Tibila Insert CK, Balanced Knee System Revision, REF563-1YYY recallZ-2257-2014 · Ortho Development Corporation · initiated Jul 11, 2014

Class II Terminated All recalls posted in August 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho Development Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.

FDA root cause category: Packaging process control

What the company is doing

Ortho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.

Recall details

Recall number
Z-2257-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 11, 2014
Posted by the FDA
Aug 17, 2014
Recalling firm
Ortho Development Corporation
Quantity
1
Product code
JWH
FDA submissions
K060569
Distribution
Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.

Product description

Tibila Insert CK, Balanced Knee System Revision, REF563-1YYY Product Usage: UHMWPE component used in TKA and revision TKA procedures to serve as a tibial bearing surface for articulation with the femoral component where additional joint constraint is needed

Code information (models, lots, serial numbers)

CK Tibial Insert Sz 6 12mm Part 563-1612, Lot A135666

Other devices under the same FDA submission

4 devices cleared under K060569 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 68898).

Sources and updates