JET-X(R) 6 MM X 35 MM SHORT HALF PIN TI-6AL-4V, REF… recallZ-1751-2012 · Smith & Nephew Inc · initiated Apr 13, 2012
Terminated All recalls posted in June 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, voiding sterility.
FDA root cause category: Packaging
What the company is doing
Smith & Nephew Sales Representatives and Smith & Nephew International Distributors were notified of the problem with the affected product and the recall via email on April 13, 2012. All affected direct consignees were notified via Fed Ex on April 13, 2012. The Urgent Medical Device Recall letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to cease use and distribution of the product and quarantine them immediately for return. Customers were instructed to complete the Inventory Return Certification Form included with the letter. Customers should fax or email the form to 901-566-7975 or [email protected]. The affected product should be returned to: Smith & Nephew Attn: Global Field Actions Safety Affairs 1450 Brooks Road Memphis, Tennessee 38116 If the affected product was distributed further customers were instructed to provide a copy of the recall letter to these customers.
Recall details
- Recall number
- Z-1751-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 13, 2012
- Posted by the FDA
- Jun 6, 2012
- Recalling firm
- Smith & Nephew Inc
- Quantity
- 552 units
- Product code
- JDW
- FDA submissions
- K994143
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Australia, Malaysia, Germany, Japan, Portugal, United Arab Emirates, Spain, France, South Africa, Norway, Sweden, Denmark, Switzerland, Belgium, China, The Netherlands, Finland, Mexico, Italy, UK, Venezuela, Colombia, Thailand, Chile, South Korea, India, and Turkey
Product description
JET-X(R) 6 MM X 35 MM SHORT HALF PIN TI-6AL-4V, REF 71056358, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
Code information (models, lots, serial numbers)
Lot Numbers: 09GM12206, 09JM10185, 10DM00584, 10DM00586, 10HM07426, 11BM06871, 11GM08744
Other devices under the same FDA submission
27 devices cleared under K994143 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SPATIAL FRAME 71070720Smith & Nephew, Inc. · PIN, FIXATION, THREADED15 reports
- SPATIAL FRAME 71070710Smith & Nephew, Inc. · PIN, FIXATION, THREADED9 reports
- SPATIAL FRAME 71070705Smith & Nephew, Inc. · PIN, FIXATION, THREADED7 reports
- SPATIAL FRAME 71070200Smith & Nephew, Inc. · PIN, FIXATION, THREADED7 reports
- SPATIAL FRAME 71070220Smith & Nephew, Inc. · PIN, FIXATION, THREADED5 reports
- ILIZAROV 102107Smith & Nephew, Inc. · PIN, FIXATION, THREADED5 reports
- Jet-X 71062250Smith & Nephew, Inc. · PIN, FIXATION, THREADED3 reports
- SPATIAL FRAME 71070310Smith & Nephew, Inc. · PIN, FIXATION, THREADED2 reports
- SPATIAL FRAME 71070125Smith & Nephew, Inc. · PIN, FIXATION, THREADED2 reports
- SPATIAL FRAME 71070730Smith & Nephew, Inc. · PIN, FIXATION, THREADED2 reports
- Jet-X 71065401Smith & Nephew, Inc. · PIN, FIXATION, THREADED2 reports
- Jet-X 71064002Smith & Nephew, Inc. · PIN, FIXATION, THREADED2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 61767).
- Z-1725-2012 · Terminated4.5 MM X 10 MM X 150 MM HA COATED HALF PIN SST, REF 71070823, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1726-2012 · Terminated4.5 MM X 20 MM X 150 MM HA COATED HALF PIN SST, REF 71070824, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1727-2012 · Terminated4.5 MM X 30 MM X 150 MM HA COATED HALF PIN SST, REF 71070825, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1728-2012 · Terminated4.5 MM X 40 MM X 150 MM HA COATED HALF PIN SST, REF 71070826, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1729-2012 · Terminated4.5 MM X 50 MM X 150 MM HA COATED HALF PIN SST, REF 71070827, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1730-2012 · Terminated6 MM X 15 MM X 150 MM HA COATED HALF PIN SST, REF 71070828, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1731-2012 · Terminated6 MM X 20 MM X 150 MM HA COATED HALF PIN SST, REF 71070800, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1732-2012 · Terminated6 MM X 25 MM X 150 MM HA COATED HALF PIN SST, REF 71070802, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1733-2012 · Terminated6 MM X 30 MM X 150 MM HA COATED HALF PIN SST, REF 71070804, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1734-2012 · Terminated6 MM X 35 MM X 150 MM HA COATED HALF PIN SST, REF 71070806, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1736-2012 · Terminated6 MM X 45 MM X 150 MM HA COATED HALF PIN SST, REF 71070835, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1737-2012 · Terminated6 MM X 50 MM X 150 MM HA COATED HALF PIN, 1.6 MM CANNULATED, SST, REF 71070822, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1738-2012 · Terminated6 MM X 50 MM X 150 MM HA COATED HALF PIN SST, REF 71070810, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1739-2012 · Terminated6 MM X 55 MM X 150 MM HA COATED HALF PIN SST, REF 71070812, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1740-2012 · Terminated6 MM X 60 MM X 150 MM HA COATED HALF PIN SST, REF 71070814, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1741-2012 · Terminated6 MM X 65 MM X 150 MM HA COATED HALF PIN SST, REF 71070816, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1742-2012 · Terminated6 MM X 70 MM X 150 MM HA COATED HALF PIN SST, REF 71070818, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1743-2012 · Terminated6 MM X 20 MM X 200 MM HA COATED HALF PIN SST, REF 71070801, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1744-2012 · Terminated6 MM X 25 MM X 200 MM HA COATED HALF PIN SST, REF 71070803, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1745-2012 · Terminated6 MM X 30 MM X 200 MM HA COATED HALF PIN SST, REF 71070805, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1746-2012 · Terminated6 MM X 35 MM X 200 MM HA COATED HALF PIN SST, REF 71070807, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1747-2012 · TerminatedJET-X(R) BAR 5 MM X 45 MM BAR SHORT HALF PIN TI-6AL-4V, REF 71055451, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1748-2012 · TerminatedJET-X(R) 6 MM X 20 MM SHORT HALF PIN TI-6AL-4V, REF 71056208, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1749-2012 · TerminatedJET-X(R) 6 MM X 25 MM SHORT HALF PIN TI-6AL-4V, REF 71056258, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1750-2012 · TerminatedJET-X(R) 6 MM X 30 MM SHORT HALF PIN TI-6AL-4V, REF 71056308, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1752-2012 · TerminatedJET-X(R) 6 MM X 40 MM SHORT HALF PIN TI-6AL-4V, REF 71056408, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1753-2012 · TerminatedJET-X(R) 6 MM X 45 MM SHORT HALF PIN TI-6AL-4V, REF 71056458, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1754-2012 · TerminatedJET-X(R) 6 MM X 50 MM SHORT HALF PIN TI-6AL-4V, REF 71056508, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1755-2012 · TerminatedJET-X(R) 6 MM X 55 MM SHORT HALF PIN TI-6AL-4V, REF 71056558, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
- Z-1756-2012 · TerminatedJET-X(R) 6 MM X 60 MM SHORT HALF PIN TI-6AL-4V, REF 71056608, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page