DePuy Mitek FMS Intermediary Tubing with One-Way valve recallZ-0191-2014 · DePuy Mitek, Inc., a Johnson & Johnson Co. · initiated Aug 29, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Mitek, Inc., a Johnson & Johnson Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
FDA root cause category: Component change control
What the company is doing
DePuy Mitek inititaed telephone notification on Ausgust 29, 2013 and sent an Urgent Voluntary Product Recall letter dated September 5, 2013 to US and OUS affected customers/affiliates. The customers were notified of the affected products, problem and actions to be taken. The customers were instructed to immediately check all inventories to locate and return affected product following the enclosed instructions. Stericycle is cooridianting the returns. If you have any questions or concerns in regards to this recall, please contact Stericycle directly at 1-866-737-1928.
Recall details
- Recall number
- Z-0191-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 29, 2013
- Posted by the FDA
- Nov 8, 2013
- Recalling firm
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Quantity
- 368 pk
- Product code
- HRX
- FDA submissions
- K951843
- Distribution
- Worldwide Distribution - USA (Nationwide) and in the countries of: Australia , Austria , Belgium , Canada , Czech Republic , Denmark , Finland , France, Germany , India, Ireland , Israel, Italy , Latvia , Netherlands , Norway , Nepal, Poland , Portugal , Spain , Sweden , Switzerland , United Kingdom.
Product description
DePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The Intermediary Tubing for FMS Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Intermediary Tubing is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy .
Code information (models, lots, serial numbers)
All lots with a D identifier
Other devices under the same FDA submission
6 devices cleared under K951843 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- FMS 284508Medos International Sàrl · ARTHROSCOPE304 reports
- FMS 284504Depuy Mitek, LLC · ARTHROSCOPE81 reports
- FMS 281142Medos International Sàrl · ARTHROSCOPE38 reports
- FMS 284649Medos International Sàrl · ARTHROSCOPE34 reports
- FMS 284610Medos International Sàrl · ARTHROSCOPE20 reports
- FMS 281103Medos International Sàrl · ARTHROSCOPE6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 66145).
- Z-0192-2014 · TerminatedDePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS Fluid Management Systems connects the Inflow Tubing to the ar…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page