INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162380, TRIGEN… recallZ-0943-2013 · Smith & Nephew Inc · initiated Feb 7, 2013
Class II Terminated All recalls posted in March 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label.
FDA root cause category: Labeling mix-ups
What the company is doing
Smith & Nephew sales representatives and Smith & Nephew international distributors were notified of problem and the recall via email on February 7, 2013. They were instructed to cease use and distribution of the product and quarantine for return. Non responding consignees were notified again on February 28, 2013 by email. Hospitals were notified by letter beginning on February 7, 2013. Smith & Nephew also contacted the surgeons who implanted the screws by letter the week of February 11th. For questions regarding this recall call 978-749-1330.
Recall details
- Recall number
- Z-0943-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 7, 2013
- Posted by the FDA
- Mar 13, 2013
- Recalling firm
- Smith & Nephew Inc
- Quantity
- 69 units
- Product code
- JDS
- FDA submissions
- K981529
- Distribution
- Worldwide Distribution - USA including NM, TN, TX, NC, CA, MI, KY, MA, NJ, OH, and HI. Internationally to Canada, United Arab Emirates, Mexico, Argentina, Japan, and Chile.
Product description
INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162380, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 80 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for intramedullary rod fixation.
Code information (models, lots, serial numbers)
Batch Number 12KM01206
Other devices under the same FDA submission
27 devices cleared under K981529 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- TRIGEN 71634000Smith & Nephew, Inc. · NAIL, FIXATION, BONE5 reports
- TRIGEN 71634010Smith & Nephew, Inc. · NAIL, FIXATION, BONE5 reports
- TRIGEN 71642235Smith & Nephew, Inc. · NAIL, FIXATION, BONE3 reports
- TRIGEN 71642230Smith & Nephew, Inc. · NAIL, FIXATION, BONE2 reports
- TRIGEN 71634005Smith & Nephew, Inc. · NAIL, FIXATION, BONE2 reports
- TRIGEN 71634020Smith & Nephew, Inc. · NAIL, FIXATION, BONE2 reports
- TRIGEN 71642300Smith & Nephew, Inc. · NAIL, FIXATION, BONE2 reports
- TRIGEN 71934786Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 report
- TRIGEN 71634440Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 report
- TRIGEN 71634340Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 report
- TRIGEN 71634338Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 report
- TRIGEN 71634240Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 64342).
- Z-0944-2013 · TerminatedINTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 75 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page