Draeger Infinity Acute Care System (IACS) Monitoring… recallZ-0939-2013 · Draeger Medical, Inc. · initiated Dec 13, 2012

Class II Terminated All recalls posted in March 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

FDA root cause category: Labeling design

What the company is doing

Drager sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-800-543-5047 for questions relating to this notice.

Recall details

Recall number
Z-0939-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 13, 2012
Posted by the FDA
Mar 11, 2013
Recalling firm
Draeger Medical, Inc.
Quantity
260
Product code
MHX
FDA submissions
K093788, K113798
Distribution
Distributed in the states of CA, IL, and MA.

Product description

Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Code information (models, lots, serial numbers)

Catalog Number MS20783 (Infinity MCable-Dual Hemo)

Other devices under the same FDA submission

5 devices cleared under K093788, K113798 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page