MiniMed® 530G MMT-751LNAS (UDI 00763000161170)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-751LNAS PRDGM V3.0 SK EN PR LN · Class 3
All models of MiniMed 530G (71)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic MiniMed 530G insulin pump has 2,218 reports, which include 2 deaths and 600 injuries. Reports describe mechanical problems, blocked flow, power problems and motor errors, and some mention high blood sugar. 1 recall is open, from February 2026, over insulin delivery that can be too high or too low when the pump is raised or lowered relative to the infusion site.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,218
- Deaths reported
- 2
- Injuries reported
- 600
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2,218 adverse event reports that list this exact device (DI 00763000161170), received between Oct 16, 2018 and Apr 25, 2023 — classified as 2 deaths, 600 injuries, 1,616 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (591 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P120010); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Feb 13, 2026Class IIOpen, Classified
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Z-1751-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem591
- Obstruction of Flow590
- Device Difficult to Program or Calibrate557
- Adverse Event Without Identified Device or Use Problem391
- Power Problem350
- No Apparent Adverse Event335
- Battery Problem285
- Patient Device Interaction Problem271
Patient problems
- No Consequences Or Impact To Patient1,522
- Hyperglycemia580
- Hypoglycemia124
- Elevated Ketones/ Diabetic Ketoacidosis43
- Nausea27
- Vomiting16
- Blood Loss10
- Abdominal Pain9
Recent adverse event reports
The 25 most recent of 2,218 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2023Battery Problem; Battery Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER HAS BEEN RECEIVING VARIOUS LOW BATTERY ALERTS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED AND CUSTOMER WAS ADVISED TO MONITOR THE PUMP AND CALL BACK IF ISSUE RECURS. CUSTOMER WAS INFORMED THAT NEW BATTERY CAP WILL BE SENT. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16804734 ↗MalfunctionAug 10, 2022Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM. CUSTOMER WAS ABLE TO COMPLETE PUMP REWIND AND DRIVE SUPPORT CAP WAS NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USE OF THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15196557 ↗MalfunctionMar 7, 2021Premature Discharge of Battery; Loss of Data; Premature Discharge of Battery; Loss of Data
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A MULTIPLE PUMP ERROR ALARM. CUSTOMER GOT LOW BATTERY ALARM. THE CUSTOMER STATED THEY WERE ABLE TO CLEAR THE ALARM AND WERE ABLE TO COMPLETE THE REWIND PROCESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 11427001 ↗MalfunctionFeb 22, 2021Communication or Transmission Problem; Communication or Transmission Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE SENSOR WAS GIVING INACCURATE READINGS. BLOOD GLUCOSE WAS 300 MG/DL BUT THE SENSOR INDICATED LOW READING. CUSTOMER INSERTED ANOTHER SENSOR AND GOT LOST SENSOR ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 11361623 ↗MalfunctionOct 27, 2020Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD NO DELIVERY ALARM. THE CUSTOMER WAS DECLINED TO CONTINUE TROUBLESHOOT FOR NO DELIVERY ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10738726 ↗MalfunctionOct 8, 2020Obstruction of Flow; Communication or Transmission Problem; Obstruction of Flow; Communication or Transmission Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD UNEXPLAINED NO DELIVERY ALARM AND TRANSMITTER LOSES CONNECTION. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10646856 ↗MalfunctionOct 7, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED MOTOR ERROR ALARM. THEY WERE ABLE TO CLEAR THE ALARM, REWIND THE INSULIN PUMP AND IT PASSED THE DISPLACEMENT TEST. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10646457 ↗MalfunctionSep 29, 2020Moisture or Humidity Problem; Patient Device Interaction Problem; Moisture or Humidity Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC STATED THAT THE RESERVOIR COMPARTMENT HAD CHIPPING ON THE THREADING. CUSTOMER STATED THAT INSULIN PUMP MIGHT SUBMERGED IN WATER. NO HARM WAS REQUIRED DURING MEDICAL INTERVENTION. