Medfusion 4000 Syringe Infusion Pumps recallZ-0540-2014 · Smiths Medical ASD, Inc. · initiated Nov 22, 2013
Class II Terminated All recalls posted in December 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Smiths Medical is conducting a recall involving all Medfusion 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po
FDA root cause category: Device Design
What the company is doing
Smiths Medical sent an "Urgent Medical Device Correction" letter dated November 22, 2013, to all affected customers.The letter identified the product the problem and the action needed to be taken by the customers. All customers will be provided with instructions on inspection, cleaning, and prevention regarding this issue, for the continued safe and effective use of these pumps. Additionally, customers will be instructed to: 1. Return damaged pumps for servicing or damaged AC power cords for replacement to Smiths Medical. 2. Return the completed Confirmation Form to acknowledge receipt and understanding of the information provided in the Urgent Medical Device Correction letter, via fax or email, to Smiths Medical for processing. "The Instructions for preventing this issue", "Mitigations", Conclusions", " Action to be taken by the User", and "Transmission of this Urgent Medical Device Correction". The consignees were also asked to complete the "Urgent Medical Device confirmation Form" and fax it to 1-800-237-8033 or by sending an electronic copy via e-mail to [email protected]. For questions regarding this information, consignees should contact their local Sales Representative, Smiths Medical's Customer Service Department at 1-800-258-5361, or via e-mail at [email protected]. UPDATE: Consignees were sent a follow-up letter dated 13 January 2014 from the one dated November 22, 2013. The letter was addressed to Risk / Safety Managers, Biomedical Professionals, Clinicians who oversee the use of Medfusion Pumps, Distributors, and other users of these devices. The "Urgent Medical Device Recall" letter included Advise on Action to be Taken by the User. This included customers to have their Pumps undergo a free correction. The correction consists of the installation of a new AC power Receptacle seal. Smith Medical provided options for completing the correction process which included: 1) Consignees were a
Recall details
- Recall number
- Z-0540-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 22, 2013
- Posted by the FDA
- Dec 19, 2013
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 9560 (8601 US, 959 OUS)
- Product code
- FRN
- FDA submissions
- K111386
- Distribution
- Worldwide Distribution - USA (nationwide ) including the states of : AZ, NY, TN, MS, AR, WV, VA, VT, PA, DE, DC, OH, CO, NH, MI, CA, FL, CT, WA, MD, NM, RI, MA, TX, NC, IL, KY, WI and MI., and the countries of CANADA, SAUDI ARABIA and AUSTRALIA.
Product description
Medfusion 4000 Syringe Infusion Pumps Medfusion 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.
Code information (models, lots, serial numbers)
All Medfusion¿ 4000 Syringe Pumps in the field are affected by this Field Safety Corrective Action.
Other devices under the same FDA submission
14 devices cleared under K111386 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Medfusion 4000-0106-01Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls4,390 reports
- Medfusion 4000-0105-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls2,882 reports
- Medfusion 3500-0600-01Icu Medical, Inc. · Pump, infusion2 recalls2,041 reports
- Medfusion 4000-0101-51Icu Medical, Inc. · Pump, infusion2 recalls1 death1,795 reports
- Medfusion 3500-0600-82Icu Medical, Inc. · Pump, infusion2 recalls89 reports
- MEDFUSION 4000-0106-00Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls10 reports
- Medfusion 3500-0600-00Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls9 reports
- MEDFUSION 4000-0106-231Icu Medical, Inc. · Pump, infusion2 recalls5 reports
- Medfusion 3500-0600-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls2 reports
- Medfusion 4000-0105-249Icu Medical, Inc. · Pump, infusion2 recalls1 report
- Medfusion 4000-0105-78Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls0 reports
- MEDFUSION 4000-0107-01Icu Medical, Inc. · Pump, infusion1 recall3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page