Multiples Anspach Micro Curved Attachments (MCA) and Curved… recallZ-0708-2014 · The Anspach Effort, Inc. · initiated Oct 26, 2012

Class II Terminated All recalls posted in January 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from The Anspach Effort, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During a recent review of documentation Anspach determined that the current directions for use for the MCA cutting burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA burrs. The MCA burrs are intended for use in delicate bone in otologic procedures such as cochleostomy. The use of the device in dense bone removal may result in cu

FDA root cause category: Labeling False and Misleading

What the company is doing

Anspach sent an Urgent Medical Removal letter dated October 26, 2012, to all affected customers. The letter stated that they were initiating a Medical Device removal regarding the MCA Cutting Burrs. During a recent review of documentation it was determined that the current directions for use for the MCA cutting burrs (item 18-0079) and MCA Micro Curved Attachment (item 18-0060) did not provide clear details regarding the intended use specifically for the MCA burrs. The MCA burrs are intended for use in delicate bone in otologic procedures such as cochleostomy. The use of the device in dense bone removal may result in cutter/ burr fracture. If a fracture occurs, contact Customer Support at (800) 327-6887 or email [email protected]. If the procedure was done successfully, no additional patient follow up is necessary. Customer Immediate Actions: 1. Please remove and return any MCA items from your inventory immediately. 2. Complete the attached reply form indicating your receipt of this letter. Return the completed form by fax or email to the number or email address provided on the form. If you distribute any of the products to other services or facilities, please forward this information as appropriate. Customers with questions were instructed to contact Anspach Customer Support at (800) 327-6887 or email [email protected]. For questions regarding this recall call 1-800-327-6887.

Recall details

Recall number
Z-0708-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 26, 2012
Posted by the FDA
Jan 9, 2014
Recalling firm
The Anspach Effort, Inc.
Quantity
6,824 imits (4,7324 in USA / 2,100 Foreign)
Product code
HBE
FDA submissions
K042783
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, Great Britain, France, Sweeden, Italy, Spain, Japan, South Africa, Australia, Switzerland,

Product description

Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters and Attachments) MCA - MICRO CURVED ATTACHMENT, MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM, MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM, MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM, MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM, MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM, MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM, MCA5-1SB - lMM FLUTED BALL, 5.6CM, MCA5-1SD - lMM DIAMOND BALL, 5.6CM, MCA5-2SB - 2MM FLUTED BALL, 5.6CM, MCA5-2SD - 2MM DIAMOND BALL, 5.6CM, MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM, MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM, MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM, MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM, MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM, MCA7-15SD 1.5MM DIAMOND BALL, 7.3CM, MCA7-1SB - lMM FLUTED BALL, 7.3CM, MCA7-1SD - 1 MM DIAMOND BALL, 7 .3CM, MCA7-2SB - 2MM FLUTED BALL, 7.3CM and MCA7-2SD - 2MM DIAMOND BALL, 7.3CM. The MCA attachment is a reusable device used with dissection tools. Dissection tools are single use, disposable cutter designed for cutting and shaping delicate bone, primary in otology procedures for delicate bone cutting.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

All Lots for  MCA - MICRO CURVED ATTACHMENT,  MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM,  MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM,  MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM,  MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM,  MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM,  MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM,  MCA5-1SB - lMM FLUTED BALL, 5.6CM,  MCA5-1SD - lMM DIAMOND BALL, 5.6CM,  MCA5-2SB - 2MM FLUTED BALL, 5.6CM,  MCA5-2SD - 2MM DIAMOND BALL, 5.6CM,  MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM,  MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM,  MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM,  MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM,  MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM,  MCA7-15SD 1.5MM DIAMOND BALL, 7.3CM,  MCA7-1SB - lMM FLUTED BALL, 7.3CM,  MCA7-1SD - 1 MM DIAMOND BALL, 7 .3CM,  MCA7-2SB - 2MM FLUTED BALL, 7.3CM and  MCA7-2SD - 2MM DIAMOND BALL, 7.3CM.

Other devices under the same FDA submission

14 devices cleared under K042783 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page