Model CAT-850B Catheterization Table, a component of the… recallZ-0331-2012 · Toshiba American Medical Systems Inc · initiated Oct 3, 2011

Terminated All recalls posted in December 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Toshiba American Medical Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The recall was initiated because Toshiba has confirmed that during patient transfer from a gurney to the catherization table, the tabletop may rotate unexpectedly when sufficient force is applied to the tabletop. The result of the tabletop rotating unexpectedly is the patient being tranferred may fall from the tabletop.

FDA root cause category: Device Design

What the company is doing

Toshiba America Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 3, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers were instructed to follow the user instructions by positioning the tabletop at the foot-end position, and ensuring sufficient assistance is available, when transferring a patient from the gurney to the CAT-850B catheterization table. Customers were also instructed to follow the instructions for transferring a patient to the CAT-850B that are included in the operation manual. A Customer Reply Form was included with the letter for customers to complete and return via fax to (877) 349-3054. The customer reply form can also be sent via e-mail to [email protected]. Questions regarding the recall letter can contact the Director of Regulatory Affairs at (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.

Recall details

Recall number
Z-0331-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 3, 2011
Posted by the FDA
Dec 1, 2011
Recalling firm
Toshiba American Medical Systems Inc
Quantity
300 units
Product code
MQB
FDA submissions
K081582, K081621, K081624, K082830
Distribution
Nationwide Distribution.

Product description

Model CAT-850B Catheterization Table, a component of the Toshiba INFX-8000 series X-Ray Interventional System. These systems are single or dual plane systems that employ x-rays to obtain fluoroscopy or radiography images of the body. The system is comprised of a computer system, support arms (that contain the tube and solid state detector), x-ray generator and a patient table. This device is a digital x-ray system configured for use as a diagnostic x-ray angiography system. This x-ray angiography system is used for diagnostic and interventional procedures for cardiac blood vessels, cerebral blood vessels, abdominal blood vessels and lower extremities.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Last Four Digits of Lot #2408, 2064, 2851, 2076, 2081, 2804, 2109, 2100, 2519, 2521, 2008, 3063, 2537, 2350, 2382, 2834, 3030, 3043, 3250, 3194, 2105, 2938, 2063, 2148, 2187, 2313, 2324, 2330, 2331, 2424, 2284, 2692, 2067, 2416, 2802, 3294, 2846, 2852, 2901, 3011, 3075, 3167, 2590, 2591, 2462, 3044, 2305, 3155, 3066, 3102, 3290, 3188, 2176, 2781, 3293, 2117, 3308, 3309, 2862, 2321, 2225, 2235, 2238, 3041, 7003, 7004, 2502, 2599, 3317, 2246, 2272, 2217, 3330, 3048, 2954, 3341, 2391, 2835, 2557, 2303, 2294, 2440, 2348, 2597, 3003, 2369, 2500, 2070, 2613, 2299, 2301, 2216, 3146, 3166, 3110, 2262, 2040, 2675, 3064, 2097, 2145, 3203, 2870, 2306, 2750, 2776, 2691, 2410, 7005, 2930, 2494, 2052, 2244, 2245, 2457, 3345, 2683, 2152, 2995, 2970, 2816, 2763, 2593, 2708, 2674, 2744, 2547, 2966, 2967, 2730, 2836, 2843, 2844, 2845, 2643, 2751, 2261, 2373, 2697, 2559, 2405, 2565, 3189, 2121, 3134, 2821, 3195, 2157, 2158, 2206, 3193, 3285, 3350, 2841, 2401, 2503, 2863, 2534, 2140, 2530, 3355, 3262, 3307, 2581, 2736, 2982, 2418, 2280, 2281, 3149, 2353, 2456, 2346, 2349, 2091, 2343, 3213, 2815, 2857, 2909, 2973, 3049, 3047, 2544, 2688, 3060, 2315, 2308, 2676, 2961, 3176, 2760, 3187, 2267, 2504, 2908, 2738, 2007, 3156, 3373, 3380, 2354, 2524, 2190, 2250, 3173, 3197, 3349, 2470, 2178, 2174, 2785, 3022, 3371, 3108, 2394, 3352, 2383, 2824, 2574, 2659, 2789, 2782, 2788, 2498, 2614, 2975, 2398, 3150, 2620, 2752, 3319, 2445, 2501, 2837, 2737, 2801, 2194, 2213, 2513, 2558, 2957, 2984, 2204, 3123, 2395, 2478, 2595, 2921, 3325, 2399, 3074, 3087, 3181, 2900, 2538, 3314, 2082, 3236, 2814, 2095, 3200, 3055, 3056, 2022, 3178, 2310, 3093, 2198, 2096, 2268, 3305, 2952, 2628, 2994, 2427, 2655, 2989, 2374, 2447, 3378, 3198, 3346, 2362, 2455, 2721, 3261, 2423, 2439, 2569, 2682, 3177, 3233, 2104, 2871, 2876, 2073, 2316, 2128, 3336, 3427.

Other devices under the same FDA submission

2 devices cleared under K081582, K081621, K081624, K082830 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.