Medtronic MiniMed Paradigm Insulin Infusion Pumps recallZ-2544-2014 · Medtronic MiniMed Inc. · initiated Mar 13, 2014

Class II Terminated All recalls posted in August 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic MiniMed Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.

FDA root cause category: Device Design

What the company is doing

The firm, Medtronic, sent an "IMPORTANT MEDICAL DEVICE SAFETY INFORMATION" letter dated March 13, 2014 to its customers. The letter described the product, problem and actions to be taken. The letter informs the customers that because accidental button pressing errors may occur it is important that customers always confirm the insulin dose flashing on the display is correct before pressing ACT to start delivery. The letter informs the customers of what safety settings will minimize the risk of an accidental programming error and what actions to be taken. Should you have any questions about this letter or concerns with your keypad, please do not hesitate to contact the 24 hour helpline at (866) 222-7304 for a replacement.

Recall details

Recall number
Z-2544-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 13, 2014
Posted by the FDA
Aug 22, 2014
Recalling firm
Medtronic MiniMed Inc.
Quantity
559,374 units total (444,374 units in US)
Product code
OYC
FDA submissions
P980022, P120010
Distribution
Worldwide Distribution-Nationwide in US

Product description

Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.

Code information (models, lots, serial numbers)

Paradigm: MMT-523NAB, MMT-723KNB, MMT-523NAH, MMT-723KNH, MMT -523NAL, MMT-723KNL, MMT-523NAP, MMT-723KNP, MMT-523NAS, MMT-723KNS, MMT-723NAB, MMT-723NAH, MMT-723NAL, MMT-723NAP, MMT-723NAS, MMT-523KNB, MMT-523KNH, MMT-523KNL, MMT-523KNP, MMT-523KNS.  MiniMed 530G: MMT-551NAM, MMT-551NAH, MMT-551NAL, MMT-551NAP, MMT-551NAS, MMT-751NAB, MMT-751NAH, MMT-751NAL, MMT-751NAP, MMT-751NAS.

Other devices under the same FDA submission

353 devices cleared under P980022, P120010 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.