MRT-1504/A4 - Titan X MRI System recallZ-1101-2013 · Toshiba American Medical Systems Inc · initiated Oct 15, 2012

Class II Terminated All recalls posted in April 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Toshiba American Medical Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to operate normally, and the risk of a serious health hazard is therefore very low. The LCD Moni

FDA root cause category: Nonconforming Material/Component

What the company is doing

Toshiba sent an Urgent Medical Device Correction letter dated October 1, 2012 via USPS return receipt mail to all affected customers . The letter identified the affected product, problem and actions to be taken. The letter contains a return reply form that is to be faxed to TAMS for retention. For questions contact your local Toshiba representative at 800-521-1968.

Recall details

Recall number
Z-1101-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 15, 2012
Posted by the FDA
Apr 11, 2013
Recalling firm
Toshiba American Medical Systems Inc
Quantity
27 units
Product code
LNH
FDA submissions
K113093, K120638
Distribution
USA Nationwide Distribution including the states of : IL, NH, MA, MI, MO, PR, NV, NM, MN, ID, NY, OH, NC, AZ, MS, LA, OR, PA, CA, and CT.

Product description

MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System.

Code information (models, lots, serial numbers)

Product Numbers: MRT-1504/A4, MRT-1504/S4, MRT-3010/A5, MZKT-UPKHE/S1. Serial Numbers unique to each consignee.

Other devices under the same FDA submission

2 devices cleared under K113093, K120638 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.