ATLANTIS(R) Anterior Cervical Plate System FIXED ANGLE… recallZ-0037-2013 · Medtronic Sofamor Danek USA Inc · initiated Aug 30, 2012
Class II Terminated All recalls posted in October 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medtronic conducted a conference call on August 30, 2012 and sent an "URGENT VOLUNTARY RECALL" notice dated September 3, 2012 to all affected customers. The notice identified the product, problem, and actions to be taken by the customers. A Recall Questionnaire Form was attached to the letter for customers to complete and return. Contact your Medtronic Sales representative or the firm's Global Quality Department at 800-876-3133, ext. 3197 for questions regarding this notice.
Recall details
- Recall number
- Z-0037-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 30, 2012
- Posted by the FDA
- Oct 10, 2012
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Quantity
- 1014 units
- Product code
- KWQ
- FDA submissions
- K021461, K063100, K081038
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of India, Netherlands, Hong Kong, S. L Korea, Singapore, Canada, Australia, Japan, and Mexico.
Product description
ATLANTIS(R) Anterior Cervical Plate System FIXED ANGLE SCREW MODULE, REF 176-504, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.
Code information (models, lots, serial numbers)
All lots except those starting with SY
Other devices under the same FDA submission
29 devices cleared under K021461, K063100, K081038 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ATLANTIS® Anterior Cervical Plate System 3120515Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY12 reports
- ATLANTIS® Anterior Cervical Plate System 3120215Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY8 reports
- ATLANTIS® Anterior Cervical Plate System 3120516Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY7 reports
- ATLANTIS® Anterior Cervical Plate System 3120313Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY5 reports
- ATLANTIS® Anterior Cervical Plate System 3120314Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY4 reports
- ATLANTIS® Anterior Cervical Plate System 3120517Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY4 reports
- ATLANTIS® Anterior Cervical Plate System 3120513Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY3 reports
- ATLANTIS® Anterior Cervical Plate System 3120514Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY3 reports
- ATLANTIS® Anterior Cervical Plate System 3125517Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY3 reports
- ATLANTIS® Anterior Cervical Plate System 3120416Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- ATLANTIS® Anterior Cervical Plate System 6190062Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- ATLANTIS® Anterior Cervical Plate System 6190045Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 63157).
- Z-0036-2013 · TerminatedATLANTIS(R) Anterior Cervical Plate System VARIABLE ANGLE SCREW MODULE, REF 176-503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee…
- Z-0038-2013 · TerminatedPREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthoped…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page