Skip to content

NexGen Complete Knee Solutions Stemmed Tibial Component… recallZ-0480-2014 · Zimmer, Inc. · initiated Oct 31, 2013

Class II Terminated All recalls posted in December 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two complaints were received in which it was reported that the stem would not lock into the tibia. Evaluation of the returned device found that the stemmed portion of the tibial plate would not hold the stem as the connection was too loose. The taper diameter was found to be oversized by .010.

FDA root cause category: Device Design

What the company is doing

All distributors will be notified via electronic mail. Hospital risk managers and distributors with affected inventory will also be notified via courier. Surgeons that have implanted this manufacturing lot will be identified and notified by the distributor. Distributors will be sent a letter identifying the issue and their responsibilities. These responsibilities include locating and removing the affected product in their territory, as well as identifying and notifying surgeons that have implanted this manufacturing lot. Distributors will return on-hand affected products to Zimmer and ensure all of their products are accounted for using Attachment 1 provided in the letter. Distributors are to complete and return Attachment 2 indicating which surgeons have implanted the product and to which they have provided notice to. Hospitals that received direct shipments from Zimmer will be sent a letter identifying the issue and their responsibilities. These responsibilities include locating the affected product and returning it to their Zimmer sales representative. Communications outside of the United States will occur approximately two weeks after the United States communications. Zimmer will conduct effectiveness checks in the following way: Distributors/Hospitals o 100% of the FedEx notifications will be accounted for o An inventory certification form will need to be returned from each distributor that certifies that all available affected product from their territory has been returned. o Additional notifications via email and FedEx will be sent to those who have not completed the required certification form o Accounts will be deemed unresponsive after 3 attempts. Further questions please call 800-613-6131.

Recall details

Recall number
Z-0480-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 31, 2013
Posted by the FDA
Dec 11, 2013
Recalling firm
Zimmer, Inc.
Quantity
128 units
Product code
JWH
FDA submissions
K933785
Distribution
Worldwide Distribution - USA (nationwide) including the states of CA, IN, OH, GA, WI, IA, NE, NJ, PA, NY, MD, NC, AZ, NM, TX, CO, IL, AL, MI, WA, FL and KS, and the countries of China, Japan and Taiwan.

Product description

NexGen Complete Knee Solutions Stemmed Tibial Component Precoat NexGen Tibial Baseplates are part of the NexGen systems of semi-constrained, non-linked condylar knee prosthesis. The device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis; Collagen disorders, and/or avascular necrosis of the femoral condyle; Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; Moderate valgus, varus, or flexion deformities.

Code information (models, lots, serial numbers)

00-5980-037-02; lot 6228417

Other devices under the same FDA submission

89 devices cleared under K933785 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page