Newport HT70 and HT70 Plus Ventilators, Model Number: HT70… recallZ-2250-2012 · Newport Medical Instruments Inc · initiated Aug 3, 2012
Class I Terminated All recalls posted in April 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Newport Medical Instruments Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
May emit a continuous high priority alarm and the ventilator may stop ventilating, due to a component failure on the control board.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Newport Medical sent an Urgent Field Correction Notice letter dated August 3, 2012 to all affected customers, via DHL and FedEx courier services. The letter identified the affected product, reason for recall and actions to be taken. Customers were instructed to immediately stop using the affected ventilator and retrieve the replacement kit provided free of charge. In addition, the letter states that if the product was further distributed, to contact their customers and notify them of the recall. Customers were asked to complete and return the enclosed Replacement form as replacements are completed. For questions contact Newport Medical Technical Support Department at 800-451-3111, extension 282, for U.S. customers; 714-427-5811, extension 500, for international customers.".
Recall details
- Recall number
- Z-2250-2012
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Aug 3, 2012
- Posted by the FDA
- Apr 14, 2014
- Recalling firm
- Newport Medical Instruments Inc
- Quantity
- 72
- Product code
- CBK
- FDA submissions
- K090888, K111146
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Brazil, Chile, Egypt, France, Hong Kong, India, Japan, Norway, Philippines and South Africa.
Product description
Newport HT70 and HT70 Plus Ventilators, Model Number: HT70 and HT70 Plus. Product Usage: The HT70 ventilator system is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs.). The HT70 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. The HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
Code information (models, lots, serial numbers)
Model Number: HT70 and HT70 Plus. Serial numbers: N12HT700514226 to N12HT700514231, N12HT700514233 to N12HT700514236, N12HT700514238 to N12HT700614280, N12HT700614282 to N12HT700614299, N12HT700614301 to N12HT700614309, N12HT720410266 to N12HT720410268, N12HT720410271, N12HT720410273, N12HT720410282, N12HT720410283, N12HT720410286, N12HT720410288, N12HT720410289, N12HT720410290, N12HT720410294 to N12HT720510300, N12HT720510302, N12HT720510322 to N12HT720510330, N12HT720510332 to N12HT720510344.
Other devices under the same FDA submission
15 devices cleared under K090888, K111146 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Newport HT70M-JP-NANewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE89 reports
- Newport HT70M-SY-EUNewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall17 reports
- Newport HT70PM-WW-EUNewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall4 reports
- Newport DLHT70PM-WW-NANewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 report
- Newport HT70M-WW-EUNewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 report
- Newport HT70PM-PT-BRNewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall0 reports
- Newport HT70PM-SY-UKNewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall0 reports
- Newport HT70PM-WW-UKNewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall0 reports
- Newport HT70PM-ES-EUNewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall0 reports
- Newport HT70PM-ES-NANewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall0 reports
- Newport HT70PM-PT-BRNewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall0 reports
- Newport HT70PM-SY-NANewport Medical Instruments, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page