Contour, Contour TS Blood Glucose Test Strips and Bonus… recallZ-3268-2011 · Bayer Healthcare, LLC · initiated Aug 12, 2011
Terminated All recalls posted in September 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bayer Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
All 10- or 25- count glucose test strip vials, either loose vials or found in patient starter kits of Contour, Contour TS and Contour" USB blood glucose meters are being recalled a packaging issue which, under certain conditions, can cause some test strips packaged in some small-count vials to operate outside of US FDA-cleared performance specifications and produce a low-bias reading during blood
FDA root cause category: Packaging
What the company is doing
Bayer sent an "URGENT DIABETES TEST STRIP RECALL" letter dated August 12, 2011, to its customers with the name of the test strips, sizes recalled, reason for the recall, potential injury, and instructions to discontinue use and return the test strips to Bayer. The recall letter also included a recall response form that was to be faxed back to Bayer at 1-800-876-2243. For questions call Customer Logistics at 1-800-348-2637.
Recall details
- Recall number
- Z-3268-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 12, 2011
- Posted by the FDA
- Sep 19, 2011
- Recalling firm
- Bayer Healthcare, LLC
- Quantity
- 3,617,287 vials/ 54,555,460 test strips
- Product code
- LFR
- FDA submissions
- K062058
- Distribution
- Nationwide Distribution and Puerto Rico
Product description
Contour, Contour TS Blood Glucose Test Strips and Bonus Packs Intended for self-testing by people with diabetes and healthcare professionals to monitor glucose concentrations in whole blood
Code information (models, lots, serial numbers)
9545C, 9677, 9678, 9679, 9680, 7151G, 7190, 7183, 9615, 7393, 9591, 9591A 9579, 1802, 6181, 7070A, 9725, 9507C, and 9578. Contour 125 Count Bonus Pack (includes 2-50 Count and 1-25 Count Vials) the following lot numbers:DW0AC3B32A, DW0AC3B32C, DW0BC3B31A, DW0BC3B31B, DW0BC3D05B, DW0CC3B31A, DW0CC3B31B, DW0GC3B31C, DW0GC3B31D, DW0GC3B31F, DW0HC3B32A, DW9MC3B32D, DW9MC3B32E, DW9MC3B33A, and DW9MC3B33B.
Other devices under the same FDA submission
9 devices cleared under K062058 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CONTOUR® 7080GAscensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose61 reports
- CONTOUR® 7151HAscensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose37 reports
- CONTOUR® 9545CAscensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose21 reports
- CONTOUR® 7097CAscensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose15 reports
- CONTOUR® 9556CAscensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose10 reports
- CONTOUR® 7189Ascensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose6 reports
- CONTOUR® 7098CAscensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose3 reports
- CONTOUR® 7109BAscensia Diabetes Care US Inc. · System, Test, Blood Glucose, Over The Counter2 reports
- CONTOUR® 7099CAscensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page