Allure Quadra™ PM3242Recalls & reported problems

St. Jude Medical, Inc. · RF CRT Cardiac Resynchronization Therapy Pacemaker DDDRV · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Allure Quadra pacemaker has 391 reports in the FDA database, with 154 injuries and 18 deaths. Many reports describe the battery running down early, the device switching to back-up mode, incorrect readings, and electrical noise being picked up as heartbeats. Some describe difficulty removing old leads during replacement surgery. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
391
Deaths reported
18
Injuries reported
154
Recalls naming it
0

Summary

The FDA's MAUDE database contains 391 adverse event reports that list this exact device (DI 05414734507912), received between Jun 20, 2018 and May 17, 2026 — classified as 18 deaths, 154 injuries, 219 malfunctions.

The most frequently reported specific device problem is “Premature Discharge of Battery” (93 reports).

5 other recalls cover devices cleared under the same FDA submission (P030035); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Discharge of Battery93
  • Adverse Event Without Identified Device or Use Problem54
  • Defective Device50
  • Pacemaker Found in Back-Up Mode35
  • Incorrect Measurement34
  • Over-Sensing30
  • Failure to Disconnect25
  • Failure to Interrogate14

Patient problems

  • No Clinical Signs, Symptoms or Conditions154
  • No Known Impact Or Consequence To Patient92
  • Insufficient Information52
  • Unspecified Infection26
  • No Consequences Or Impact To Patient19
  • Death15
  • Arrhythmia8
  • Discomfort7

Recent adverse event reports

The 25 most recent of 391 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 17, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE BATTERY OF THE PATIENT'S PACEMAKER (PM) HAD PREMATURELY DEPLETED. THE PATIENT WAS ASYMPTOMATIC. THE PM WAS EXPLANTED AND REPLACED. THE PATIENT WAS STABLE THROUGHOUT THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25196822 ↗
  • MalfunctionApr 2, 2026No Apparent Adverse Event

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 24769432 ↗
  • MalfunctionFeb 13, 2026Incorrect Measurement

    IT WAS REPORTED THAT A DIAGNOSTIC ANOMALY WAS OBSERVED ON THE DEVICE. THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED. THERE WERE NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24333020 ↗
  • MalfunctionJun 26, 2025Defective Device; Defective Device

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 22338704 ↗
  • MalfunctionJun 25, 2025Failure to Disconnect; Failure to Disconnect

    IT WAS REPORTED THAT THE PACEMAKER DEPENDENT PATIENT PRESENTED FOR A GENERATOR CHANGE PROCEDURE. IT WAS NOTED THAT THE PATIENT'S LEFT VENTRICULAR (LV) FAILED TO CAPTURE FROM THE TIME OF ITS INITIAL IMPLANT AND HAD NOT BEEN UTILIZED SINCE THEN. DURING THE PROCEDURE, IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD FAILED TO BE REMOVED FROM THE PACEMAKER HEADER. THE PHYSICIAN USED EXCESSIVE TECHNIQUE TO REMOVE THE RV LEAD AND IT WAS NOTED THAT THE RV LEAD EXHIBITED INSULATION DAMAGE. TO COMPLETE THE PROCEDURE, THE PHYSICIAN CUT, CAPPED AND REPLACED THE RV LEAD DURING THE PROCEDURE ON (B)(6) 2025. THE OLD LV LEAD WAS PLUGGED INTO THE NEW PACEMAKER AND NOT UTILIZED. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22333662 ↗
  • MalfunctionJun 18, 2025Over-Sensing; Failure to Disconnect; No Pacing; Over-Sensing; Failure to Disconnect; No Pacing

    IT WAS REPORTED THAT DURING A DEVICE REPLACEMENT PROCEDURE DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI), THE SET SCREW OF THE DEVICE WAS DIFFICULT TO BE LOOSENED. NOISE RESULTING IN OVERSENSING AND PACING INHIBITION WAS NOTED ON THE RIGHT VENTRICLE CHANNEL WHILE ACTIVELY LOOSENING THE SETSCREW, RESULTING IN THE PATIENT EXPERIENCING ASYSTOLE. LOOSENING OF THE SETSCREW WAS HALTED, DEVICE PACING CONTINUED, AND THE ASYSTOLE WAS TERMINATED. THE PHYSICIAN CONSIDERED THE ASYSTOLE EPISODES TO BE SHORT AND NOT CLINICALLY SIGNIFICANT. LOOSENING OF THE SETSCREW WAS COMPLETED AND THE DEVICE WAS EXPLANTED AND REPLACED. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: Arrhythmia

    MAUDE report 22285985 ↗
  • InjuryMay 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC DUE TO THEIR PACEMAKER ERODING THROUGH THE POCKET. THE PATIENT'S RIGHT ATRIAL (RA), RIGHT VENTRICULAR (RV), AND LEFT VENTRICULAR (LV) LEADS WERE ALL EXPOSED AND VISIBLE THROUGH THE POCKET. THE RA AND RV LEADS WERE EXPLANTED. THE PACEMAKER AND LV LEAD WERE EXPLANTED AND REPLACED. PATIENT WAS STABLE.

