Anthem CRT-P models PM3110 and PM3210 recallZ-0430-2012 · St Jude Medical CRMD · initiated Sep 23, 2011
Terminated All recalls posted in January 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from St Jude Medical CRMD or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
St. Jude Medical has received reports of out of range pacing lead impedance (PLI) measurements. These measurements are low and out of range (<200 ohms) but measure in range during visits to the clinic. The out of range readings are concluded as false.
FDA root cause category: Pending
What the company is doing
The firm, St. Jude Medical, sent an "IMPORTANT PRODUCT FIELD CORRECTION" letter dated September 22, 2011 to its customers by FedEx priority overnight service for planned receipt on September 23, 2011. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed that in order to prevent a false reading, a new programmer software version (12.1.0.1) is available. This software when used to interrogate an Accent DR or Anthem CRT-P pacemaker will eliminate the potential for this anomaly to occur. The onetime upgrade is performed automatically on affected devices and will not change the operation of the implanted device. A St. Jude Medical Sales Representative will assist the customers in loading the new programmer software onto their Merlin programmer. Customers with questions or concerns, can contact their local St. Jude Medical representative or their Technical services department at (800) 722-3774.
Recall details
- Recall number
- Z-0430-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 23, 2011
- Posted by the FDA
- Jan 11, 2012
- Recalling firm
- St Jude Medical CRMD
- Quantity
- 14,000 units
- Product code
- DXY
- FDA submissions
- P030035, P880086
- Distribution
- Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Australia, Asia Pacific, Belgium, Bulgaria, Canada, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, India, Iran, Israel, Italy, Japan, Jordan, Korea, Kuwait, Latin America, Lebanon, Malaysia, Netherlands, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, UK, and U.A.E.
Product description
Anthem CRT-P models PM3110 and PM3210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
Code information (models, lots, serial numbers)
Models PM3110, 3112, 3210,3212 (Anthem) and All lots and serial numbers
Other devices under the same FDA submission
48 devices cleared under P030035, P880086 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Quadra Allure MP™ PM3562St. Jude Medical, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)26 deaths1,180 reports
- Zephyr™ XL DRSt. Jude Medical, Inc. · Pulse generator, permanent, implantable24 deaths944 reports
- QUADRA ALLURE MP™ PM3262St. Jude Medical, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)27 deaths518 reports
- Allure™ PM3222St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)8 deaths463 reports
- Allure Quadra™ PM3242St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)18 deaths391 reports
- Assurity™ PM1240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall20 deaths306 reports
- Victory™ 5816St. Jude Medical, Inc. · implantable pacemaker Pulse-generator5 deaths347 reports
- Endurity™ PM2160St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall2 deaths237 reports
- Zephyr® DRSt. Jude Medical, Inc. · Pulse generator, permanent, implantable9 deaths219 reports
- Accent™ DRSt. Jude Medical, Inc. · Pulse generator, permanent, implantable20 deaths214 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 60134).
- Z-0429-2012 · TerminatedImplantable Pacemaker Accent DR models PM2110 and PM2210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page