Versaport Bladeless Optical Trocar With (2) Fixation… recallZ-1470-2013 · Covidien LP · initiated May 6, 2013
Class II Terminated All recalls posted in June 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Seals may disengage from the cannula which may result in a component inadvertently disengaging into the patients cavity
FDA root cause category: Device Design
What the company is doing
Covidien sent an Urgent Medical Device Recall letter dated May 6, 2013,via federal express to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the product and return all inventory to Covidien Surgical Solutions, 195 McDermott Road, North Haven, CT 06473. Customers were requested to return completed Response Form. Customers with questions were instructed to contact their Covidien representative or Covidien Customer Service, Monday through Friday, Sam - 6:30pm ET, at (800) 962-9888, option 1, and then option 2. For questions regarding this recall call 203-492-8165.
Recall details
- Recall number
- Z-1470-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 6, 2013
- Posted by the FDA
- Jun 5, 2013
- Recalling firm
- Covidien LP
- Quantity
- 5364 units
- Product code
- GCJ
- FDA submissions
- K112349
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to the following countries: Canada Austria Bahrain Benoni Centurion Chatsworth Czech Republic Denmark Egypt Finland France Germany Ireland Israel Italy Japan Kuwait Netherlands Norway Portugal QATAR Saudi Arabia Serbia Singapore South Africa Spain OMAN Sweden Switzerland UAE United Kingdom Venezuela
Product description
Versaport Bladeless Optical Trocar With (2) Fixation Cannulae - 5mm Product Code: ONB5STF2C The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynecologic, general, thoracic and urologic endoscopic procedures to create and maintain a port of entry.
Code information (models, lots, serial numbers)
Lot Numbers: N2J0211X through N3D0118X Note: The specific lot numbers listed below are not affected by this recall and are acceptable for use: N2H0045X,N2H0166X,N2H0286X, N2H0357X,N2 H051 6UX
Other devices under the same FDA submission
10 devices cleared under K112349 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VersaOne ONB5STFCovidien LP · Laparoscope, general & plastic surgery1 death143 reports
- VersaOne UNVCA5STFCovidien LP · Laparoscope, General & Plastic Surgery76 reports
- VersaOne ONB5SHFCovidien LP · Laparoscope, general & plastic surgery65 reports
- Versaport ONB5SHFCovidien LP · Laparoscope, general & plastic surgery48 reports
- Versaport ONB5STFCovidien LP · Laparoscope, general & plastic surgery44 reports
- VersaOne UNVCA5SHFCovidien LP · Laparoscope, General & Plastic Surgery17 reports
- VersaOne ONB5STF2CCovidien LP · Laparoscope, general & plastic surgery7 reports
- Versaport ONB5LGFCovidien LP · Laparoscope, general & plastic surgery3 reports
- VersaOne ONB5LGFCovidien LP · Laparoscope, general & plastic surgery2 reports
- Versaport ONB5STF2CCovidien LP · Laparoscope, general & plastic surgery1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 65093).
- Z-1467-2013 · TerminatedVersaport Bladeless Optical Trocar With Fixation Cannula - 5mm Long Product Code: ONB5LGF The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gyneco…
- Z-1468-2013 · TerminatedVersaport Bladeless Optical Trocar With Fixation Cannula - 5mm Short Product Code: ONB5SHF The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynec…
- Z-1469-2013 · TerminatedVersaport Bladeless Optical Trocar With Fixation Cannula - 5mm Product Code: ONB5STF The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynecologic,…
- Z-1471-2013 · TerminatedVersaport Bladeless Optical Fixation Cannula - 5mm Short Product Code: ONBFCA5SH The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynecologic, ge…
- Z-1472-2013 · TerminatedVersaport Bladeless Optical Fixation Cannula - 5mm Product Code: ONBFCA5ST The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynecologic, general, …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page