G7 Acetabular System Positioning Guide Rod, orthopedic… recallZ-1837-2014 · Biomet, Inc. · initiated May 7, 2014
Class II Terminated All recalls posted in June 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Guide rod may fracture during impaction of the acetabular shell and be retained in the inserter or the patient and could result in a delay in surgery.
FDA root cause category: Use error
What the company is doing
The firm issued the following notice on May 7, 2014: URGENT MEDICAL DEVICE RECALL NOTICE: This notification is to inform you of an Urgent Medical Device Recall initiated by Biomet Orthopedics ("Biomet") which involves Part Numbers: 110003458 G7 Positioning Guide Rod and 110003500 G7 Positioning Guide Post. These instruments have been consigned and/or invoiced to your account. Biomet has initiated this action following an investigation which identified that positioning guide rod may fracture during impaction of the acetabula shell. If the positioning guide rod fractures, then the positioning guide post may become stuck to the insertion handle. There is no adverse health outcome expected for the patient. This action requires the immediate location and discontinued use of the product and its return to Biomet. The following actions are REQUIRED: Immediately locate and remove the identified device(s) listed below from circulation. Carefully follow the instructions on the enclosed "FAX Back Response Form. Fax a copy of the Response Form to 574-372-1683 prior to return of product. Use priority carrier for your shipment. If you have further distributed this product, you MUST notify hospital personnel of this action via the enclosed "Dear Risk/Recall Manager" notice. This letter MUST be given to hospital personnel responsible for receiving recall notices. However, you are charged with the location and return of these products.
Recall details
- Recall number
- Z-1837-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 7, 2014
- Posted by the FDA
- Jun 19, 2014
- Recalling firm
- Biomet, Inc.
- Quantity
- 1047 total
- Product code
- PBI
- FDA submissions
- K121874
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Canada, Netherlands, New Zealand, and Japan.
Product description
G7 Acetabular System Positioning Guide Rod, orthopedic surgical instrument for hip prosthesis PN 110003458.
Code information (models, lots, serial numbers)
PN:110003458. Lots 424960, ZB130702, ZB130703, ZB130704, ZB130705, ZB130701, ZB131201, 360877, 377304 and ZB140201.
Other devices under the same FDA submission
62 devices cleared under K121874 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- G7® Acetabular System 010000663Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall30 reports
- G7® Acetabular System 010000667Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 recall3 reports
- G7® Acetabular System 010000665Biomet Orthopedics, LLC · prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive31 reports
- G7® 110003450Biomet Orthopedics, LLC · prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive29 reports
- G7® Acetabular System 010000662Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented27 reports
- G7 STRAIGHT MONOBLOCK ACETABULAR 110003450Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED23 reports
- G7 ACETABULAR SCREW 010000997Biomet Orthopedics, LLC · Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented21 reports
- G7® Acetabular System 010000664Biomet Orthopedics, LLC · Hip, semi-constrained, cemented, metal/polymer + additive, cemented20 reports
- G7 ACETABULAR SCREW 010000998Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED20 reports
- G7 STRAIGHT INSERTER THREADED SHAFT 110003452Biomet Orthopedics, LLC · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer20 reports
- G7® 110003454Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented20 reports
- G7 ACETABULAR SCREW 010000999Biomet Orthopedics, LLC · Hip, semi-constrained, cemented, metal/polymer + additive, cemented19 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 68310).
- Z-1838-2014 · TerminatedG7 Acetabular System Positioning Guide Post, orthopedic surgical instrument for hip prosthesis. PN:110003500.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page