    Patient problems: Pocket Erosion

    MAUDE report 22101469 ↗
  • MalfunctionMay 8, 2025Failure to Disconnect; Failure to Disconnect

    DURING A ROUTINE DEVICE REPLACEMENT PROCEDURE, THE RIGHT VENTRICULAR (RV) LEAD WAS UNABLE TO BE DISCONNECTED FROM THE DEVICE. THE RV LEAD WAS CUT, CAPPED AND REPLACED DURING THE PROCEDURE TO RESOLVE THE EVENT. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21989681 ↗
  • MalfunctionApr 21, 2025Break; Loose or Intermittent Connection; Difficult to Remove; Break; Loose or Intermittent Connection; Difficult to Remove

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL FOR A SCHEDULED PACEMAKER CHANGEOUT. DURING THE PROCEDURE, THE SET SCREW OF THE PACEMAKER WAS STRIPPED, DIFFICULT TO REMOVE AND BROKE OFF BELOW THE SCREW HEAD. THE PACEMAKER WAS EXPLANTED AND REPLACED ON (B)(6) 2025. THE PATIENT WAS DISCHARGED WITH NO REPORTS OF ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21870939 ↗
  • InjuryJan 13, 2025Unexpected Mode Switch; Unexpected Mode Switch

    IT WAS REPORTED THAT THE PACEMAKER EXHIBITED A BACKUP VVI MODE. THE PATIENT HAD BEEN EXPERIENCING DISCOMFORT. THE PACEMAKER WAS EXPLANTED AND REPLACED. THE PATIENT WAS STABLE.

    Patient problems: Discomfort

    MAUDE report 21134888 ↗
  • InjuryNov 27, 2024Signal Artifact/Noise; Over-Sensing; Signal Artifact/Noise; Over-Sensing

    RELATED MANUFACTURER REFERENCE NUMBERS: 2017865-2024-71105, 2017865-2024-71106, AND 2017865-2024-71107. IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY TO THE CLINIC VIA MERLIN.NET TRANSMISSION. TRANSMISSIONS REVEALED THAT THE PACEMAKER, RIGHT ATRIAL LEAD AND LEFT VENTRICULAR LEAD EXHIBITED ELECTROMAGNETIC INTERFERENCE (EMI). THE RIGHT VENTRICULAR LEAD EXHIBITED NOISE OVERSENSING RESULTING PACING INHIBITION. NO INTERVENTION AND PATIENT CONSEQUENCES WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20793601 ↗
  • MalfunctionNov 1, 2024Defective Device; Defective Device

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 20589555 ↗
  • MalfunctionOct 7, 2024Failure to Disconnect; Failure to Disconnect

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A DEVICE CHANGE OUT PROCEDURE. DURING THE PROCEDURE, THE RIGHT VENTRICULAR (RV) LEAD WAS STUCK IN THE HEADER AND WAS UNABLE TO BE REMOVED. UPON FURTHER INTERROGATION, IT WAS NOTED THAT THE RV LEAD EXHIBITED FAILURE TO CAPTURE, LOW PACING IMPEDANCE, AND THE INSULATION WAS DAMAGED. THE RV LEAD WAS CAPPED, AND THE PACEMAKER WAS EXPLANTED AND REPLACED. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20380733 ↗
  • MalfunctionJun 17, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING ROUTINE GENERATOR EXCHANGE THAT THE PATIENT NEEDED EXTERNAL DEFIBRILLATION AND THE PHYSICIAN WONDERED IF THE PACEMAKER WAS RESPONSIBLE FOR THE ARRHYTHMIA. THE PHYSICIAN ELECTED TO EXPLANT AND REPLACED THE PACEMAKER. THE PATIENT WAS IN UNSTABLE CONDITION.

    Patient problems: Arrhythmia

    MAUDE report 19554744 ↗
  • MalfunctionMar 15, 2024Failure to Interrogate; Failure to Interrogate

    IT WAS REPORTED THAT THE DEVICE WAS UNABLE TO BE INTERROGATED VIA RADIOFREQUENCY TELEMETRY. ABBOTT TECHNICAL SUPPORT WAS CONTACTED AND THE DEVICE WAS UPGRADED AND REPROGRAMMED TO RESOLVE THE EVENT. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18916296 ↗
  • MalfunctionFeb 19, 2024Defective Device; Defective Device

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 18731528 ↗
  • InjuryJan 2, 2024Premature Discharge of Battery; No Pacing; Interrogation Problem; Premature Discharge of Battery; No Pacing; Interrogation Problem

    IT WAS REPORTED A PATIENT'S PACEMAKER PRESENTED WITH A TRANSMISSION PROBLEM VIA MERLIN.NET. DURING INITIALIZATION OF THE PROGRAMMER SESSION, THE PATIENT EXPERIENCED SYNCOPE WITH LOSS OF PACING. THERE WAS ALSO AN ALLEGATION OF PREMATURE BATTERY DEPLETION. THE PACEMAKER WAS EXPLANTED AND REPLACED IN A PROCEDURE ON (B)(6) 2023. POST-PROCEDURE THE PATIENT STATUS WAS UNKNOWN.

