90-5970-020-17, NexGen CR Complete Knee Solution, Cruciate… recallZ-1567-2012 · Zimmer Inc. · initiated Feb 24, 2012
Terminated All recalls posted in May 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer has received 5 complaints where a NexGen CR micro articular surface was used with a standard CR femur, even though the compatibility chart and product labeling indicates that these components are NOT compatible. Investigation identified additional instances where a CR micro articular surface and standard CR femur were sold together on the same purchase order. CR micro articular surfaces are
FDA root cause category: Labeling design
What the company is doing
Zimmer sent an Urgent Correction Notice dated March 5, 2012, to all affected customers via telephone, E-mail and letter notifying users of the appropriate compatibility of micro sizes of NexGen CR and CR-Flex Articular Surfaces and informing them of known cases of incompatible devices being implanted, including associated risks. Customers and surgeons were asked to acknowledge receipt by calling 1-888-912-7349. Questions and concerns or help in notiying accounts about this correction were to be addressed to 1-888-548-8514. For questions regarding this recall call 574-372-4807.
Recall details
- Recall number
- Z-1567-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 24, 2012
- Posted by the FDA
- May 14, 2012
- Recalling firm
- Zimmer Inc.
- Quantity
- 83,322 all sizes
- Product code
- JWH
- FDA submissions
- K003910, K892800, K933785, K951185
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Poland, Taiwan, Denmark, Korea, Singapore, China, Canada, Australia, Japan, France, Italy, UK, Germany, Spain, Belgium, Portugal, Switzerland, Sweden, India, Thailand, Peru, The Netherlands, Malaysia, and Austria.
Product description
90-5970-020-17, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 17mm Height , Use With Plate 1,2 Sterile R, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Complete Knee System ,integrated, with an extensive cruciate retaining, cruciate substituting, and fully constrained component configurations, design specific, conforming surfaces, femoral and tibial augmentation. The femoral, patellar, articular surface, and tibial base plate components should be design specific to the condition and availability of the bone stock, and the condition and functionality of soft tissues. This allows the surgeon to select the most appropriate component combination to meet specific patient needs. Articulating surfaces on micro femoral and tibial components have a smaller bearing spacing than those on the standard and macro components.
Code information (models, lots, serial numbers)
Part 90-5970-020-17, all lots
Other devices under the same FDA submission
99 devices cleared under K003910, K892800, K933785, K951185 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NexGen® 5980-47-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall13 reports
- NexGen® 5988-20-14Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- NexGen® 5988-20-17Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- NexGen® 5986-37-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER32 reports
- NexGen® 5986-37-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER32 reports
- NexGen® 5980-37-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER23 reports
- NexGen® 5986-47-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER21 reports
- NexGen® 5972-65-32Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER20 reports
- NexGen® 5980-37-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER18 reports
- NexGen® 5980-47-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- NexGen® 5988-12-15Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- NexGen® 5986-47-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER16 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 61468).
- Z-1547-2012 · Terminated00595202010, NexGen CompleteKnee Solution, Prolong, Highly CrossLinked Polyethylene Micro Articular Surface Size Purple/A E Micro 10mm Height; Sterile, Rx ,Zimmer, Inc. Warsaw, I…
- Z-1548-2012 · Terminated00595202012, NexGen CompleteKnee Solution, Prolong, Highly CrossLinked Polyethylene Micro Articular Surface Size Purple/A E Micro 12mm Height; Sterile, Rx ,Zimmer, Inc. Warsaw, I…
- Z-1549-2012 · Terminated00595202014, NexGen CompleteKnee Solution, Prolong, Highly CrossLinked Polyethylene Micro Articular Surface Size Purple/A E Micro 14mm Height; Sterile, Rx ,Zimmer, Inc. Warsaw, I…
- Z-1550-2012 · Terminated00595202017, NexGen CompleteKnee Solution, Prolong, Highly CrossLinked Polyethylene Micro Articular Surface Size Purple/A E Micro 17 mm Height; Sterile, Rx ,Zimmer, Inc. Warsaw, …
