PS Tibial Insert, REF 163-1XXX recallZ-2255-2014 · Ortho Development Corporation · initiated Jul 11, 2014
Class II Terminated All recalls posted in August 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho Development Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
FDA root cause category: Packaging process control
What the company is doing
Ortho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.
Recall details
- Recall number
- Z-2255-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 11, 2014
- Posted by the FDA
- Aug 17, 2014
- Recalling firm
- Ortho Development Corporation
- Quantity
- 11
- Product code
- JWH
- FDA submissions
- K994370
- Distribution
- Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
Product description
PS Tibial Insert, REF 163-1XXX Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component
Code information (models, lots, serial numbers)
PS Tibial Insert Size 3 8mm,Part 163-1308,Lot A135943,PS Tibial Insert Size 3 9mm,Part 163-1309,Lot A136492,PS Tibial Insert Size 3 12mm,Part 163-1312,Lot A136453,PS Tibial Insert Size 4 7mm,Part 163-1407,Lot A135945,PS Tibial Insert Size 4 7mm,Part 163-1407,Lot A136454,PS Tibial Insert Size 5 9mm,Part 163-1509,Lot A135827,
Other devices under the same FDA submission
4 devices cleared under K994370 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Balanced Knee System 162-1400AOrtho Development Corporation · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER3 reports
- Balanced Knee System 161-1402Ortho Development Corporation · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Balanced Knee System 163-1509Ortho Development Corporation · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Balanced Knee System 162-1300AOrtho Development Corporation · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 68898).
- Z-2253-2014 · TerminatedPivot Bipolar Cup 22x42/43-51/53mm, REF 133-YYYY Product Usage: Used in a hemiarthroplasty procedure as the bearing surface to articulate with both the CoCr femoral head and the n…
- Z-2254-2014 · TerminatedPatella 38mm, REF164-0038 Product Usage: UHMWPE component used in a TKA procedure to resurface the patella and serve as a bearing surface for sliding contact with the femoral co…
- Z-2257-2014 · TerminatedTibila Insert CK, Balanced Knee System Revision, REF563-1YYY Product Usage: UHMWPE component used in TKA and revision TKA procedures to serve as a tibial bearing surface for artic…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page