Alaris PC unit model 8015 with software version 9.12 recallZ-1112-2013 · Carefusion Corporation · initiated Mar 6, 2013

Class II Terminated All recalls posted in April 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Carefusion Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The recall was initiated because Carefusion has identified potential risk associated with bolus programming with the Alaris PC unit (model 8015) software version 9.12. Automating infusion pump parameter input from either the electronic medical record (EMR) or the Alaris Auto-ID module for a continuous infusion with a bolus dose option may result in a loss of information from the Guardrails Data s

FDA root cause category: Software design

What the company is doing

Carefusion sent an Urgent Medical Device Recall Notification letter dated March 6, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers are required to confirm receipt of the notification by returning the Recall response card to Carefusion by postage-paid, self-addressed mail, fax, or email. The letter stated that a Carefusion representatives will contact customers by telephone to schedule field remediation. For questions call Carefusion Support Center at 888-562-6018, Customer Advocacy at 888-812-3266, and/or Technical Support at 888-812-3229.

Recall details

Recall number
Z-1112-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 6, 2013
Posted by the FDA
Apr 15, 2013
Recalling firm
Carefusion Corporation
Quantity
4,090 total units
Product code
FRN
FDA submissions
K051641
Distribution
Worldwide Distribution USA Nationwide and the countries of Australia and Canada.

Product description

Alaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System

Code information (models, lots, serial numbers)

not available

Other devices under the same FDA submission

1 device cleared under K051641 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.