Medtronic Disposable Pressure Display Set, Models 61000… recallZ-0078-2012 · Medtronic Perfusion Systems · initiated Sep 9, 2011
Terminated All recalls posted in October 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the system resulting in a pressure reading that is not representative of the actual line pressure.
FDA root cause category: Device Design
What the company is doing
Medtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification. For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.
Recall details
- Recall number
- Z-0078-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 9, 2011
- Posted by the FDA
- Oct 20, 2011
- Recalling firm
- Medtronic Perfusion Systems
- Quantity
- 193,590 units
- Product code
- DXS
- FDA submissions
- K852232
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela, and Yemen.
Product description
Medtronic Disposable Pressure Display Set, Models 61000, 61006, 62000, 62006. Sterilized by Ethylene Oxide. Non-pyrogenic. Coiled PVC tubing with threaded transducer dome (acrylic) with silicone diaphragm on proximal end. Product utilizes a stopcock on proximal and/or distal end. Distal end of tubing features a male luer port. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604. Product Usage: This product is intended for use in monitoring catheter, cannula, or line pressures associated with cardiopulmonary bypass equipment and/or related products.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
Lot numbers: 2008070788, 2008070792, 2008070793, 2008070955, 2008080357, 2008080359, 2008080385, 2008080426, 2008080441, 2008080525, 2008080545, 2008081109, 2008081195, 2008090080, 2008090244, 2008090980, 2008091122, 2008091123, 2008091125, 2008100001, 2008100531, 2008100583, 2008100605, 2008100607, 2008101323, 2008101324, 2008101325, 2008101368, 2008101375, 2008101379, 2008101380, 2008110008, 2008110173, 2008110287, 2008110301, 2008110530, 2008110531, 2008110612, 2008110720, 2008110876, 2008110877, 2008110878, 2008111157, 2008111209, 2008111210, 2008120149, 2008120150, 2008120478, 2008120848, 2008120849, 2008120850, 2008120851, 2008121107, 2009010121, 2009010122, 2009010603, 2009010604, 2009010605, 2009010953, 2009010954, 2009010955, 2009011280, 2009011281, 2009011282, 2009011283, 2009011284, 2009020119, 2009020581, 2009020841, 2009020842, 2009020843, 2009030102, 2009030327, 2009030328, 2009030329, 2009030330, 2009030626, 2009030955, 2009031171, 2009031172, 2009040538, 2009040539, 2009040540, 2009040541, 2009040834, 2009041092, 2009041093, 2009041094, 2009050183, 2009050527, 2009050528, 2009050529, 2009050867, 2009050868, 2009050975, 2009050976, 2009060054, 2009060381, 2009060382, 2009060383, 2009060633, 2009060640, 2009060641, 2009060642, 2009060696, 2009061139, 2009070210, 2009070487, 2009070488, 2009070912, 2009071231, 2009080403, 2009080404, 2009080405, 2009080406, 2009080407, 2009081128, 2009081129, 2009081130, 2009081131, 2009081851, 2009090137, 2009090589, 2009090590, 2009090591, 2009091085, 2009100116, 2009100118, 2009100443, 2009100444, 2009100445, 2009100446, 2009100882, 2009100974, 2009100975, 2009100976, 2009100977, 2009100978, 2009101597, 2009101598, 2009102220, 2009102221, 2009102222, 2009110186, 2009110859, 2009110861, 2009111388, 2009112071, 2009120316, 2009120317, 2009120788, 2009120789, 2009120790, 2009120791, 2009121304, 2009121305, 2009121306, 2009121701, 2009121704, 2010010236, 2010010237, 2010010238, 2010011329, 2010011330, 2010011525, 2010011526, 2010011527, 2010021286, 2010022073, 2010022074, 2010031328, 2010031329, 2010031330, 2010031331, 2010031332, 2010031333, 2010032084, 