Portex Low Dead Space Connector with Sideport, 3.5mm recallZ-1641-2014 · Smiths Medical ASD, Inc. · initiated Apr 10, 2014
Class I Terminated All recalls posted in June 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
One lot of 3.0mm sized Connectors, Product Reorder No. 10035-05PS, Lot No. 2553426 were placed in packages labeled as size 3.5mm.
FDA root cause category: Packaging process control
What the company is doing
Smiths Medical sent an Urgent Medical Device Recall Notice dated April 10, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory and quarantine all affected products and to complete the attached Response Form and return it to Smith Medical within 7 days of receipt of the notice. Customers with questions were instructed to contact Smiths Medical's Customer Service Department at +1-800-258-5361.
Recall details
- Recall number
- Z-1641-2014
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Apr 10, 2014
- Posted by the FDA
- Jun 11, 2014
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 100 devices
- Product code
- BTR
- FDA submissions
- K790312
- Distribution
- US distribution in the state of North Carolina.
Product description
Portex Low Dead Space Connector with Sideport, 3.5mm Product Usage: The connector is used to attach the endotracheal tube to the patient breathing circuit. The low dead space connector minimizes mechanical dead space and subsequent carbon dioxide rebreathing. The connector's sideport permits access for administration of artificial surfactant, airway pressure monitoring, gas sampling, and instillation of sterile saline for airway leakage.
Code information (models, lots, serial numbers)
Lot or Serial No: 2553426
Other devices under the same FDA submission
28 devices cleared under K790312 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Portex 100/199/075Icu Medical, Inc. · Tube, tracheal (w/wo connector)4 deaths222 reports
- Portex 100/111/025Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall13 reports
- Portex 100/141/030Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall13 reports
- Portex 100/111/030Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall6 reports
- Portex 100/126/030Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall6 reports
- Portex 100/111/035Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall5 reports
- Portex 100/141/025Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall5 reports
- Portex 100/127/025Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall5 reports
- Portex 100/126/025Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall5 reports
- Portex 100/111/020Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall1 death1 report
- Portex 100/127/030Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall1 report
- Portex 100/105/025Icu Medical, Inc. · Tube, tracheal (w/wo connector)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page