measurements on true size printouts may be inaccurate
- Z-0015-2012 · Deroyal Industries, Inc. Lafollette · posted Oct 6, 2011 · Terminated
The firm manufactured kits containing a HCH device, manufactured by Vital Signs, that was subsequently recalled because the product may have an occlusion that could prevent proper flow of medical gases or oxygen to the patient.
- Z-0016-2012 · Deroyal Industries, Inc. Lafollette · posted Oct 6, 2011 · Terminated
The firm manufactured kits containing a HCH device, manufactured by Vital Signs, that was subsequently recalled because the product may have an occlusion that could prevent proper flow of medical gases or oxygen to the patient.
- Z-0012-2012 · Biomet Microfixation, Inc. · posted Oct 5, 2011 · Terminated
On 09/18/2011, Biomet Microfixation, Jacksonville, FL, recalled their SP-2411, Rubber Band, Lot 307850 because it was labeled and sold as "LATEX FREE" although it contains "LATEX". The firm's HHE identified this as a potential high health risk to the…
- Z-0013-2012 · Convatec Inc. · posted Oct 5, 2011 · Terminated
Post-operative kits 57 mm may contain 70 mm pouches.
- Z-0004-2012 · Smiths Medical ASD, Inc. · posted Oct 4, 2011 · Terminated
Reports of non-delivery/ significant under-delivery of fluids when the CADD Medication Cassette Reservoir is not correctly latched to the pump. EXPANDED: On 7/14/11, Smith Medical expanded this recall to include all consignees with affected pumps.
- Z-0005-2012 · Cyberonics, Inc · posted Oct 4, 2011 · Terminated
An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a patient's Model 103…
- Z-0006-2012 · Cyberonics, Inc · posted Oct 4, 2011 · Terminated
An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a patient's Model 103…
- Z-0008-2012 · Osteomed, Lp · posted Oct 4, 2011 · Terminated
Two of the Implant Driver Assembly tips were reported to break during surgery.
- Z-0009-2012 · Hospira Inc. · posted Oct 4, 2011 · Terminated
Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This condition could…
- Z-0010-2012 · Hospira Inc. · posted Oct 4, 2011 · Terminated
Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This condition could…
- Z-0011-2012 · Hospira Inc. · posted Oct 4, 2011 · Terminated
Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This condition could…
- Z-0001-2012 · Medtronic Navigation, Inc · posted Oct 3, 2011 · Terminated
Distal end of tactile probe may break off during use.
- Z-0002-2012 · Medtronic Navigation, Inc · posted Oct 3, 2011 · Terminated
Distal end of tactile probe may break off during use.
- Z-0003-2012 · Cochlear Americas Inc. · posted Oct 3, 2011 · Terminated
Cochlear implant may shut down and cease to function.
- Z-0007-2012 · B Braun Medical, Inc · posted Oct 3, 2011 · Terminated
The Outlook ES Dose Guard Care Area can inadvertently be exited when the pump enters the KVO (Keep Vein Open) state. The same condition can occur when a sequence of HOLD-HOLD key strokes are performed on the pump key panel.