St. Jude Medical Riata ST Silicone Endocardial… recallZ-0458-2012 · St Jude Medical CRMD · initiated Nov 28, 2011
Terminated All recalls posted in January 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from St Jude Medical CRMD or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The recall was initiated because St. Jude Medical has confirmed the failures associated with all cause insulation failure on the St. Jude Medical Riata and Riata ST Silicone Endocardial Defibrillation Leads with specific emphasis on externalized conductors.
FDA root cause category: Device Design
What the company is doing
St. Jude Medical sent a Medical Device Advisory Important Product Information Update letter dated November 28, 2011, Physician Communication dated December 15, 2010, Product Performance Report Extracts, and a Physician Device Advisory Notification to physicians by FedEx priority overnight service for receipt on November 29, 2011. The letters provides the customers with an explanation of the problem, the product, and the actions to be taken. The issuance of the recall letter is in conjunction with St. Jude's recent release November 2011 Product Performance Report (available online at http://sjmprofessional.com). The letter also explains about the Lead Peformance Summary, Root Cause, Clinical Implications, Rate of Occurrence from Complaints and Returns, New Peer Reviewed Publications, Data on Optim Insulated Leads (Durata and Riata ST Optim), and Recommendations and Mitigations. Physicians with any questions or concerns can contact their local St. Jude Medical representative or the Technical Services Department at (800) 722-3774.
Recall details
- Recall number
- Z-0458-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 28, 2011
- Posted by the FDA
- Jan 11, 2012
- Recalling firm
- St Jude Medical CRMD
- Quantity
- 128,000 units total
- Product code
- LWS
- FDA submissions
- P950022
- Distribution
- USA (Nationwide) and worldwide to: Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hungary, Iceland, Italy, Romania, Russia, Netherlands, Norway, Poland, Portugal, Russia, Serbia, Slovakia, Spain, Sweden, Switzerland, The Netherlands, UK, Australia, China, Hong Kong, Indonesia, Japan, Korea, Malaysia, Philippines, Singapore, Taiwan, Thailand, Vietnam, New Zealand, Argentina, Brazil, Colombia, Costa Rica, Ecuador, Mexico, Puerto Rico, Uruguay, Egypt, Greece, India, Iran, Israel, Joran, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, South Africa, Turkey, Malta, Cyprus, Iraq, Egypt, Lybia, Yemen, and U.A.E.
Product description
St. Jude Medical Riata ST Silicone Endocardial Defibrillation Leads Riata(7Fr), Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042 Product Usage: These Silicone transvenous high voltage (HV) leads connect an implantable cardioverter-defibrillator (lCD or CRT-D) to cardiac tissue to monitor and regulate a patients heart rate by providing pacing therapy and/or tachyarrhythmia therapy.
Code information (models, lots, serial numbers)
Riata ST models (700X, 701X, 704X)
Other devices under the same FDA submission
38 devices cleared under P950022 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- OPTISURE™ LDA210Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes69 deaths2,441 reports
- Durata™ 7120Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes59 deaths2,092 reports
- OPTISURE™ LDA210Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes43 deaths1,706 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
- Durata™ 7120Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes32 deaths1,140 reports
- Durata™ 7120St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)31 deaths857 reports
- Durata™ 7121Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes27 deaths529 reports
- Durata™ 7122Q-52St. Jude Medical, Inc. · Permanent defibrillator electrodes13 deaths510 reports
- Durata™ 7121Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes21 deaths478 reports
- Durata™ 7121-65St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)23 deaths429 reports
- Durata™ 7120St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)25 deaths345 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 60571).
- Z-0457-2012 · TerminatedSt. Jude Medical Riata (8Fr), Models: 1560, 1561, 1562, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592 Product Usage: These Silicone transvenous high voltage (HV) leads co…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.