MiniMed® 530G MMT-751LNAS (UDI 00643169933224)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-751LNAS PRDGM V3.0 SK EN PR LN · Class 3

All models of MiniMed 530G (71)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This MiniMed 530G is an insulin pump in a threshold-suspend artificial pancreas system. Its 3,201 reports mostly describe programming and calibration difficulties, blocked flow, and mechanical or power problems. The reports include 7 deaths and 1,020 injuries. There is 1 open recall, started in February 2026, because insulin delivery can be too high or too low when the pump is raised or lowered relative to the infusion site.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
3,201
Deaths reported
7
Injuries reported
1,020
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3,201 adverse event reports that list this exact device (DI 00643169933224), received between Dec 14, 2017 and Apr 14, 2023 — classified as 7 deaths, 1,020 injuries, 2,174 malfunctions.

The most frequently reported specific device problem is “Device Difficult to Program or Calibrate” (786 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P120010); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 13, 2026Class IIOpen, Classified

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

    Z-1751-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Difficult to Program or Calibrate786
  • Obstruction of Flow696
  • Mechanical Problem649
  • Adverse Event Without Identified Device or Use Problem548
  • Power Problem525
  • Patient Device Interaction Problem501
  • No Apparent Adverse Event474
  • Battery Problem306

Patient problems

  • No Consequences Or Impact To Patient2,011
  • Hyperglycemia983
  • Hypoglycemia258
  • Elevated Ketones/ Diabetic Ketoacidosis68
  • Nausea18
  • Blood Loss14
  • No Known Impact Or Consequence To Patient9
  • Death8

Recent adverse event reports

The 25 most recent of 3,201 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2023Battery Problem; Power Problem; Battery Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTS A LOW BATTERY ALARM. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE ISSUE WAS NOT RESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE DEVICE AND THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16741435 ↗
  • MalfunctionNov 4, 2021Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A MOTOR ERROR ALARM. CUSTOMER STATED THEY WERE ABLE TO CLEAR THE ALARM AND ABLE TO REWIND THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 12750555 ↗
  • MalfunctionAug 16, 2021Obstruction of Flow; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A NO DELIVERY ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 12328633 ↗
  • MalfunctionFeb 4, 2021Power Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD FAILED BATTERY TEST ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11277201 ↗
  • MalfunctionOct 13, 2020Mechanical Problem; Battery Problem; Communication or Transmission Problem; Mechanical Problem; Battery Problem; Communication or Transmission Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BATTERY LIFE WAS NOW QUICKER AND REWINDING NOW WAS MUCH ROUGHER THAN BEFORE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER ROUTINELY GOT A CALIBRATION ERROR, AND AT TIMES UNABLE TO DOWNLOAD INTO CARE LINK. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10675730 ↗
  • MalfunctionAug 26, 2020Obstruction of Flow; Power Problem; Obstruction of Flow; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD RANDOM NO DELIVERY ALARM AND INSULIN PUMP REJECTS NEW BATTERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10452715 ↗
  • MalfunctionAug 13, 2020Failure to Cycle; Failure to Cycle

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM ERROR WHILE CHANGING THE RESERVOIR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10401073 ↗
  • MalfunctionAug 11, 2020Device Difficult to Program or Calibrate; Moisture or Humidity Problem; Device Difficult to Program or Calibrate; Moisture or Humidity Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON WONT READ WHEN PRESSED AND RAIN BOW EFFECT ON SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10389440 ↗
  • InjuryAug 11, 2020Priming Problem; Priming Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY HAD LOW BLOOD GLUCOSE LEVEL. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 40 MG/DL. CUSTOMER REPORTED THAT THE INSULIN WAS DRIPPING. THE INSULIN WAS SQUIRTING DURING MANUAL PRIME. THE TROUBLESHOOTING WAS PERFORMED FOR PRIMING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 10389310 ↗
  • MalfunctionAug 10, 2020Premature Discharge of Battery; Premature Discharge of Battery

