4.5mm TI Multiloc Screw Length 28MM-Sterile recallZ-2060-2012 · Synthes USA HQ, Inc. · initiated May 22, 2012
Class II Terminated All recalls posted in July 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Two lots of 4.5mm multi-loc screws were recalled due to incorrect labeling of screw size.
FDA root cause category: Error in labeling
What the company is doing
Synthes sent a "NOTICE: MEDICAL DEVICE RECALL" letter dated May 22, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Verification Form was enclosed for customers to complete and return via fax to 610-251-9005. Contact the firm at 610-719-5450 for questions regarding this recall.
Recall details
- Recall number
- Z-2060-2012
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 22, 2012
- Posted by the FDA
- Jul 23, 2012
- Recalling firm
- Synthes USA HQ, Inc.
- Quantity
- 20
- Product code
- HSB
- FDA submissions
- K103002
- Distribution
- Nationwide Distribution-including the states of AZ, CA, FL, GA, IL, LA, MD, MN, NJ, NY, OH, PA, TN, TX, and VA.
Product description
4.5mm TI Multiloc Screw Length 28MM-Sterile Expiration: 04/29/2012 REF 04.019.028S Manufactured or Distributed by Synthes (USA) 1101 Synthes Avenue, Monument CO 80132 4.5MM TI Multiloc Screw Length 28MM-Sterile Ref 04.019,038S. The Multiloc Screw 4.5 is used with a Multiloc PHN for treatment of proximal humerus fractures and is intended to be inserted through the Multiloc PHN until it engages the subchondral bone below the relative distal cortex (cortex closest to the humeral head, below the articular surface).
Code information (models, lots, serial numbers)
Lot #2721745 Expiration 4/29/2012 REF 04.019.028S
Other devices under the same FDA submission
69 devices cleared under K103002 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MULTILOC 03.019.006Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES41 reports
- MULTILOC 03.019.008Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES39 reports
- MULTILOC 03.019.007Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES38 reports
- NA 04.005.434Synthes GmbH · Screw, fixation, bone11 reports
- MULTILOC 03.019.024Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES10 reports
- NA 04.005.424Synthes GmbH · Screw, fixation, bone9 reports
- NA 04.005.422Synthes GmbH · Screw, fixation, bone8 reports
- NA 04.005.428Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES8 reports
- NA 04.005.432Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES7 reports
- NA 04.005.436Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES7 reports
- NA 04.005.418Synthes GmbH · Screw, fixation, bone5 reports
- NA 04.005.420Synthes GmbH · Screw, fixation, bone5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 62307).
- Z-2059-2012 · Terminated4.5mm TI Multiloc Screw Length 38MM-Sterile Expiration: 04/29/2012 REF 04.019.038S Manufactured or Distributed by Synthes (USA) 1101 Synthes Avenue, Monument CO 80132 4.5MM TI Mu…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page