Model CAT-850B Catheterization Table, a component of the… recallZ-0331-2012 · Toshiba American Medical Systems Inc · initiated Oct 3, 2011
Terminated All recalls posted in December 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Toshiba American Medical Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The recall was initiated because Toshiba has confirmed that during patient transfer from a gurney to the catherization table, the tabletop may rotate unexpectedly when sufficient force is applied to the tabletop. The result of the tabletop rotating unexpectedly is the patient being tranferred may fall from the tabletop.
FDA root cause category: Device Design
What the company is doing
Toshiba America Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 3, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers were instructed to follow the user instructions by positioning the tabletop at the foot-end position, and ensuring sufficient assistance is available, when transferring a patient from the gurney to the CAT-850B catheterization table. Customers were also instructed to follow the instructions for transferring a patient to the CAT-850B that are included in the operation manual. A Customer Reply Form was included with the letter for customers to complete and return via fax to (877) 349-3054. The customer reply form can also be sent via e-mail to [email protected]. Questions regarding the recall letter can contact the Director of Regulatory Affairs at (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.
Recall details
- Recall number
- Z-0331-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 3, 2011
- Posted by the FDA
- Dec 1, 2011
- Recalling firm
- Toshiba American Medical Systems Inc
- Quantity
- 300 units
- Product code
- MQB
- FDA submissions
- K081582, K081621, K081624, K082830
- Distribution
- Nationwide Distribution.
Product description
Model CAT-850B Catheterization Table, a component of the Toshiba INFX-8000 series X-Ray Interventional System. These systems are single or dual plane systems that employ x-rays to obtain fluoroscopy or radiography images of the body. The system is comprised of a computer system, support arms (that contain the tube and solid state detector), x-ray generator and a patient table. This device is a digital x-ray system configured for use as a diagnostic x-ray angiography system. This x-ray angiography system is used for diagnostic and interventional procedures for cardiac blood vessels, cerebral blood vessels, abdominal blood vessels and lower extremities.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Last Four Digits of Lot #2408, 2064, 2851, 2076, 2081, 2804, 2109, 2100, 2519, 2521, 2008, 3063, 2537, 2350, 2382, 2834, 3030, 3043, 3250, 3194, 2105, 2938, 2063, 2148, 2187, 2313, 2324, 2330, 2331, 2424, 2284, 2692, 2067, 2416, 2802, 3294, 2846, 2852, 2901, 3011, 3075, 3167, 2590, 2591, 2462, 3044, 2305, 3155, 3066, 3102, 3290, 3188, 2176, 2781, 3293, 2117, 3308, 3309, 2862, 2321, 2225, 2235, 2238, 3041, 7003, 7004, 2502, 2599, 3317, 2246, 2272, 2217, 3330, 3048, 2954, 3341, 2391, 2835, 2557, 2303, 2294, 2440, 2348, 2597, 3003, 2369, 2500, 2070, 2613, 2299, 2301, 2216, 3146, 3166, 3110, 2262, 2040, 2675, 3064, 2097, 2145, 3203, 2870, 2306, 2750, 2776, 2691, 2410, 7005, 2930, 2494, 2052, 2244, 2245, 2457, 3345, 2683, 2152, 2995, 2970, 2816, 2763, 2593, 2708, 2674, 2744, 2547, 2966, 2967, 2730, 2836, 2843, 2844, 2845, 2643, 2751, 2261, 2373, 2697, 2559, 2405, 2565, 3189, 2121, 3134, 2821, 3195, 2157, 2158, 2206, 3193, 3285, 3350, 2841, 2401, 2503, 2863, 2534, 2140, 2530, 3355, 3262, 3307, 2581, 2736, 2982, 2418, 2280, 2281, 3149, 2353, 2456, 2346, 2349, 2091, 2343, 3213, 2815, 2857, 2909, 2973, 3049, 3047, 2544, 2688, 3060, 2315, 2308, 2676, 2961, 3176, 2760, 3187, 2267, 2504, 2908, 2738, 2007, 3156, 3373, 3380, 2354, 2524, 2190, 2250, 3173, 3197, 3349, 2470, 2178, 2174, 2785, 3022, 3371, 3108, 2394, 3352, 2383, 2824, 2574, 2659, 2789, 2782, 2788, 2498, 2614, 2975, 2398, 3150, 2620, 2752, 3319, 2445, 2501, 2837, 2737, 2801, 2194, 2213, 2513, 2558, 2957, 2984, 2204, 3123, 2395, 2478, 2595, 2921, 3325, 2399, 3074, 3087, 3181, 2900, 2538, 3314, 2082, 3236, 2814, 2095, 3200, 3055, 3056, 2022, 3178, 2310, 3093, 2198, 2096, 2268, 3305, 2952, 2628, 2994, 2427, 2655, 2989, 2374, 2447, 3378, 3198, 3346, 2362, 2455, 2721, 3261, 2423, 2439, 2569, 2682, 3177, 3233, 2104, 2871, 2876, 2073, 2316, 2128, 3336, 3427.
Other devices under the same FDA submission
2 devices cleared under K081582, K081621, K081624, K082830 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- INTERVENTIONAL ANGIOGRAPHY SYSTEM INFX-8000FCanon Inc. · Solid state x-ray imager (flat panel/digital imager)1 recall0 reports
- INTERVENTIONAL ANGIOGRAPHY SYSTEM INFX-8000VCanon Inc. · Interventional fluoroscopic x-ray system1 death3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page