CARTO 3 Systems, Product Code M-4800-01 recallZ-1230-2012 · Biosense Webster, Inc. · initiated Feb 9, 2012
Terminated All recalls posted in March 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biosense Webster, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Three Urgent Customer Notification recall communications were initiated 2/9/12 with Biosense forwarding the notification to customers with affected backplane cards (electrophysiology lab managers, a notification for customers with affected back plane cards (risk managers), and a notification for customers with affected backplane cards and smarttouch module along with customer acknowledgement forms. The letters provides the customers with an explanation of the problem identified and steps to take in the event that the issue is observed. If the steps provided do not resolve the issue, customers are to contact their Biosense Webster representative. A Biosense Webster representative will contact customers to schedule an appointment for the replacement of the Backplane card. Customers were instructed to complete and return the attached Acknowledgement Form in accordance with the instructions listed on the form. If customers have questions, they can contact their BWI representative.
Recall details
- Recall number
- Z-1230-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 9, 2012
- Posted by the FDA
- Mar 20, 2012
- Recalling firm
- Biosense Webster, Inc.
- Quantity
- 134 units total (32 units in US)
- Product code
- DQK
- FDA submissions
- K090017, K093455, K103746, K112007
- Distribution
- Worldwide Distribution -- US, Japan, China, Ireland, Germany, Italy, Slovenia, Indonesia, Poland, Spain, United Kingdom, Switzerland, Belgium, and Canada.
Product description
CARTO 3 Systems, Product Code M-4800-01 The intended use of the CARTO 3 System is catheter-based cardiac electrophysiological (EP) procedures.
Code information (models, lots, serial numbers)
Catalog # FG-5400-00 Serial numbers: 11996, 11998, 11999, 12002, 12026, 12031, 12040, 12046, 12047, 12051, 12052, 12057, 12061, 12062, 12063, 12065, 12066, 12067, 12069, 12070, 12075, 12077, 12078, 12086, 12087, 12088, 12089, 12094, 12096, 12098, 12099, and 12111.
Other devices under the same FDA submission
6 devices cleared under K090017, K093455, K103746, K112007 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Carto 3 CR3434CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE2 reports
- Carto 3 CB3410CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 report
- Carto 3 CB3412CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 report
- Carto 3 CB3434CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 report
- Carto 3 CY1212CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.