- Z-1551-2012 · Terminated00595202020, NexGen CompleteKnee Solution, Prolong, Highly Crosslinked Polyethylene Micro Articular Surface Size Purple/A E Micro 20 mm Height; Sterile, Rx ,Zimmer, Inc. Warsaw, …
- Z-1552-2012 · Terminated00-5970-02023, NexGen CompleteKnee Solution, Prolong, Highly Crosslinked Polyethylene Micro Articular Surface with locking screw & insert, Size Purple/A E Micro 23mm Height; Ster…
- Z-1553-2012 · Terminated90595202012, NexGen CompleteKnee Solution, Prolong, Highly Crosslinked Polyethylene Micro Articular Surface with locking screw & insert, Size Purple/A E Micro 12 mm Height; Steri…
- Z-1554-2012 · Terminated00-5976-020-23 NexGen Complete Knee Solution, Prolong, Highly Crosslinked Polyethylene Micro Articular Surface with locking screw & insert, Size Purple/A E Micro 23 mm Height; St…
- Z-1555-2012 · Terminated90595202017, NexGen Complete Knee Solution, Prolong, Highly Crosslinked Polyethylene Micro Articular Surface with locking screw & insert, Size Purple/A E Micro 17 mm Height; Ster…
- Z-1556-2012 · Terminated90-5952-020-20, NexGen Complete Knee Solution, Prolong, Highly Crosslinked Polyethylene Micro Articular Surface with locking screw & insert, Size Purple/A E Micro 20mm Height; St…
- Z-1557-2012 · Terminated00-5970-020-09, NexGen CR Complete Knee Solution, Articular Surface regular constraint, Size Purple, Height; 9 mm CR Sterile, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Articular Su…
- Z-1558-2012 · Terminated00-5970-020-10, NexGen CR Complete Knee Solution, Articular Surface regular constraint, Size Purple, Height; 10 mm CR Sterile, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Articular S…
- Z-1559-2012 · Terminated00-5970-020-12, NexGen CR Complete Knee Solution, Articular Surface regular constraint, Size Purple, Height; 12 mm CR Sterile, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Articular S…
- Z-1560-2012 · Terminated00-5970-020-14, NexGen CR Complete Knee Solution, Articular Surface regular constraint, Size Purple, Height; 14 mm CR Sterile, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Articular S…
- Z-1561-2012 · Terminated00-5970-020-17, NexGen CR Complete Knee Solution, Articular Surface regular constraint, Size Purple, Height; 17 mm CR Sterile, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Complete Kne…
- Z-1562-2012 · Terminated00-5970-020-20 NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 20 mm Height , Use With Plate 1,2 (5.5 MM un…
- Z-1563-2012 · Terminated90-5970-020-09, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 9 mm Height , Use With Plate 1,2 (5.5 MM un…
- Z-1564-2012 · Terminated90-5970-020-10, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) 10 MM Height , Use With Plate 1,2 Sterile R, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Complete Knee System…
- Z-1565-2012 · Terminated90-5970-020-12, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 12mm Height , Use With Plate 1,2 Sterile R,…
- Z-1566-2012 · Terminated90-5970-020-14, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 14mm Height , Use With Plate 1,2 Sterile R,…
- Z-1568-2012 · Terminated90-5970-020-20, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 20mm Height , Use With Plate 1,2 Sterile R,…
- Z-1569-2012 · Terminated00-5972-020-09, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 9mm Height , ( 5.5MM under the Condyles) Ste…
- Z-1570-2012 · Terminated00-5972-020-10, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 10mm Height , Sterile R, Rx ,Zimmer, Inc. Wa…
- Z-1571-2012 · Terminated00-5972-020-12, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 12mm Height , Sterile R, Rx ,Zimmer, Inc. Wa…
- Z-1572-2012 · Terminated00-5972-020-14, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 14 mm Height , Sterile R, Rx ,Zimmer, Inc. W…
- Z-1573-2012 · Terminated00-5972-020-17, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 17mm Height , Sterile R, Rx ,Zimmer, Inc. Wa…
- Z-1574-2012 · Terminated00-5972-020-20, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 20mm Height , Sterile R, Rx ,Zimmer, Inc. Wa…
- Z-1575-2012 · Terminated00-5976-020-10, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Anterior Constrained AC, Size Purple, AE Micro 10mm Height , Sterile R, Rx ,Zimmer, In…
- Z-1576-2012 · Terminated00-5976-020-12, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Anterior Constrained AC, Size Purple, AE Micro 12mm Height , Sterile R, Rx ,Zimmer, In…
- Z-1577-2012 · Terminated00-5976-020-17, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Anterior Constrained AC, Size Purple, AE Micro 17mm Height , Sterile R, Rx ,Zimmer, In…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page