Creatinine, Part Numbers OSR6178, OSR6678 recallZ-0958-2013 · Beckman Coulter Inc. · initiated Dec 28, 2012

Class II Terminated All recalls posted in March 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Beckman Coulter is including additional information and instructions in the Creatinine (OSR6x78) IFU and Setting Sheet, relating to calibration stability and quality control frequency to assure product performance.

FDA root cause category: Labeling design

What the company is doing

Beckman Coulter sent an Urgent Product Correction letter to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed the customers that Beckman Coulter is including additional information and instructions in the Creatinine (OSR6x78) IFU and Setting Sheet, relating to calibration stability and quality control frequency to assure product performance. Customers are instructed to complete and return the enclosed response form. Customers with questions about the notification letter are instructed to contact the call center (hotline) at (800) 854-3633 in the US and Canada.

Recall details

Recall number
Z-0958-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 28, 2012
Posted by the FDA
Mar 15, 2013
Recalling firm
Beckman Coulter Inc.
Quantity
57,962 units (50,452 units in the US)
Product code
CGX
FDA submissions
K924692, K934361
Distribution
Worldwide Distribution - USA Nationwide and the countries of: Canada, Brazil, Colombia, Mexico, Puerto Rico, Venezuela, Guam, Honduras, Peru, Ecuador, Costa Rica.

Product description

Creatinine, Part Numbers OSR6178, OSR6678 Product Usage: System reagent for the quantitative determination of Creatinine in human serum and urine on Beckman Coulter AU analyzers.

Code information (models, lots, serial numbers)

All

Other devices under the same FDA submission

1 device cleared under K924692, K934361 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.