    Patient problems: Syncope/Fainting

    MAUDE report 18432140 ↗
  • InjuryDec 18, 2023Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH A PACEMAKER THAT WAS EXHIBITING PREMATURE BATTERY DEPLETION. THE PACEMAKER WAS EXPLANTED, AND NEW PACEMAKER WAS IMPLANTED. THE PATIENT CONDITION WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18352626 ↗
  • MalfunctionNov 27, 2023Defective Device; Defective Device

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 18212922 ↗
  • InjuryNov 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC DUE TO POCKET EROSION. THE PATIENT WAS GIVEN ANTIBIOTICS. THE ENTIRE RIGHT VENTRICULAR, LEFT VENTRICULAR, AND ATRIAL LEADS WERE EXPLANTED AND REPLACED BY THE PHYSICIAN. THE PACEMAKER WAS REPOSTIONED. THE PATIENT WAS IN STABLE CONDITION THROUGHOUT THE PROCEDURE.

    Patient problems: Pocket Erosion

    MAUDE report 18190400 ↗
  • MalfunctionNov 1, 2023Loose or Intermittent Connection; Difficult to Remove; Loose or Intermittent Connection; Difficult to Remove

    RELATED MANUFACTURER REFERENCE NUMBER: (B)(4) IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL FOR A SCHEDULED IMPLANT PROCEDURE. DURING PROCEDURE, THE RIGHT VENTRICULAR (RV) LEAD WAS FOUND DIFFICULT TO BE REMOVED FROM THE PACEMAKER HEADER. THE RV LEAD WAS REMOVED USING ANOTHER HEX WRENCH COMPATIBLE TO THE PACEMAKER HEADER SETSCREW WHILE USING SILICON OIL TO DISCONNECT THE RV LEAD PIN. THE PACEMAKER WAS SUCCESSFULLY EXPLANTED AND REPLACED. THE RV LEAD WAS CONNECTED TO THE NEW PACEMAKER AND WAS FOUND TO FUNCTION AS INTENDED. PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18051084 ↗
  • InjuryOct 17, 2023Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT PATIENT PRESENTED FOR ROUTINE GENERATOR CHANGE PROCEDURE. DURING PROCEDURE IT WAS FOUND THAT PATIENT'S ATRIAL LEAD AND RIGHT VENTRICULAR LEAD WAS NOT ABLE TO DISENGAGE FROM DEVICE HEADER IT WAS FURTHER NOTED THAT SET SCREWS OF ATRIAL AND RIGHT VENTRICULAR PORT WERE STRIPPED. THE DEVICE WAS REMAINED IMPLANTED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17951315 ↗
  • DeathSep 26, 2023Pacemaker Found in Back-Up Mode; Pacemaker Found in Back-Up Mode

    IT WAS REPORTED THAT THE PATIENT PASSED AWAY DUE TO AN UNKNOWN CAUSE. UPON INTERROGATION, THE DEVICE WAS OBSERVED TO BE IN BACKUP (BVVI) MODE. FURTHER INFORMATION WAS REQUESTED, BUT NOT YET AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17818159 ↗
  • InjurySep 13, 2023Premature Discharge of Battery; Failure to Capture; Failure to Interrogate; Premature Discharge of Battery; Failure to Capture; Failure to Interrogate

    IT WAS REPORTED THAT THE PATIENT PRESENTED DURING CLINICAL FOLLOW-UP. IN CLINIC, THE PACEMAKER PRESENTED WITH LOSS OF CAPTURE, FAILURE TO INTERROGATE, AND NO MAGNET RESPONSE. PREVIOUS INTERROGATION NOTED THAT THE DEVICE EXHIBITED PREMATURE BATTERY DEPLETION AND IT WAS NOTED THAT THE PATIENT FELT FATIGUE SINCE (B)(6) OF 2023. THE REPORTED DEVICE WAS EXPLANTED. THE PATIENT'S CONDITION WAS UNKNOWN.

    Patient problems: Fatigue

    MAUDE report 17737283 ↗
  • InjurySep 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2023-42543. IT WAS REPORTED THAT THE PATIENT PRESENTED WITH A SYSTEMIC INFECTION. THE PHYSICIAN EXPLANTED THE ENTIRE PACEMAKER SYSTEM TO RESOLVE THE ISSUE. THE PATIENT EXPERIENCED NO CONSEQUENCES.

    Patient problems: Unspecified Infection

    MAUDE report 17709257 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030035. One submission often covers many models, so check the product and lot details in each record.

  • Z-1497-2022 · Jun 17, 2022 · Open, Classified
    Merlin.net model MN5000 Software
  • Z-1496-2022 · Jun 17, 2022 · Open, Classified
    Merlin 2 PCS MER37000 programmer Model MER3400 software
  • Z-1495-2022 · Jun 17, 2022 · Open, Classified
    Merlin PCS 3650 programmer Model 3330 software
  • Z-0429-2012 · Sep 23, 2011 · Terminated
    Implantable Pacemaker Accent DR models PM2110 and PM2210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
  • Z-0430-2012 · Sep 23, 2011 · Terminated
    Anthem CRT-P models PM3110 and PM3210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →