Medical device recalls: June 2013
The FDA posted 177 medical device recall records in June 2013, from 95 recall events by 75 companies. 23 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 5 events, 5 records
- Baxter Healthcare Corp. · 3 events, 10 records
- Medtronic Neuromodulation · 3 events, 8 records, 8 Class I
- Bard Peripheral Vascular Inc · 3 events, 6 records
- American Optisurgical Inc · 3 events, 3 records
- Biomet, Inc. · 3 events, 3 records
- Hill-Rom, Inc. · 2 events, 5 records
- Prism Medical Services USA · 2 events, 5 records
FDA root cause
- Nonconforming Material/Component · 40
- Device Design · 35
- Under Investigation by firm · 15
- Software design · 12
- Process control · 12
- Use error · 8
- Employee error · 7
- No Marketing Application · 6
Class I recalls (23)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1570-2013 · Medtronic Neuromodulation · posted Jun 25, 2013 · Terminated
Medtronic is providing Healthcare Professionals with important safety information and patient management recommendations regarding the unintended delivery of drug during the priming bolus function for the SynchroMed implantable infusion pump. The…
- Z-1571-2013 · Medtronic Neuromodulation · posted Jun 25, 2013 · Terminated
Medtronic is providing Healthcare Professionals with important safety information and patient management recommendations regarding the unintended delivery of drug during the priming bolus function for the SynchroMed implantable infusion pump. The…
- Z-1579-2013 · Medtronic Neuromodulation · posted Jun 25, 2013 · Terminated
Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a…
- Z-1580-2013 · Medtronic Neuromodulation · posted Jun 25, 2013 · Terminated
Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a…
- Z-1573-2013 · Medtronic Neuromodulation · posted Jun 24, 2013 · Terminated
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous…
- Z-1574-2013 · Medtronic Neuromodulation · posted Jun 24, 2013 · Terminated
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous…
- Z-1575-2013 · Medtronic Neuromodulation · posted Jun 24, 2013 · Terminated
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous…
- Z-1576-2013 · Medtronic Neuromodulation · posted Jun 24, 2013 · Terminated
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous…
- GlideScope Video Laryngoscope (GVL) 3, 4, 5 and GlideScope (AVL) 2, 3, 4 and 5 reusable blades ↗Class IZ-1558-2013 · Verathon, Inc. · posted Jun 20, 2013 · Terminated
Verathon GlideScope Reusable GVL and AVL Blades are being recalled due to the potential risk of premature failure/breakage of the blade tip that may not be readily visible during routine inspection before or after intubation.
- Z-1537-2013 · Respironics California Inc · posted Jun 17, 2013 · Terminated
Respironics California, Inc. has initiated a recall on the V60 ventilator because of an issue with the software on the V60 Power Management Board Assembly. If a component fails on the Power Management Board Assembly, it may cause ventilator support to be…
- Z-1436-2013 · Endologix Inc · posted Jun 7, 2013 · Terminated
Endologix, Inc. initiated this voluntary recall of certain lots of Endologix, Inc. due to reports of dilator breakage during clinical procedures.
- Z-1452-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1453-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1454-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1455-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1456-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1457-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1458-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1459-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1460-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1461-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1462-2013 · Symbios Medical Products, LLC · posted Jun 6, 2013 · Terminated
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being…
- Z-1420-2013 · Medtronic Xomed, Inc. · posted Jun 4, 2013 · Terminated
In March 2013, Medtronic issued a recall of the NIM TriVantage EMG Endotracheal Tube due to cuff leakage in the intraoperative cuff deflation which then requires re-inflation or replacement of the deflated tube in order to ensure the continued ventilation…
Class II recalls (148)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1616-2013 · GE Healthcare, LLC · posted Jun 28, 2013 · Terminated
GE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.
- Z-1617-2013 · Biosense Webster, Inc. · posted Jun 28, 2013 · Terminated
Biosense Webster is recalling the 20-Pole LASSO NAV Catheter Eco because it has been incorrectly calibrated causing them to be recognized and displayed as a 10-Pole LASSO NAV Catheter Eco by the CARTO 3 System.
- Z-1618-2013 · Bard Access Systems · posted Jun 28, 2013 · Terminated
BARD Access Systems is recalling lot JUWJF283 of StatLock CV Plus Stabilization Butterfly Devices due to incorrect pad assembly used during manufacture. The absence of the hydrocolloid strip on the bottom may pose an incremental risk including localized…
- Z-1600-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality.…
- Z-1601-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality.…
- Z-1602-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality.…
- Z-1603-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality.…
- Z-1604-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality.…
- Z-1605-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality.…
- Z-1606-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality.…
- Z-1607-2013 · Medtronic Inc. Cardiac Rhythm Disease Management · posted Jun 27, 2013 · Terminated
Medtronic recently identified an issue with a subset of Consulta CRT-P and Syncra CRT-P devices during production. As of May 30, 2013, there have been no reported or confirmed device failures. However, because of the potential for malfunction, Medtronic is…
- Z-1608-2013 · Greatbatch Medical · posted Jun 27, 2013 · Terminated
Cordis Corporation, Inc and Greatbatch Medical are recalling one lot of the cordis RADIALSOURCE Spring Wire Transradial Access Kit product. Some units were manufactured with an undersized vessel dilator tip, too small to slide over the guidewire.
- Z-1609-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.
- Z-1610-2013 · Baxter Healthcare Corp. · posted Jun 27, 2013 · Terminated
Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.
- Pinnacle Destination Peripheral Guiding Sheath, 8 French, 90cm, Straight Tip, Cross Cut Valve. ↗Class IIZ-1611-2013 · Terumo Medical Corporation · posted Jun 27, 2013 · Terminated
Medical devices were incorrectly labeled with extended expiration dates.
- Z-1612-2013 · Terumo Medical Corporation · posted Jun 27, 2013 · Terminated
Medical devices were incorrectly labeled with extended expiration dates.
- Z-1613-2013 · Beckman Coulter Inc. · posted Jun 27, 2013 · Terminated
Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may…
- Z-1519-2013 · GE Healthcare, LLC · posted Jun 26, 2013 · Terminated
GE Healthcare has recently become aware of a potential issue due to inaccessible Operator Instructions associated with the defective Operator Manual CD of the Senographe Essential.
- Quick Connect QLC1677E, QRC1699E, QPC1721E and QFC1683E. STERIS Corporation, Mentor, OH 44060Class IIZ-1596-2013 · Steris Corporation · posted Jun 26, 2013 · Terminated
During a routine label review, the firm identified four SYSTEM 1E Quick Connects that will be updated to reflect the removal of obsolete device models. This labeling review process also highlighted the opportunity to combine two existing quick connects…
- Z-1597-2013 · RF Technologies, Inc. · posted Jun 26, 2013 · Terminated
RF Technologies, Inc. has initiated a recall involving CA520 Code Alert Integrated Care Management Advanced 4-Way Care Solution . Due to a firmware error, the alarm may not be transmitted to the system server immediately.
- Z-1599-2013 · ConBio, a CynoSure Company · posted Jun 26, 2013 · Terminated
An inconsistency was identified related to electronic display identification of the optional Multilite Dye Handpieces of the RevLite SI Laser system that may confuse user.
- Z-1586-2013 · Medrad Inc · posted Jun 25, 2013 · Terminated
The firm is recalling these power cables due to a latent design reliability issue and the potential for shorting which can result in heating/melting of the cable jacket.
- Z-1588-2013 · Philips Medical Systems (Cleveland) Inc · posted Jun 25, 2013 · Terminated
If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information page is the patient name selected with…
- Z-1589-2013 · Medtronic Sofamor Danek USA Inc · posted Jun 25, 2013 · Terminated
Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
- Z-1590-2013 · Medtronic Sofamor Danek USA Inc · posted Jun 25, 2013 · Terminated
Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
- Z-1591-2013 · Medtronic Sofamor Danek USA Inc · posted Jun 25, 2013 · Terminated
Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
- Z-1592-2013 · Medtronic Sofamor Danek USA Inc · posted Jun 25, 2013 · Terminated
Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
- Z-1593-2013 · Medtronic Sofamor Danek USA Inc · posted Jun 25, 2013 · Terminated
Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
- Z-1594-2013 · Medtronic Sofamor Danek USA Inc · posted Jun 25, 2013 · Terminated
Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
- Z-1595-2013 · Medtronic Sofamor Danek USA Inc · posted Jun 25, 2013 · Terminated
Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
- CoaguChek XS System (PST)Class IIZ-1581-2013 · Roche Diagnostics Operations, Inc. · posted Jun 24, 2013 · Terminated
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
- CoaguChek XS Pro SystemClass IIZ-1582-2013 · Roche Diagnostics Operations, Inc. · posted Jun 24, 2013 · Terminated
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
- CoaguChek XS Plus SystemClass IIZ-1583-2013 · Roche Diagnostics Operations, Inc. · posted Jun 24, 2013 · Terminated
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
- CoaguChek XS System (HCP)Class IIZ-1584-2013 · Roche Diagnostics Operations, Inc. · posted Jun 24, 2013 · Terminated
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
- Z-1585-2013 · GE Healthcare It · posted Jun 24, 2013 · Terminated
If environment variables MICRO_CORRECT_MSG_TYPE and CORRECT_MSG_TYPE are set to different values, in Microbiology result entry when released results (Final Report released) are changed and the Corrected Result messages are edited on the free-text form,…
- Z-1587-2013 · U-systems Inc · posted Jun 24, 2013 · Terminated
Devices labeled for an intended use not included in the existing 510(k) or PMA, in that, the user manual contains the Indication for Use of the soma v Platinum ABUS [a 510(k) product].
- Z-1562-2013 · Siemens Medical Solutions USA, Inc · posted Jun 21, 2013 · Terminated
The potential for a safety issue, for the image quality degradation, that is addressed by the Customer Safety Advisory Notice when verifying the patient position based on 2D images that were acquired as single-exposure MV or In-Line kView images using the…
- Z-1563-2013 · Microgenics Corporation · posted Jun 21, 2013 · Terminated
It has been confirmed from in-house testing that Direct Bilirubin (DBIL), and Total Bilirubin (TBIL) claimed in Thermo Scientific MAS Bilirubin Systems are displaying reduced stability.
- Z-1564-2013 · Medtronic Inc. Cardiac Rhythm Disease Management · posted Jun 21, 2013 · Terminated
Medtronic has identified a labeling issue for 24 units of the Launcher coronary Guide Catheter. The two affected catheter lots are mislabeled and do not accurately reflect the catheters contained within each package.
- Z-1565-2013 · Baxter Healthcare Corp. · posted Jun 21, 2013 · Terminated
leaking pouches
- Z-1566-2013 · Loma Vista Medical · posted Jun 21, 2013 · Terminated
Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum.
- CARESCAPE Monitor B850 ↗Class IIZ-1567-2013 · GE Healthcare, LLC · posted Jun 21, 2013 · Terminated
GE Healthcare become aware of 4 potential issues associated with the CARESCAPE Monitor B850. These issues include CARESCAPE Monitor B850 uDOM failure, CARESCAPE Monitor B850 unstable USB communication, CARESCAPE Monitor B850 remote printing issue with…
- Z-1569-2013 · Terumo Cardiovascular Systems Corporation · posted Jun 21, 2013 · Terminated
Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail.
- Z-1572-2013 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2013 · Terminated
The specified part numbers and lot numbers of Maquet Pediatric Arterial Cannula Packs may contain an incorrectly sized connector.
- Z-1577-2013 · Becton Dickinson & Company · posted Jun 21, 2013 · Terminated
There has been an increase in complaints regarding the needle/safety mechanism disengaging from the syringe.
- Z-1578-2013 · Prism Medical Services USA · posted Jun 21, 2013 · Terminated
The firm replaced the recalled carry bar with a plastic insert with a carry bar with a metal insert due to improper installation of the recalled carry bars.
- Z-1561-2013 · Linde Gas North America Llc · posted Jun 20, 2013 · Terminated
A number of MEDICYL-e Portable Oxygen Systems may have misapplied valve flow knobs.
- Z-1552-2013 · Prism Medical Services USA · posted Jun 19, 2013 · Terminated
The product does not meet the firm's size standards, the product was produced in the incorrect size.
- Z-1553-2013 · Prism Medical Services USA · posted Jun 19, 2013 · Terminated
The product does not meet the firm's size standards, the product was produced in the incorrect size.
- Z-1554-2013 · Prism Medical Services USA · posted Jun 19, 2013 · Terminated
The product does not meet the firm's size standards, the product was produced in the incorrect size.
- Z-1555-2013 · Prism Medical Services USA · posted Jun 19, 2013 · Terminated
The product does not meet the firm's size standards, the product was produced in the incorrect size.
- Z-1556-2013 · Ortho-Clinical Diagnostics · posted Jun 19, 2013 · Terminated
The firm is recalling the specific lot due to lower than expected results.
- Z-1557-2013 · Synthes USA HQ, Inc. · posted Jun 19, 2013 · Terminated
The Synthes One-Third Tubular DCL Plate (11-Hole, 1/3 Tubular Plate w/Collar), was released to the United States market in error. This action is not being initiated as a result of adverse events.
- Z-1559-2013 · Hitachi Medical Systems America Inc · posted Jun 19, 2013 · Terminated
Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to…
- Z-1560-2013 · Tecan US, Inc. · posted Jun 19, 2013 · Terminated
WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt and the door has been left open, the script will progress (continue without…
- Z-1550-2013 · IDEV Technologies, Inc. · posted Jun 18, 2013 · Terminated
There was an error on the outer labeling of this device. The inner label was correct.
- GE Healthcare, PET VCAR. ↗Class IIZ-1551-2013 · GE Healthcare, LLC · posted Jun 18, 2013 · Terminated
GE Healthcare has recently become aware of a potential safety issue with respect to the automated PERCIST response criteria calculation applied in the PET VCAR application on Advantage Workstation or AW Server.
- Z-1447-2013 · Philips Healthcare Inc. · posted Jun 17, 2013 · Terminated
Installation Acceptance Test Document (IATD) non compliant with regulations IATDs (Installation Acceptance Test could not be located or different templates have been used
- Z-1448-2013 · Philips Healthcare Inc. · posted Jun 17, 2013 · Terminated
Installation Acceptance Test Document (IATD) non compliant with regulations. Records shows table height measurements out of specification with the consequence of potential system mis-alignment.
- Z-1449-2013 · Philips Healthcare Inc. · posted Jun 17, 2013 · Terminated
An inspection shows several IATDs (Installation Acceptance Test could not be located or different templates have been used
- Z-1450-2013 · Philips Healthcare Inc. · posted Jun 17, 2013 · Terminated
An inspection of IATD (Installation Acceptance Test Document)IA SID (Source Image Distance) records shows table height measurements out of specification with the consequence of potential system mis-alignment
- Z-1548-2013 · Biomet, Inc. · posted Jun 17, 2013 · Terminated
Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to…
- Z-1549-2013 · Aperio Technologies Inc · posted Jun 17, 2013 · Terminated
Aperio initiated this recall because the User Guides incorrectly state the devices had been cleared for specific applications. The User Guides should state the devices are for Research Use Only and are not for use in diagnostic procedures.
- Z-1547-2013 · Biomet, Inc. · posted Jun 15, 2013 · Terminated
Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part…
- Z-1544-2013 · Ormco/Sybronendo · posted Jun 14, 2013 · Terminated
A recall has been issued by SybronEndo because a Canadian Standards Association (CSA)-directed cautionary statement label was inadvertently omitted from the Elements Obturation Unit and its Replacement A/C Power Cord.
- Z-1545-2013 · Ormco/Sybronendo · posted Jun 14, 2013 · Terminated
A recall has been issued by SybronEndo because a Canadian Standards Association (CSA)-directed cautionary statement label was inadvertently omitted from the Elements Obturation Unit and its Replacement A/C Power Cord.
- Z-1546-2013 · Zimmer, Inc. · posted Jun 14, 2013 · Terminated
The bed frame adapter assemblies may have a weak weld where the patient helper adapter mounting bars attach to the tube.
- Z-1539-2013 · Abbott Medical Optics, Inc. · posted Jun 13, 2013 · Terminated
Under specific conditions, there is an unlikely potential to result in inadvertent laser firing of the laser.
- Z-1540-2013 · Abbott Medical Optics, Inc. · posted Jun 13, 2013 · Terminated
Under specific conditions, there is an unlikely potential to result in inadvertent laser firing of the laser.
- Z-1541-2013 · Abbott Medical Optics, Inc. · posted Jun 13, 2013 · Terminated
Under specific conditions, there is an unlikely potential to result in inadvertent laser firing of the laser.
- Z-1542-2013 · Edwards Lifesciences, LLC · posted Jun 13, 2013 · Terminated
Edwards is recalling the QuickDraw Venous Cannula due to complaints of cannula separation during withdrawal of the device when the Percutaneous insertion method is used.
- Z-1543-2013 · Utah Medical Products, Inc · posted Jun 13, 2013 · Terminated
Utah Medical Products, Inc. is recalling certain Deltran Disposable Pressure Transducer Kits due to complaints related to packaging integrity of the sterile packaged devices.
- Z-1533-2013 · Bard Peripheral Vascular Inc · posted Jun 12, 2013 · Terminated
Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the hub may read 7mm x 4cm, when the…
- Z-1534-2013 · OrthoPediatrics Corp · posted Jun 12, 2013 · Terminated
Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.
- Z-1535-2013 · OrthoPediatrics Corp · posted Jun 12, 2013 · Terminated
Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.
- Z-1536-2013 · OrthoPediatrics Corp · posted Jun 12, 2013 · Terminated
Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.
- ECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack with Tyvek lid. ↗Class IIZ-1530-2013 · Ethicon Endo-Surgery Inc · posted Jun 11, 2013 · Terminated
Ethicon Endo-Surgery is initiating a voluntary worldwide recall for ECHELON 60mm Black Reload (ECR60T) due to the potential for incomplete staple line formation from reload damage during the firing sequence.
- VIA Medical Collection Bag ↗Class IIZ-1531-2013 · International Biomedical, Ltd. · posted Jun 11, 2013 · Terminated
International Biomedical has received reports of interference between the collection bag and the extension set connectors. Evaluation of the affected components indicates that the collection bag connector can become occluded due to an over-tightening of…
- embla Embletta Gold (It functions as a Polysomnography Recorder and Ventilator Effort Recorder). ↗Class IIZ-1532-2013 · Embla System Llc · posted Jun 11, 2013 · Terminated
Calibration error for Embla Embletta Gold units calibrated through service center between June 15, 2012 to February 27, 2013. This calibration error could cause a clinician to over titrate the patient and prescribe a therapy pressure higher than is necessary.
- Z-1526-2013 · Bard Peripheral Vascular Inc · posted Jun 10, 2013 · Terminated
Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard UltraClip Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization…
- Z-1527-2013 · Bard Peripheral Vascular Inc · posted Jun 10, 2013 · Terminated
Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard UltraClip Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization…
- Z-1528-2013 · Bard Peripheral Vascular Inc · posted Jun 10, 2013 · Terminated
Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard UltraClip Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization…
- Z-1529-2013 · Bard Peripheral Vascular Inc · posted Jun 10, 2013 · Terminated
Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard UltraClip Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization…
- AMICUS Exchange KitClass IIZ-1523-2013 · Fenwal Inc · posted Jun 8, 2013 · Terminated
Fenwal has initiated a voluntary Urgent Product Recall for lot FA12C07070 of product code R4R2339, Amicus Exchange Kits. Fenwal identified a labeling issue with this batch of Product Code R4R2339 (which was cleared for European use only) in which this…
- Z-1524-2013 · Bard Peripheral Vascular Inc · posted Jun 8, 2013 · Terminated
Bard Peripheral Vascular (BPV) has confirmed that some devices with the recalled product code / lot number combination may have the metal tissue marker separated from the needle prior to use.
- Siemens brand MicroScan Dried Negative Breakpoint Combo 47 panels (B1017-417), SMN #: 10483099), ↗Class IIZ-1525-2013 · Siemens Healthcare Diagnostics, Inc. · posted Jun 8, 2013 · Terminated
The product has the potential for false negative or delayed reaction for Arginine (ARG) results. In addition, 42 hour identification panel holds may occur due to a weak ARG reaction.
- Z-1517-2013 · Invatec Llc · posted Jun 7, 2013 · Terminated
Devices labeled for an intended use was not included in the existing 510(k) or PMA, in that, the Indications for use was modified.
- Z-1518-2013 · American Optisurgical Inc · posted Jun 7, 2013 · Terminated
American Optisurgical, Inc. is recalling certain lots of TX1 Tissue Removal System disposable handpiece due to inventory being released into distribution without a cleared FDA 510(k) to cover recent product changes. Use of this product may lead to an…
- Z-1520-2013 · DePuy Orthopaedics, Inc. · posted Jun 7, 2013 · Terminated
The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as…
- Applied Medicals Epix and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. ↗Class IIZ-1521-2013 · Applied Medical Resources Corp · posted Jun 7, 2013 · Terminated
Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the…
- Z-1522-2013 · Dental EZ Stardental Division · posted Jun 7, 2013 · Terminated
Recall was initiated because a manufacturing issue was found that prevented the diamond coating from adhering properly to the shank of the diamond bur for these lots.
- Z-1487-2013 · AGFA Corp. · posted Jun 6, 2013 · Terminated
Customers could potentially experience intermittent, unintended and illogical movement when using the product.
- Z-1488-2013 · Berchtold Corp. · posted Jun 6, 2013 · Terminated
The firm received complaints for unintended movements.
- Z-1489-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1490-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1491-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1492-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1493-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1494-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1495-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1496-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1497-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1498-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1499-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1500-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1501-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1502-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1503-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- Z-1504-2013 · Stryker Instruments Div. of Stryker Corporation · posted Jun 6, 2013 · Terminated
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life.…
- FAST Procedure Kit. ↗Class IIZ-1505-2013 · American Optisurgical Inc · posted Jun 6, 2013 · Terminated
American Optisurgical, Inc initiated a recall of ChloraPrep Hi-Lite Orange 3mL Applicator, in response to a sub-recall letter from Medline Industries, Inc due to lack of sterility assurance caused by an incomplete seal on the individual unit packaging.
- Z-1506-2013 · Alcon Grieshaber AG · posted Jun 6, 2013 · Terminated
Alcon is conducting a Medical Device Removal for the Constellation Pneumatic Handpiece, DSP, product number 725.01. This action is being taken because the forceps or scissors tips (sold separately) may dislodge from the hand piece, if not attached by the…
- Z-1507-2013 · Cordis Corporation · posted Jun 6, 2013 · Terminated
Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
- Z-1508-2013 · Cordis Corporation · posted Jun 6, 2013 · Terminated
Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
- Z-1509-2013 · Cordis Corporation · posted Jun 6, 2013 · Terminated
Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
- Z-1510-2013 · Cordis Corporation · posted Jun 6, 2013 · Terminated
Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
- Z-1511-2013 · Cordis Corporation · posted Jun 6, 2013 · Terminated
Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
- Z-1512-2013 · Siemens Healthcare Diagnostics, Inc · posted Jun 6, 2013 · Terminated
iPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability
- Z-1513-2013 · Siemens Healthcare Diagnostics, Inc · posted Jun 6, 2013 · Terminated
iPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability
- Z-1514-2013 · Siemens Healthcare Diagnostics, Inc · posted Jun 6, 2013 · Terminated
iPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability
- Z-1515-2013 · Siemens Healthcare Diagnostics, Inc · posted Jun 6, 2013 · Terminated
iPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability
- Z-1516-2013 · American Optisurgical Inc · posted Jun 6, 2013 · Terminated
The recall has been initiated due to an overheating protection voltage suppressor on the redundant safety protection circuit board. Secondary to the recall, American Optisurgical, Inc. will be making a correction to add spacers between the valve and…
- Z-1463-2013 · Intuitive Surgical, Inc. · posted Jun 5, 2013 · Terminated
Intuitive Surgical is implementing a voluntary Field Action to implement a modified inspection process for the 5 mm Cannula (PN 420011) and 5 mm Flared Cannula (PN 420262).
- da Vinci 5 mm Flared Instrument Cannulae for use with the da Vinci S and Si Surgical Systems. ↗Class IIZ-1464-2013 · Intuitive Surgical, Inc. · posted Jun 5, 2013 · Terminated
Intuitive Surgical is implementing a voluntary Field Action to implement a modified inspection process for the 5 mm Cannula (PN 420011) and 5 mm Flared Cannula (PN 420262).
- Active Articulation E1.Class IIZ-1465-2013 · Biomet, Inc. · posted Jun 5, 2013 · Terminated
Biomet is recalling Active Articulation E1 Hip Bearing implants. Biomet has initiated this action following an investigation which identified that outer label indicates two conflicting O.D. sizes while the inner labels indicate an incorrect O.D. size. The…
- Z-1466-2013 · Hill-Rom, Inc. · posted Jun 5, 2013 · Terminated
Complaints have been reported for brakes disengaging when the bed is subjected to a hard jolt or force to the frame.
- Z-1467-2013 · Covidien LP · posted Jun 5, 2013 · Terminated
Seals may disengage from the cannula which may result in a component inadvertently disengaging into the patients cavity
- Z-1468-2013 · Covidien LP · posted Jun 5, 2013 · Terminated
Seals may disengage from the cannula which may result in a component inadvertently disengaging into the patients cavity
- Z-1469-2013 · Covidien LP · posted Jun 5, 2013 · Terminated
Seals may disengage from the cannula which may result in a component inadvertently disengaging into the patients cavity
- Z-1470-2013 · Covidien LP · posted Jun 5, 2013 · Terminated
Seals may disengage from the cannula which may result in a component inadvertently disengaging into the patients cavity
- Z-1471-2013 · Covidien LP · posted Jun 5, 2013 · Terminated
Seals may disengage from the cannula which may result in a component inadvertently disengaging into the patients cavity
- Z-1472-2013 · Covidien LP · posted Jun 5, 2013 · Terminated
Seals may disengage from the cannula which may result in a component inadvertently disengaging into the patients cavity
- SlingBar 350. ↗Class IIZ-1473-2013 · Hill-Rom, Inc. · posted Jun 5, 2013 · Terminated
Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting…
- SlingBar 360. ↗Class IIZ-1474-2013 · Hill-Rom, Inc. · posted Jun 5, 2013 · Terminated
Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting…
- SlingBar Wide 670. ↗Class IIZ-1475-2013 · Hill-Rom, Inc. · posted Jun 5, 2013 · Terminated
Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting…
- SlingBar Standard. ↗Class IIZ-1476-2013 · Hill-Rom, Inc. · posted Jun 5, 2013 · Terminated
Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting…
- E-Z WIDER BARIATRIC CHAIR BED ↗Class IIZ-1480-2013 · Kreg Medical Inc. · posted Jun 5, 2013 · Terminated
An EZ Wider bed (capacity of 1000 lb. safe working load collapsed during patient use, leaving the patient in a forward tilt position.
- Z-1481-2013 · Philips Medical Systems (Cleveland) Inc · posted Jun 5, 2013 · Terminated
Philips has received one report from the field that there was an unexpected motion of the detector due to a failure of the mechanical assembly.
- Z-1482-2013 · SpineFrontier, Inc. · posted Jun 5, 2013 · Terminated
During two separate occasions an impactor plate on the handle broke off while being struck by a mallet.
