HeartMate II System Controller recallZ-1228-2014 · Thoratec Corporation · initiated Mar 4, 2014
Class I Terminated All recalls posted in March 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Thoratec Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
FDA root cause category: Device Design
What the company is doing
Thoratec sent an Urgent Medical Device Correction letter dated March 4, 2014 to all affected customers. The letter was sent via e-mail. All distributors outside of the USA and Canada to whom affected product has been distributed have been notified of this field correction notice and have been provided copies of the Customer Letter to forward to their customers. The letter identified the affected product , problem and actions to be taken. Customers were requested to complete and sign the attached Acknowledgement Form and return it to Thoratec via fax (925-847-8571) or e-mail a scanned copy to [email protected]. For questions call 1-800-528-2577.
Recall details
- Recall number
- Z-1228-2014
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Mar 4, 2014
- Posted by the FDA
- Mar 21, 2014
- Recalling firm
- Thoratec Corporation
- Quantity
- 4120
- Product code
- DSQ
- FDA submissions
- P060040
- Distribution
- Worldwide distribution: US (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI; and countries of: Austria, Belgium, Czech Republic, France, Germany, Ireland, Israel, Italy, Lebanon, Lithuania, Poland, Saudi Arabia, Singapore, Slovenia, Sweden, Switzerland, The Netherlands, Turkey, and United Kingdom.
Product description
HeartMate II System Controller Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump blood adequately on its own. It is sometimes called a a heart pump or VAD. HeartMate II is a small implantable LVAD. HeartMate II attaches to the heart and is designed to assist or take over the pumping function of the patients left ventricle, the main pumping chamber of the heart.
Code information (models, lots, serial numbers)
Catalog number 106762 (NAm) 106017 - EU all serial numbers
Other devices under the same FDA submission
7 devices cleared under P060040 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HeartMate™ 107754Thoratec Corporation · Ventricular (assist) bypass2 recalls447 reports
- HeartMate Touch™ HMT1100Thoratec Corporation · Ventricular (assist) bypass1 recall50 reports
- HeartMate™ 107760Thoratec Corporation · Ventricular (assist) bypass1 recall6 reports
- HeartMate Touch™ HMT1100-RThoratec Corporation · Ventricular (assist) bypass1 recall3 reports
- HeartMate™ 106128Thoratec Corporation · Ventricular (assist) bypass1 recall0 reports
- HeartMate Touch™ HMT2100Thoratec Corporation · Ventricular (assist) bypass10 reports
- HeartMate™ 1340Thoratec Corporation · Ventricular (assist) bypass1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 67639).
- Z-1227-2014 · TerminatedHeartMate II Implant Kit with Pocket Controller; with sealed grafts (NAm) with sealed grafts (EU) Product Usage: HeartMate II is a mechanical device that circulates blood through…
- Z-1230-2014 · TerminatedHeart Mate II Pocket Controllers removed from packaging Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak t…
- Z-229-2014 · TerminatedHeartMate II LVAD Pump and Pock Controller Kit; Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump b…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page