2010032085, 2010032086, 2010032087, 2010032816, 2010032817, 2010032818, 2010032819, 2010032820, 2010033477, 2010033478, 2010033479, 2010040563, 2010040564, 2010040947, 2010040948, 2010040949, 2010041433, 2010050370, 2010050371, 2010050726, 2010050963, 2010051276, 2010051363, 2010060108, 2010060280, 2010060558, 2010060689, 2010061703, 2010061759, 2010062023, 2010062558, 2010062870, 2010062871, 2010070115, 2010071372, 2010071373, 2010071375, 2010071377, 2010071570, 2010071868, 2010080229, 2010080231, 2010080810, 2010080811, 2010080812, 2010081270, 2010081424, 2010081602, 2010081604, 2010081606, 2010082011, 2010082012, 2010082218, 2010082219, 2010082439, 2010090223, 2010090420, 2010090497, 2010090615, 2010090759, 2010090939, 2010090952, 2010091040, 2010091666, 2010091802, 2010092208, 2010092209, 2010092472, 2010092505, 2010100038, 2010100237, 2010100249, 2010100374, 2010100376, 2010100681, 2010101291, 2010101293, 2010101374, 2010101385, 2010101493, 2010101961, 2010101974, 2010110030, 2010110033, 2010110974, 2010120220, 2010120424, 2010120605, 2010120606, 2010121586, 2010121587, 2010122170, 2010122208, 2011010069, 2011010367, 2011010368, 2011011063, 2011011067, 2011011311, 2011011454, 2011011931, 2011020018, 2011020019, 2011020217, 2011020218, 2011020581, 2011020583, 2011020585, 2011020955, 2011020956, 2011021012, 2011021014, 2011021155, 2011021156, 2011021157, 2011021668, 2011021669, 2011021991, 2011021993, 2011022187, 2011022188, 2011030015, 2011030089, 2011030090, 2011030095, 2011030155, 2011030156, 2011030311, 2011030312, 2011031091, 2011031094, 2011031450, 2011031610, 2011031611, 2011031685, 2011032335, 2011032336, 2011040004, 2011040025, 2011040026, 2011040260, 2011040261, 2011040365, 2011040366, 2011040811, 2011040812, 2011041038, 2011041070, 2011041092, 2011041194, 2011041306, 2011041307, 2011041573, 2011041574, 2011041590, 2011041592, 2011050006, 2011050362, 2011050363, 2011050397, 2011050398, 2011050657, 2011050697, 2011051009, 2011051238, 2011051239, 2011051381, 2011051382, 2011051696, 2011051704, 2011060013, 2011060014, 2011060184, 2011060373, 2011060375, 2011060551, 2011060680, 2011060726, 2011060786, 2011061269, 2011061271, 2011062118, 2011062119, 2011062400, 2011062401, 2011062402, 2011070035, 2011070036, 2011070285, 2011070287, 2011070847, 2011070848, 200806C084, 200806C089, 200808C026, 200808C094, 200808C109, 200811C004, 200811C005, 200811C038, 200812C005, 200903C075, 200905C126, 200908C163, 200908C164, 200910C047, 200910C175, 200910C182, 200910C183, 200911C085, 200912C010, 201001C087, 201002C045, 201002C135, 201005C040, 201006C018, 201007C033, 201008C040, 201009C132, 201009C180, 201009C181, 201011C104, 201011C132, 201011C139, 201012C042, 201101C032, 201102C103, 201103C061, 201104C019, 201104C064, 201104C106, 201105C057.
Other devices under the same FDA submission
2 devices cleared under K852232 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DLP® 62000Medtronic, Inc. · GAUGE, PRESSURE, CORONARY, CARDIOPULMONARY BYPASS2 recalls0 reports
- DLP® 61000Medtronic, Inc. · GAUGE, PRESSURE, CORONARY, CARDIOPULMONARY BYPASS1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 59948).
- Z-0077-2012 · TerminatedMedtronic Pressure Display Box, Models 66000, 64000, 63000, 60000. It is capable of measuring pressure between -100 mmHg and +500 mmHg with an accuracy of +/-5 mmHg from -100 mmHg…
- Z-0079-2012 · TerminatedMedtronic Intersept Custom Tubing Pack, which are designed as specified by the user. The following models contain a Pressure Display Set. Sterilized by Ethylene Oxide. Non-pyro…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page