- Z-1483-2013 · Zimmer Surgical Inc · posted Jun 5, 2013 · Terminated
This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This…
- Z-1484-2013 · Zimmer Surgical Inc · posted Jun 5, 2013 · Terminated
This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This…
- Z-1485-2013 · Zimmer Surgical Inc · posted Jun 5, 2013 · Terminated
This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This…
- Z-1486-2013 · Coloplast Manufacturing US, LLC · posted Jun 5, 2013 · Terminated
Coloplast is conducting a recall on three lots of Digitex Absorbable Polydiaxanone Suture, Size 2-0 because it is mislabeled with a three year expiration date.
- Z-1451-2013 · Maytex Corp · posted Jun 4, 2013 · Terminated
Recent Lab testing conducted in March 2013 indicates that the product does not meet the standard of "Differential Pressure of only 1.6 Delta P (mm H2O/cm2)" as indicated on the box/package of the product; and also product manufactured between 2008 - 2012…
- Z-1369-2013 · Shimadzu Medical Systems · posted Jun 3, 2013 · Terminated
The recall was initiated because Shimadzu Corporation has identified a potential issue in certain manufacturing lots of the Shimadzu MobileDaRt Evolution in combination with CXDI-70C, CXDI-55C, and CXDI-55G. As a result of Shimadzu's investigation, there…
- Z-1443-2013 · GE Healthcare, LLC · posted Jun 3, 2013 · Terminated
GE Healthcare has recently become aware of a potential safety issue involving the reusable CO2 absorbent canister accessory used with Aespire, Aespire View, Avance, Avance CS2, Aisys, ADU, 9100, M900, M904, 9300 systems. Reusable CO2 absorbent canisters…
- Z-1444-2013 · I-Flow LLC · posted Jun 3, 2013 · Terminated
I-Flow initiated two voluntary product removals of the ON-Q Pain Relief System: T-Block Needle/Catheter Tray and T-Block Needle/Catheter Kit because they may have potentially malfunctioned.
- Z-1445-2013 · I-Flow LLC · posted Jun 3, 2013 · Terminated
I-Flow initiated two voluntary product removals of the ON-Q Pain Relief System: T-Block Needle/Catheter Tray and T-Block Needle/Catheter Kit because they may have potentially malfunctioned.
- Z-1446-2013 · Leica Microsystems, Inc. · posted Jun 3, 2013 · Terminated
An LPC Fine Mesh Cassette failed to stay closed during processing operations.
Class III recalls (6)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1614-2013 · Ortho-Clinical Diagnostics · posted Jun 27, 2013 · Terminated
While performing interval testing, Ortho Clinical Diagnostics (OCD) determined that a surfactant was unintentionally added to VITROS Chemistry Products K+ Slides during the manufacturing process.
- VITROS Chemistry Products CRP Slides (Reagent), REF/Catalog Number 192 6740 and 809 7990, ↗Class IIIZ-1615-2013 · Ortho-Clinical Diagnostics · posted Jun 27, 2013 · Terminated
While performing interval testing Ortho Clinical Diagnostics (OCD) determined that an excess amount of surfactant was unintentionally added to VITROS Chemistry Products CRP Slides during the manufacturing process.
- CMC Saddle ↗Class IIIZ-1568-2013 · Integra LifeSciences Corp. · posted Jun 21, 2013 · Terminated
As a result of a quality review, labelling content discrepancies were identified between the Instruction for Use (IFU) distributed in the US with the Ascension CMC and the content that is cleared by FDA.
- Z-1477-2013 · Bion Enterprises Ltd · posted Jun 5, 2013 · Terminated
Affected products have a labeled expiration date of one year too long.
- Z-1478-2013 · Bion Enterprises Ltd · posted Jun 5, 2013 · Terminated
Affected products have a labeled expiration date of one year too long.
- Z-1479-2013 · Bion Enterprises Ltd · posted Jun 5, 2013 · Terminated
Affected products have a labeled expiration date of one year too long.