Medtronic Arctic Front Catheter, Sterile EO, Medtronic… recallZ-0149-2012 · Medtronic Inc. Cardiac Rhythm Disease Management · initiated Aug 29, 2011
Terminated All recalls posted in November 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc. Cardiac Rhythm Disease Management or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic CryoCath has identified specific lots of Arctic Front CryoAblation Catheters that were distributed with incorrect programming that may result in false expired catheter notifications. Impacted lots are 21571, 25375, 25376, 25377, and 25381 of the Arctic Front Catheter Model Number 2AF282. Medtronic CryoCath is communicating this information to the FDA. On affected catheters, the incor
FDA root cause category: Employee error
What the company is doing
Medtronic sent an "Important Device Recall" letter dated August 2011, hand delivered by Medtronic Field Representatives on August 29, 2011 to all affected customers. The letter described the product affected, the problem, and the actions needed to be taken. The customers were instructed to return all unused catheters from affected lots for credit and replacement. The letter states that Medtronic AF Solution representative will assist customers with the return of affected product. For further questions contact your local Medtronic AF Solutions representative.
Recall details
- Recall number
- Z-0149-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 29, 2011
- Posted by the FDA
- Nov 2, 2011
- Recalling firm
- Medtronic Inc. Cardiac Rhythm Disease Management
- Quantity
- 263
- Product code
- OAE
- FDA submissions
- P100010
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of: AK, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MN, MS, MO, NE, NY, NC, OH, OK, PA, SC, SD, TN, TX, VA, WA, and WI. and the countries of: Austria, United Kingdom.
Product description
Medtronic Arctic Front Catheter, Sterile EO, Medtronic CryoCath LP, Kirkland, Quebec, H9H 5 H3, Canada, Model 2AF282 and 2AF231 ( not distributed in US) Product Usage: The Medtronic CryoCath System consists of a Cryoablation console, cardiac Cryoablation catheters, connection components and accessories. The Arctic Front Cardiac Cryoablation Catheter is an over the wire Balloon catheter. The inside of the Balloon at the distal end of the Cryoablation Catheter reaches temperatures of -50C or colder when refrigerant is injected from the console to the inside of the balloon. Arctic Front Cardiac Cryoablation catheters are introduced into the vasculature by traditional minimally invasive techniques. The CryoConsole houses the electrical and mechanical components as well as proprietary software for controlling and recording a cryotherapy procedure with Arctic Front and/or the Freezor family of catheters. It stores and controls the delivery of the liquid refrigerant through the coaxial umbilical to the catheter, recovers the refrigerant vapor from the catheter under constant vacuum, and disposes of the refrigerant through the hospital scavenging system. Multiple features are built into the catheter and the CryoConsole system to ensure safety.
Code information (models, lots, serial numbers)
Model Number: 2AF282 Lot Numbers: 21571, 25375, 25376, 25377, and 25381
Other devices under the same FDA submission
22 devices cleared under P100010 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Arctic Front Advance™ 2AF284Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation9 deaths144 reports
- Arctic Front Advance Pro™ AFAPRO28Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation3 deaths104 reports
- Arctic Front Advance Pro™ AFAPRO28Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation2 deaths97 reports
- Arctic Front Advance Pro™ AFAPRO28Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation4 deaths97 reports
- Arctic Front Advance™ 2AF284Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 death75 reports
- Arctic Front Advance™ 2AF284Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation58 reports
- Arctic Front Advance Pro™ AFAPRO28Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation38 reports
- Arctic Front Advance™ 2AF284Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation37 reports
- Arctic Front Advance Pro™ AFAPRO28Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation32 reports
- Arctic Front Advance™ 2AF284Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation25 reports
- Arctic Front Advance™ 2AF284Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 death23 reports
- Arctic Front Advance™ 2AF284Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page