INTERVENTIONAL ANGIOGRAPHY SYSTEM INFX-8000VRecalls & reported problems

Canon Inc. · INTERVENTIONAL ANGIOGRAPHY SYSTEM · Class 2

All models of INTERVENTIONAL ANGIOGRAPHY SYSTEM (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04987670100222), received between Jun 13, 2024 and May 14, 2026 — classified as 1 death, 2 malfunctions.

The most frequently reported specific device problem is “Inadequate User Interface” (1 report).

12 other recalls cover devices cleared under the same FDA submission (K081582, K112054, K120073, K133535, K143225, K152696, K162614, K170909, K172863, K181670, K203551, K251602, K253584); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Inadequate User Interface1
  • Material Fragmentation1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Perforation of Vessels1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2026Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2026 DURING THE TRANSFER FROM THE HOSPITAL BED / GURNEY TO THE PROCEDURE TABLE A PATIENT BECAME UNSTABLE WHILE BEING MOVED TO THE PROCEDURE TABLE AND SLIPPED. HOSPITAL STAFF CAUGHT THE PATIENT BEFORE HE HIT THE FLOOR AND THE PATIENT RECEIVED NO INJUIRES AS A RESULT OF THIS INCIDENT.

    Patient problems: Insufficient Information

    MAUDE report 25166893 ↗
  • MalfunctionApr 2, 2026Material Fragmentation

    DURING AUTOMATIC POSITIONING, THE C-ARM AND OMEGA CONTACTED. AFTER THAT, A USER ATTEMPTED TO KEEP THE ARMS SEPARATED BUT OPERATED AS THE ARMS WERE FURTHER STRONGLY CONTACTED BY ADDITIONAL USER OPERATIONS. THIS CAUSED DAMAGE TO THE X-RAY BEAM LIMITING DEVICE UNIT AND THE OCCURRENCE OF AN ERROR OF THIS UNIT. AFTER THE STUDY WAS FINISHED, A CUSTOMER ENGINEER FOUND THAT THE ROTATION MECHANISM CAME OFF. THERE WAS NO HARM TO THE PATIENT OR USER, NOR ANY DAMAGE TO SURROUNDING EQUIPMENT. THIS EVENT DID NOT AFFECT THE STUDY BECAUSE THIS HAPPENED AT THE END OF THE STUDY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24772377 ↗
  • DeathJun 13, 2024Inadequate User Interface; Inadequate User Interface

    IN A HOSPITAL, LOCATED IN SPAIN, ON (B)(6) 2024 IT WAS ALLEGED BY THE USER THAT "IMAGE DEFINITION IS LOST WHEN MAGNIFYING, THIS PREVENTS THE CORRECT VISUALIZATION OF THE CATHETER NAVIGATION USING FLUORO AND ALSO PREVENTS THE CORRECT RELEASE OF THE STENTS OR FLOW DIVERTER DEVICES. ALSO, THE GRAIN OF THE IMAGE AND THE HALO EFFECT AROUND THE COILS ARE PRONOUNCED". THE PATIENT FROM (B)(6) ALLEGEDLY SUFFERED A PERFORATION DURING THE PROCEDURE THAT RESULTED IN SEVERE BLEEDING. THE MANUFACTURED RECEIVED INFORMED ON MAY 22ND THAT THE PATIENT HAD PASSED AWAY.

    Patient problems: Perforation of Vessels

    MAUDE report 19534300 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081582, K112054, K120073, K133535, K143225, K152696, K162614, K170909, K172863, K181670, K203551, K251602, K253584. One submission often covers many models, so check the product and lot details in each record.

  • Z-3014-2026 · Aug 4, 2026 · Open, Classified
    Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The sys
  • Z-1666-2022 · Jan 31, 2022 · Open, Classified
    INFX-8000V. For radiographic and fluoroscopic studies and intervention.
  • Z-0492-2022 · Dec 20, 2021 · Open, Classified
    Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.
  • Z-1174-2021 · Feb 5, 2021 · Open, Classified
    System INFX-8000V
  • Z-1269-2021 · Dec 8, 2020 · Open, Classified
    Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000V Catheterization Table CAT-880B Canon CATHETERIZATION TABLE, MODEL CAT-880B INPUT 220V - Product Usage: indicated for u
  • Z-2325-2020 · Mar 3, 2020 · Open, Classified
    Infinix-8000V with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angio
  • Z-0844-2018 · Dec 29, 2017 · Terminated
    INFX-8000V Fluoroscopic X-Ray Systems; INFINIX CF-I/BP PLUS CARDIAC SYSTEM
  • Z-0752-2016 · Aug 14, 2015 · Terminated
    INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems
  • Z-1394-2015 · Mar 11, 2015 · Terminated
    INFX-8000V Interventional Angiography System. CAT-880B catheterization table. Cardiac and Vascular intervention with fluoroscopic and fluorographic images. For the USA and Canad
  • Z-1714-2014 · May 12, 2014 · Terminated
    Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.
  • Z-0994-2013 · Nov 8, 2012 · Terminated
    Toshiba Infinix-i X-Ray System. A radiographic and fluoroscopic x-ray system for use in diagnostic and angiographic procedures.
  • Z-0331-2012 · Oct 3, 2011 · Terminated
    Model CAT-850B Catheterization Table, a component of the Toshiba INFX-8000 series X-Ray Interventional System. These systems are single or dual plane systems that employ x-rays to

Other Interventional fluoroscopic x-ray system devices