Skip to content

JET-X(R) 6 MM X 20 MM SHORT HALF PIN TI-6AL-4V, REF… recallZ-1748-2012 · Smith & Nephew Inc · initiated Apr 13, 2012

Terminated All recalls posted in June 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, voiding sterility.

FDA root cause category: Packaging

What the company is doing

Smith & Nephew Sales Representatives and Smith & Nephew International Distributors were notified of the problem with the affected product and the recall via email on April 13, 2012. All affected direct consignees were notified via Fed Ex on April 13, 2012. The Urgent Medical Device Recall letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to cease use and distribution of the product and quarantine them immediately for return. Customers were instructed to complete the Inventory Return Certification Form included with the letter. Customers should fax or email the form to 901-566-7975 or [email protected]. The affected product should be returned to: Smith & Nephew Attn: Global Field Actions Safety Affairs 1450 Brooks Road Memphis, Tennessee 38116 If the affected product was distributed further customers were instructed to provide a copy of the recall letter to these customers.

Recall details

Recall number
Z-1748-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 13, 2012
Posted by the FDA
Jun 6, 2012
Recalling firm
Smith & Nephew Inc
Quantity
124 units
Product code
JDW
FDA submissions
K994143
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia, Malaysia, Germany, Japan, Portugal, United Arab Emirates, Spain, France, South Africa, Norway, Sweden, Denmark, Switzerland, Belgium, China, The Netherlands, Finland, Mexico, Italy, UK, Venezuela, Colombia, Thailand, Chile, South Korea, India, and Turkey

Product description

JET-X(R) 6 MM X 20 MM SHORT HALF PIN TI-6AL-4V, REF 71056208, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.

Code information (models, lots, serial numbers)

Lot Numbers: 10GM03181, 11BM06853, 11CM12389A, 11GM06015

Other devices under the same FDA submission

27 devices cleared under K994143 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 61767).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page