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD AN PUMP ERROR ALARM. THE CUSTOMER STATED THEY WERE ABLE TO CLEAR THE ALARM AND WERE ABLE TO COMPLETE THE REWIND PROCESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10383760 ↗
  • MalfunctionAug 7, 2020Device Alarm System; Device Difficult to Program or Calibrate; Priming Problem; Device Alarm System; Device Difficult to Program or Calibrate; Priming Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP SQUIRTED INSULIN DURING PRIMING, THE DOWN ARROW BUTTON WAS STICKING AND ALARMS WERE HARDER TO HEAR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10378145 ↗
  • MalfunctionAug 7, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER REPORTED THEY WERE UNABLE TO CLEAR THE ALARM AND UNABLE TO REWIND THE INSULIN PUMP. CUSTOMER STATED DRIVE SUPPORT CAP WAS NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10377982 ↗
  • MalfunctionAug 6, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERRORS AND ALSO IT WAS NOT WORKING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10371948 ↗
  • MalfunctionAug 6, 2020Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM, NO DELIVERY ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10371890 ↗
  • MalfunctionAug 6, 2020Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM ALARM. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS LOOSE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10371824 ↗
  • MalfunctionAug 5, 2020Obstruction of Flow; Battery Problem; Priming Problem; Obstruction of Flow; Battery Problem; Priming Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN WAS SQUIRTS WHILE PRIMING. CUSTOMER REPORTE THAT THE BATTERY LAST LESS THAN SEVEN DAYS, RANDOM NO DELIVERY ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10371408 ↗
  • MalfunctionAug 5, 2020Mechanical Problem; Battery Problem; Mechanical Problem; Battery Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR POSITION ENCODER ERROR AND HAD MOTOR ERROR ALARM. THE CUSTOMER STATED THAT THE INSULIN PUMP BATTERY WAS LOW. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10366315 ↗
  • MalfunctionAug 5, 2020Mechanical Problem; Obstruction of Flow; Patient Device Interaction Problem; Mechanical Problem; Obstruction of Flow; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR ERROR ALARMS. THE CUSTOMER STATED THAT INSULIN PUMP HAD MINOR SCRATCHES ON SCREEN, ALSO IT ALARMED FOR RANDOM NO DELIVERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10366118 ↗
  • MalfunctionAug 4, 2020Mechanical Problem; Battery Problem; Power Problem; Mechanical Problem; Battery Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR, HAD REJECTED NEW BATTERIES AND THE BATTERY LASTED ONLY FOR A DAY SINCE A MONTH OR TWO. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10362627 ↗
  • MalfunctionAug 4, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED MOTOR ERROR ALARM. CUSTOMER REPORTED THAT THEY WERE ABLE TO CLEAR THE ALARM. THEY WERE ABLE TO REWIND THE INSULIN PUMP. CUSTOMER REPORTED THE HISTORY OF MOTOR ERROR ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10360576 ↗
  • MalfunctionAug 4, 2020Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD STICK BUTTON AND HARD TO PRESS. CUSTOMER REPORTED THAT THE INSULIN PUMP BUTTON UNRESPONSIVE AT TIME AND UNEXPLAINED NO INSULIN DELIVERY ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10360527 ↗
  • MalfunctionAug 3, 2020Failure to Cycle; Failure to Cycle

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM. THE CUSTOMER STATED THE INSULIN PUMP WAS RESETTING SO THEY WERE UNABLE TO USE THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10355585 ↗
  • MalfunctionAug 3, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP'S BUTTON HAD TO PRESS HARD FOR IT WORK. TRANSMITTER WAS GIVING UPDATE ERROR AND INACCURATE READING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10354958 ↗
  • MalfunctionAug 3, 2020Obstruction of Flow; Battery Problem; Patient Device Interaction Problem; Obstruction of Flow; Battery Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC THE RETAINER RING WAS BROKEN. THE BATTERY WAS DIMINISHED AND ALSO RECEIVED NO DELIVERY ALARMS. BATTERY CAP WAS BROKEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10354836 ↗
  • MalfunctionAug 1, 2020No Display/Image; Moisture or Humidity Problem; Patient Device Interaction Problem; No Display/Image; Moisture or Humidity Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BLANK DISPLAY. CUSTOMER STATED THAT THE INSULIN PUMP WAS EXPOSED TO MOISTURE. CUSTOMER STATED THEY SEE FLUID UNDER THE LCD. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10353993 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120010. One submission often covers many models, so check the product and lot details in each record.

  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Artificial pancreas device system, threshold suspend devices

All Artificial pancreas device system, threshold suspend devices →