Panda Freestanding Infant Warmers, integrated with Nellcor… recallZ-1549-2014 · GE Healthcare · initiated Apr 2, 2014
Class II Terminated All recalls posted in May 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
FDA root cause category: Software design
What the company is doing
GE Healthcare notified Biomedical/Clinical Engineering managers, Labor & Delivery/ NICU Nurse managers, Respiratory Therapy Managers and Directors of Risk Management by letter 04/02/2014. End users were advised that they may continue to use the system provided they follow the GE recommended actions: 1) If in use with a patient, turn off the Nellcor SpO2 function in the control panel and switch to an approved alternate form of SpO2 measurement. 2) Prior to using the affected GE warmer with the Nellcor SpO2 function, or resuming the use of this function with any patient, contact your hospital Bio-Med or qualified technician to check the Nellcor SpO2 alarm functionality using either an adult Nellcor SpO2 sensor or a Nellcor SpO2 simulator following the applicable instructions in the Appendix 3) If the test activates a Low Pulse Rate alarm, the user-adjusted Nellcor SpO2 alarm limits are functioning properly and the warmer can be placed back in clinical service. End users were also advised that alarm functionality must be re-checked following every service event requiring access to the internal electronics and during annual preventative maintenance, until revised software is installed. 4) If the test does NOT activate a Low Pulse Rate alarm, remove the warmer from clinical service and contact your local GE Healthcare Service Representative.
Recall details
- Recall number
- Z-1549-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 2, 2014
- Posted by the FDA
- May 6, 2014
- Recalling firm
- GE Healthcare
- Quantity
- 21 units
- Product code
- FMT
- FDA submissions
- K122267, K123309
- Distribution
- Worldwide Distribution.
Product description
Panda Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).
Code information (models, lots, serial numbers)
HDJS51142 HDJS52094 HDJS52095 HDJS52153 HDJS52154 HDJS52155 HDJS52156 HDJS52223 HDJS52224 HDJS52517 HDJS52518 HDJS52519 HDJS52520 HDJS52521 HDJS52522 HDJS52974 HDJS52975 HDJS52976 HDJS52977 HDJS52978 HDJS52979
Other devices under the same FDA submission
3 devices cleared under K122267, K123309 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Panda i-Res Warmer 1.0Datex-Ohmeda, Inc. · WARMER, INFANT RADIANT1 recall15 reports
- Panda iRes WarmerDatex-Ohmeda, Inc. · WARMER, INFANT RADIANT34 reports
- Giraffe WarmerDatex-Ohmeda, Inc. · WARMER, INFANT RADIANT27 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 68067).
- Z-1547-2014 · TerminatedGiraffe Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitorin…
- Z-1548-2014 · TerminatedPanda iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monito…
- Z-1550-2014 · TerminatedPanda Wall-Mount Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page