Medical device recalls: December 2011
The FDA posted 44 medical device recall records in December 2011, from 27 recall events by 19 companies. 44 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips Medical Systems (Cleveland) Inc · 3 events, 6 records
- Varian Medical Systems, Inc. Oncology Systems · 3 events, 4 records
- SCC Soft Computer · 3 events, 3 records
- Varian Medical Systems, Inc. · 2 events, 4 records
- Elekta, Inc. · 2 events, 2 records
- Atom Medical Corporation · 1 event, 4 records
- Zimmer Inc. · 1 event, 4 records
- Integra LifeSciences Corp. · 1 event, 3 records
FDA root cause
- Device Design · 13
- Software design · 11
- Nonconforming Material/Component · 5
- Process control · 4
- Error in labeling · 4
- Vendor change control · 3
- No Marketing Application · 2
- Packaging · 1
Not yet classified (44)
The FDA has posted these records but has not published a class for them yet.
- Z-0346-2012 · Nanotherapeutics, Inc · posted Dec 2, 2011 · Terminated
On 03/10/2011 Nanotherapeutics Inc, Alachua, FL initiated a recall of their Origen DBM with Bioactive Glass Catalog # 22-2002 (2cc), 22-2005 (5cc) and 22-2010 (10cc). The same product is also distributed as NanoFUSE DBM Catalog # NAN109-02 (2cc) NAN109-05…
- Z-0347-2012 · Nanotherapeutics, Inc · posted Dec 2, 2011 · Terminated
On 03/10/2011 Nanotherapeutics Inc, Alachua, FL initiated a recall of their Origen DBM with Bioactive Glass Catalog # 22-2002 (2cc), 22-2005 (5cc) and 22-2010 (10cc). The same product is also distributed as NanoFUSE DBM Catalog # NAN109-02 (2cc) NAN109-05…
- Z-0348-2012 · Synthes USA (HQ), Inc. · posted Dec 2, 2011 · Terminated
The potential exists for the Large Oscillating Saw Attachment to disengage from the Small Battery Drive handpiece. If the attachment disengages from the handpiece, there is a potential for injury to the patient and/or user.
- Z-0349-2012 · Encore Medical, Lp · posted Dec 2, 2011 · Terminated
The strike plate may dislodge from the broach stem handle during impaction of the device.
- Z-0351-2012 · Bard Peripheral Vascular Inc · posted Dec 2, 2011 · Terminated
This recall was voluntarily initiated by the firm due to the product 's damaged packaging trays; cracked EnCor (Trademark) packaging trays, plastic PETG tray the probe is packaged in was cracked.
- Z-0352-2012 · Maquet Inc. · posted Dec 2, 2011 · Terminated
Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical error message 24 (TE 24).
- Z-0353-2012 · Maquet Inc. · posted Dec 2, 2011 · Terminated
Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical error message 24 (TE 24).
- Z-0354-2012 · Maquet Inc. · posted Dec 2, 2011 · Terminated
Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical error message 24 (TE 24).
- Z-0355-2012 · SCC Soft Computer · posted Dec 2, 2011 · Terminated
On 06/13/2011, SCC Soft Computer initiated a correction on the following SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. A client reported when they verified the contents of the fax reports, they found one of the specimen diagnoses to be missing.…
- Z-0356-2012 · Integra LifeSciences Corp. · posted Dec 2, 2011 · Terminated
A loose screw may result in shifting of components and improper function of the turret or actuator assembly.
- Z-0357-2012 · Integra LifeSciences Corp. · posted Dec 2, 2011 · Terminated
A loose screw may result in shifting of components and improper function of the turret or actuator assembly.
- Z-0358-2012 · Integra LifeSciences Corp. · posted Dec 2, 2011 · Terminated
A loose screw may result in shifting of components and improper function of the turret or actuator assembly.
- Z-0359-2012 · Raysearch Laboratories AB · posted Dec 2, 2011 · Terminated
The firm has found an issue with DICOM RT Image representation of DRR images from RayStation 2.0 which needs to be clarified in the user manual. The symptoms involve the Pixel Spacing and Image Position attributes in the RT images being incorrect, thus…
- Z-0360-2012 · Philips Medical Systems (Cleveland) Inc · posted Dec 2, 2011 · Terminated
Software errors. A number of non-conformances were logged during in-house testing and external validation testing of Brilliance 64 and Ingenuity CT software version 4.0 Beta 2. The non-conformances are due to software errors identified in these device…
- Z-0361-2012 · Philips Medical Systems (Cleveland) Inc · posted Dec 2, 2011 · Terminated
Software errors. A number of non-conformances were logged during in-house testing and external validation testing of Brilliance 64 and Ingenuity CT software version 4.0 Beta 2. The non-conformances are due to software errors identified in these device…
- Z-0362-2012 · SCC Soft Computer · posted Dec 2, 2011 · Terminated
On 10/25/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. When SoftReports Designer opened a layout, the system did not load the layout with the previously saved value…
- Z-0363-2012 · SCC Soft Computer · posted Dec 2, 2011 · Terminated
On 07/20/2011 SCC Soft Computer, Clearwater, FL, initiated a correction on the following SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. A client reported that the HPV results on a patient report were incorrect. The HPV results were from the…
- Z-0364-2012 · Varian Medical Systems, Inc. Oncology Systems · posted Dec 2, 2011 · Terminated
An anomaly has been identified with the TrueBeam and TrueBeam STx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape.
- Z-0320-2012 · Elekta, Inc. · posted Dec 1, 2011 · Terminated
The software did not operate as expected.
