Skip to content

Pivot Bipolar Cup 22x42/43-51/53mm, REF 133-YYYY recallZ-2253-2014 · Ortho Development Corporation · initiated Jul 11, 2014

Class II Terminated All recalls posted in August 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho Development Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.

FDA root cause category: Packaging process control

What the company is doing

Ortho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.

Recall details

Recall number
Z-2253-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 11, 2014
Posted by the FDA
Aug 17, 2014
Recalling firm
Ortho Development Corporation
Quantity
188
Product code
KWY
FDA submissions
K050966
Distribution
Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.

Product description

Pivot Bipolar Cup 22x42/43-51/53mm, REF 133-YYYY Product Usage: Used in a hemiarthroplasty procedure as the bearing surface to articulate with both the CoCr femoral head and the natural acetabulum

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Pivot Bipolar Cup 22x42mm,Part 133-2242,Lot A136254,Pivot Bipolar Cup 22x43mm,Part 133-2243,Lot A136256,Pivot Bipolar Cup 22x43mm,Part 133-2243,Lot A136255,Pivot Bipolar Cup 22x44mm,Part 133-2244,Lot A136257,Pivot Bipolar Cup 22x45mm,Part 133-2245,Lot A136258,Pivot Bipolar Cup 22x46mm,Part 133-2246,Lot A136259,Pivot Bipolar Cup 22x46mm,Part 133-2246,Lot A136260,Pivot Bipolar Cup 22x47mm,Part 133-2247,Lot A136261,Pivot Bipolar Cup 22x48mm,Part 133-2248,Lot A136262,Pivot Bipolar Cup 22x49mm,Part 133-2249,Lot A136263,Pivot Bipolar Cup 22x50mm,Part 133-2250,Lot A136265,Pivot Bipolar Cup 22x51mm,Part 133-2251,Lot A136266,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136269,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136268,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136267,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136270,

Other devices under the same FDA submission

10 devices cleared under K050966 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 68898).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page