Active Articulation E1 recallZ-1465-2013 · Biomet, Inc. · initiated Apr 26, 2013
Class II Terminated All recalls posted in June 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Biomet is recalling Active Articulation E1 Hip Bearing implants. Biomet has initiated this action following an investigation which identified that outer label indicates two conflicting O.D. sizes while the inner labels indicate an incorrect O.D. size. The implant itself is etched with the correct O.D. size.
FDA root cause category: Labeling mix-ups
What the company is doing
Biomet sent an Urgent Medical Device Recall letter dated April 26, 2013 to consignees detailing that the outer label indicates two conflicting O.D. sizes while the inner labels indicate an incorrect O.D. size. This action requires the immediate location and discontinued use of the product and its return to Biomet. Ciustomers were advised to: Immediately locate and remove the identified device(s) listed from circulation . -Carefully follow the instructions on the enclosed "FAX Back Response Form" . - Fax a copy of the Response Form to 574-372-1683 prior to return of product. -Use priority carrier for your shipment. - If you have further distributed this product, you MUST notify hospital personnel of this action via the enclosed "Dear O.R. Manager" notice. This letter MUST be given to hospital personnel responsible for receiving recall notices. However, you are charged with the location and return of these products.
Recall details
- Recall number
- Z-1465-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 26, 2013
- Posted by the FDA
- Jun 5, 2013
- Recalling firm
- Biomet, Inc.
- Quantity
- 8 units
- Product code
- LPH
- FDA submissions
- K101336
- Distribution
- Distributed in New York.
Product description
Active Articulation E1. Intended for use with either primary or revision hip arthroplasty.
Code information (models, lots, serial numbers)
Part numbers: EP-200162, EP-200162, EP-200164, and EP-200166 with Lot numbers: 027050, 101470, 115950, 101480, 247990, 948810, 967270, 027070, 287820, 729870, and 860880.
Other devices under the same FDA submission
20 devices cleared under K101336 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Active Articulation Hip System EP-200148Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented19 reports
- Active Articulation Hip System EP-200146Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED18 reports
- Active Articulation Hip System EP-200152Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented17 reports
- Active Articulation Hip System EP-200150Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented14 reports
- Active Articulation Hip System EP-200144Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented9 reports
- M2A MAGNUM HIP SYSTEM US157852Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- Active Articulation Hip System EP-200154Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Active Articulation Hip System EP-200156Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented2 reports
- M2a Magnum™ Hip System US157852Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- M2A MAGNUM HIP SYSTEM 139258Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- M2A MAGNUM HIP SYSTEM 139254Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- M2A MAGNUM HIP SYSTEM 139252Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page