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10600592 ↗MalfunctionSep 29, 2020Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD NO DELIVERY ALARM. CUSTOMER STATED THAT THE INSULIN PUMP WAS NOT DELIVERING INSULIN CORRECTLY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10598774 ↗MalfunctionSep 21, 2020No Display/Image; Material Integrity Problem; Power Problem; No Display/Image; Material Integrity Problem; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BLANK DISPLAY AND THE BATTERY CAP WAS BROKE OFF. CUSTOMER STATED INSULIN PUMP WAS NOT EXPOSED TO MOISTURE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10558718 ↗MalfunctionSep 18, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS DAMAGED. CUSTOMER STATED THAT THE INSULIN PUMP HAD COSMETIC DAMAGE. CUSTOMER STATED THAT THE RESERVOIR AND INSULIN PUMP DOES NOT SHOW SIGNS OF DAMAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10552122 ↗MalfunctionSep 18, 2020Failure to Interrogate; Failure to Interrogate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD INSULIN PUMP ERROR ALARM. CUSTOMER REPORTED THAT THEY WERE ABLE TO CLEAR AND REWIND THE INSULIN PUMP. SELF TEST WAS PASSED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10551893 ↗MalfunctionSep 14, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER REPORTS THE DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM. CUSTOMER WAS ABLE TO COMPLETE PUMP REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10526591 ↗MalfunctionSep 2, 2020Mechanical Problem; Environmental Compatibility Problem; Mechanical Problem; Environmental Compatibility Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS EXPOSED TO HIGH MAGNETIC FIELD AND HAD MULTIPLE PUMP ERROR ALARMS. THE CUSTOMER STATED THEY WERE ABLE TO CLEAR THE ALARMS AND ABLE TO COMPLETE REWIND PROCESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10481521 ↗MalfunctionAug 26, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED MOTOR ERROR ALARM. THEY ALSO REPORTED THAT THE RESERVOIR HAD BEEN ALMOST FULL OR AROUND THE HALF WAY MARK. THE CUSTOMER WAS ABLE TO CLEAR THE ALARM AND REWIND THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10452473 ↗MalfunctionAug 24, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CLIP WAS BROKE THAT'S WHY THE INSULIN PUMP FELL. THE CUSTOMER STATED THAT THE INSULIN PUMP HIT THE FLOOR AND THE SCREEN SHATTERED. THE SCREEN WAS CRACKED AND THEY CAN NOT READ IT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10441606 ↗MalfunctionAug 21, 2020Premature Discharge of Battery; Premature Discharge of Battery
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THEIR INSULIN PUMP HAD UNEXPECTED RESTART ALARM. CUSTOMER WAS ABLE TO CLEAR ALARM AND ABLE TO COMPLETE REWIND THE INSULIN PUMP. CUSTOMER REPORTED THAT THE ALARM DID NOT OCCUR AFTER INSERTING A NEW BATTERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10435711 ↗MalfunctionAug 12, 2020Mechanical Problem; Environmental Compatibility Problem; Mechanical Problem; Environmental Compatibility Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR POSITION ENCODED ERROR AND MOTOR ERROR ALARM. CUSTOMER STATED DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER STATED INSULIN PUMP WAS EXPOSED TO HIGH MAGNETIC ENVIRONMENT. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10395441 ↗MalfunctionAug 12, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BUTTONS OF INSULIN PUMP WERE HARDER TO PRESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10395190 ↗InjuryAug 11, 2020Mechanical Problem; Power Problem; Mechanical Problem; Power Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 425 MG/DL AT THE TIME OF INCIDENT AND OTHER BLOOD GLUCOSE VALUE WAS 133 MG/DL. CUSTOMER REPORTED THAT THE INSULIN PUMP HAD REJECTED NEW BATTERIES AND GRINDS WHEN REWIND. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND DECLINED FOR HIGH BLOOD GLUCOSE. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS. ¿
Patient problems: Hyperglycemia
MAUDE report 10390505 ↗MalfunctionAug 11, 2020Material Integrity Problem; Material Integrity Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT BATTERY CAP WAS BROKE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10389489 ↗MalfunctionAug 11, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS GIVING MOTOR ERROR ALERTS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10389414 ↗MalfunctionAug 11, 2020Device Difficult to Program or Calibrate; Wireless Communication Problem; Device Difficult to Program or Calibrate; Wireless Communication Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD PUMP ERROR ALARM ALSO HAD KEYPAD ANOMALY. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER WAS UNABLE TO CLEAR THE ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10389297 ↗MalfunctionAug 10, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Mechanical Problem; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE ACT BUTTON TAKES A COUPLE PUSHES TO GET IT TO RESPOND. THE CUSTOMER HAD TO HIT UP AND DOWN ARROWS JUST RIGHT TO GET THEM TO RESPOND. THE INSULIN PUMP HAD GOTTEN MOTOR ALARMS RECENTLY, AND INSULIN PUMP MAKES THEM TAKE VIAL OUT AND REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10384392 ↗InjuryAug 10, 2020Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE EXPERIENCING HIGH BLOOD GLUCOSE. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 420 MG/DL. THE INSULIN PUMP HAD A MOTOR ERROR ALARM. THE CUSTOMER WAS UNABLE TO CLEAR THE ALARM AND UNABLE TO COMPLETE THE PUMP REWIND. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10384144 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120010. One submission often covers many models, so check the product and lot details in each record.
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Automated insulin dosing , threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page