Medtronic Intersept Custom Tubing Pack, which are designed… recallZ-0079-2012 · Medtronic Perfusion Systems · initiated Sep 9, 2011

Terminated All recalls posted in October 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the system resulting in a pressure reading that is not representative of the actual line pressure.

FDA root cause category: Device Design

What the company is doing

Medtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification. For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.

Recall details

Recall number
Z-0079-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 9, 2011
Posted by the FDA
Oct 20, 2011
Recalling firm
Medtronic Perfusion Systems
Quantity
74,804
Product code
DXS
FDA submissions
K852232
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela, and Yemen.

Product description

Medtronic Intersept Custom Tubing Pack, which are designed as specified by the user. The following models contain a Pressure Display Set. Sterilized by Ethylene Oxide. Non-pyrogenic. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604. Product Usage: This product is intended for use in monitoring catheter, cannula, or line pressures associated with cardiopulmonary bypass equipment and/or related products.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Models 0K57R2, 0K82R15, 0M63R7, 0M68R6, 0N41R14, 0T50R5, 0V67R27, 0X52R3, 0Y35R, 0Y95R30, 0Z30R11, 1D41R16, 1F66R12, 1G59R1, 1J74R4, 1M90R8, 1P09R2, 2031R7, 2C88R8, 2C89R7, 2C90R7, 2C91R5, 2D61R2, 2D64R16, 2H08R3, 2K85R1, 2K87R1, 2K88R1, 2N90R9, 2T28R1, 3B74R2, 3T73R1, 3W29R, 4B42R5, 4C03R, 4D82R3, 4M55R, 4N04R, 4R43R, 5788R4, 5A45R3, 5B51R9, 5B94R, 5C33R2, 5C45R3, 5C92R9, 5D83R2, 5F68R1, 5G28R4, 5H65R, 5J00R2, 5P62R, 5W13R6, 5Z18R, 6G05R, 6J55R1, 6J88R4, 6J95R8, 6N12R, 6Q49R1, 6Q49R1, 6R08R, 6R09R, 6R44R2, 6S66R, 6S75R1, 6V25R, 6Y04R5, 6Y07R6, 6Y59R3, 7A69R4, 7A70R1, 7B56R1, CB0L44R16, CB0M02R5, CB0M79R8, CB1Q18R11, CB1S33R9, CB2D32R5, CB2W45R7, CB2W86R4, CB3K45R9, CB3K91R1, CB3M40R17, CB3Y46R9, CB3Y47R10, CB4800R18, CB4953R16, CB4958R16, CB4B68R8, CB4E78R7, CB4F45R15, CB4K64R6, CB4L83R4, CB4M60R2, CB4S43R9, CB4W10R10, CB4X84R, CB4Y71R14, CB4Z87R6, CB5A30R10, CB5F84R12, CB5J33R6, CB5L15R13, CB5L79R, CB5L90R16, CB5Q71R2, CB5Q74R1, CB5R04R1, CB5R46R5, CB5U81R, CB6A80R6, CB6G02R4, CB6H12R, CB6H45R4, CB6H76R, CB6H77R, CB6H78R, CB6J66R2, CB6J68R3, CB6J75R3, CB6L12R3, CB6L13R, CB6L14R, CB6L60R2, CB6M11R1, CB6Q15R, CB6Q16R1, CB6Q62R, CB6Q78R2, CB6S17R, CB6U12R, CB6U76R1, CB6X94R2, CB6X98R3, CB6Y02R6, CB6Y34R2, CB7C31R, DLP2X17R, HY0F32R17, HY2280R16, HY2U14R2, HY3U59R12, HY4A96R9, HY4V52R4, HY4V82R3, HY5D07R2, HY6B26R2, HY6G55R6, HY6G56R6, HY6H03R5, HY6H03R5, HY6J33R3, HY6J99R, HY6R36R1, HY6R36R1, HY6U70R2, HY6V34R2, HY6Z04R1, M331105H, M331105K, M336902L, M392810D, M392810E, M392811C, M392811D, M392811E, M440115G, M440115H, M440115J, M440115K, M440115L, M440129A, M440130A, M441002E, M443824F, M443825E, M492609A, M499301B, M530710A, M530710C, M530710D, M960208A, M960454B, P390806A, S330115A, SS6X67R1, SS6Z09R1, SSHY6E81R2, TL2H89R7, TL3Y45R1, TL3Z67R4, TL4E64R, TL4E66R10, TL4E66R10, TL5398R21, TL5C09R4, TL5M01R7, TL5R15R4, TL6A91R2, TL6C91R2, TL6D87R2, TL6H39R5, TL6W08R3, TL6X29R4, TL6X36R2, TL6X62R5, TL6Y92R, TL7A15R2.

Other devices under the same FDA submission

2 devices cleared under K852232 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 59948).