- Z-0321-2012 · Zimmer Inc. · posted Dec 1, 2011 · Terminated
This recall was initiated because it may not be clear to the surgeon whether the device can be used based on the instructions for use.
- Z-0322-2012 · Zimmer Inc. · posted Dec 1, 2011 · Terminated
This recall was initiated because it may not be clear to the surgeon whether the device can be used based on the instructions for use.
- Z-0323-2012 · Zimmer Inc. · posted Dec 1, 2011 · Terminated
This recall was initiated because it may not be clear to the surgeon whether the device can be used based on the instructions for use.
- Z-0324-2012 · Zimmer Inc. · posted Dec 1, 2011 · Terminated
This recall was initiated because it may not be clear to the surgeon whether the device can be used based on the instructions for use.
- Z-0325-2012 · Varian Medical Systems, Inc. · posted Dec 1, 2011 · Terminated
VariSource iX console software may cause incorrect dwell times to be delivered to patient and may result in a mistreatment.
- Z-0326-2012 · Varian Medical Systems, Inc. · posted Dec 1, 2011 · Terminated
Plastic needle tip could possibly crack or separate.
- Z-0327-2012 · Varian Medical Systems, Inc. · posted Dec 1, 2011 · Terminated
Plastic needle tip could possibly crack or separate.
- Z-0328-2012 · Varian Medical Systems, Inc. · posted Dec 1, 2011 · Terminated
Plastic needle tip could possibly crack or separate.
- Z-0329-2012 · Varian Medical Systems, Inc. Oncology Systems · posted Dec 1, 2011 · Terminated
The Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the source cable in the event the bronchial catheter is dislodged from the connector.
- Z-0330-2012 · Varian Medical Systems, Inc. Oncology Systems · posted Dec 1, 2011 · Terminated
The Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the source cable in the event the bronchial catheter is dislodged from the connector.
- Z-0331-2012 · Toshiba American Medical Systems Inc · posted Dec 1, 2011 · Terminated
The recall was initiated because Toshiba has confirmed that during patient transfer from a gurney to the catherization table, the tabletop may rotate unexpectedly when sufficient force is applied to the tabletop. The result of the tabletop rotating…
- Z-0332-2012 · Varian Medical Systems, Inc. Oncology Systems · posted Dec 1, 2011 · Terminated
An anomaly has been identified with the Head Ring posts used by both the Optical Guidance Platform FrameArray module and the Floorstand where the head ring posts may be damaged due to excessive mechanical stress resulting in possible failure during usage.
- Z-0333-2012 · Boston Scientific CRM Corp · posted Dec 1, 2011 · Terminated
Current Boston Scientific defibrillator systems measure shock lead impedances up to 200 ohms, and export results to the programmer and LATITUDE monitoring system via the Daily Measurements feature. A measured shock lead impedance that is greater than 125…
- Z-0334-2012 · Philips Medical Systems (Cleveland) Inc · posted Dec 1, 2011 · Terminated
Software errors. A number of non-conformances were found for the Brilliance Big Bore and Brilliance 64 Computed Tomography X-ray Systems equipped with software version; 3.5 Beta1. The non-conformances stem from software errors identified in the operating…
- Z-0335-2012 · Philips Medical Systems (Cleveland) Inc · posted Dec 1, 2011 · Terminated
Software errors. A number of non-conformances were found for the Brilliance Big Bore and Brilliance 64 Computed Tomography X-ray Systems equipped with software version; 3.5 Beta1. The non-conformances stem from software errors identified in the operating…
- Z-0336-2012 · Acute Innovations Llc · posted Dec 1, 2011 · Terminated
RibLoc screw from lot L1008007 was out of specification. It would not engage with the hex tip properly.
- Z-0337-2012 · Atom Medical Corporation · posted Dec 1, 2011 · Terminated
The caster with lock on the incubator or infant warmer may become loose or break.
- Z-0338-2012 · Atom Medical Corporation · posted Dec 1, 2011 · Terminated
The caster with lock on the incubator may become loose or break.
- Z-0339-2012 · Atom Medical Corporation · posted Dec 1, 2011 · Terminated
The caster with lock on the incubator may become loose or break.
- Z-0340-2012 · Atom Medical Corporation · posted Dec 1, 2011 · Terminated
The caster with lock on the radiant-warming open type incubator may become loose or break.
- Z-0341-2012 · Radiometer America Inc · posted Dec 1, 2011 · Terminated
Transcutaneous pCO2-electrodes were wrongly coded as: pO2 electrode on the hardware plug for the firm's TCM series monitors. This causes the monitor to interpret the electrode wrongly, causing it to fail all calibrations and never get ready for use.
- Z-0342-2012 · Elekta, Inc. · posted Dec 1, 2011 · Terminated
A small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board.
- Z-0343-2012 · Philips Medical Systems (Cleveland) Inc · posted Dec 1, 2011 · Terminated
In response to complaints received by Philips Medical concerning certain identified software errors reported by their customers, the firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm…
- Z-0344-2012 · Philips Medical Systems (Cleveland) Inc · posted Dec 1, 2011 · Terminated
In response to complaints received by Philips Medical concerning certain identified software errors reported by their customers, the firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm…
- Z-0345-2012 · PerMedics, Inc. · posted Dec 1, 2011 · Terminated
Anomaly 1 - Invalid Export of Beam Table and or Collimator Angle for DICOM RT Ion, DICOM Image and RTP Link export files: During PerMedics' internal code reviews for lEC 61217 compliance verification, an issue was identified by PerMedics engineers in…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.