90 Degree Cannulated Infant Blade Plate 25mm x 5mm x 3 Hole recallZ-1534-2013 · OrthoPediatrics Corp · initiated Apr 19, 2013
Class II Terminated All recalls posted in June 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from OrthoPediatrics Corp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.
FDA root cause category: Device Design
What the company is doing
OrthoPediatrics sent an Urgent Voluntary Recall notifiation via telephone and email to all affected consignees. The letter instructed consignees to take possession of any affected devices and quarantine immediately. A Mandatory Reply Form is included with the notification letter to be completed and returned. Please contact Greg Teghtmeyer if you have any questions (574-268-6379).
Recall details
- Recall number
- Z-1534-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 19, 2013
- Posted by the FDA
- Jun 12, 2013
- Recalling firm
- OrthoPediatrics Corp
- Quantity
- 130 devices were shipped
- Product code
- HWC
- FDA submissions
- K110959
- Distribution
- USA Nationwide Distribution
Product description
90 Degree Cannulated Infant Blade Plate 25mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, intertrochanteric valgus osteotomies, proximal and distal tibial osteotomies and humeral fractures and osteotomies.
Code information (models, lots, serial numbers)
All lot numbers are affected. Catalog No. 00-1200-1000
Other devices under the same FDA submission
15 devices cleared under K110959 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LCB 00-1200-1002Orthopediatrics Corp. · PLATE, FIXATION, BONE6 reports
- LCB 00-1200-2001Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-2002Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-4503Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-4518Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-4519Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-5000Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-5002Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-5003Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-7001Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-7002Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
- LCB 00-1200-7003Orthopediatrics Corp. · PLATE, FIXATION, BONE1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 65091).
- Z-1535-2013 · Terminated90 Degree Cannulated Infant Blade Plate 30mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractur…
- Z-1536-2013 · Terminated90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractur…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page