Persona Cemented Tibial Broach Inserter/Extractor recallZ-0853-2015 · Zimmer, Inc. · initiated Dec 4, 2014
Class II Terminated All recalls posted in December 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi
FDA root cause category: Process design
What the company is doing
Notifications will be sent to the affected distributors and hospital risk managers with instructions for returning the affected product the week of December 8, 2014. Recalled product will be quarantined and returned to Zimmer Sales Representatives to be destroyed. The public contact for the recall is: Consumer Relations Call Center Zimmer, Inc. P O Box 708 Warsaw, IN 46581-0708 [email protected] Phone: 1-800-447-5633 Fax: 574-371-8755
Recall details
- Recall number
- Z-0853-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 4, 2014
- Posted by the FDA
- Dec 23, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 83 units
- Product code
- JWH
- FDA submissions
- K113369
- Distribution
- Distributed in the states of MN; NJ; NC; PA; AL; MI; OR; FL; OH; CO; IL and CA, and the countries of Australia, India, UAE, Germany, and Italy.
Product description
Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.
Code information (models, lots, serial numbers)
Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754
Other devices under the same FDA submission
249 devices cleared under K113369 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- Persona® The Personalized Knee® 42-5270-005-05Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER234 reports
- Persona® The Personalized Knee® 42-5170-005-05Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER185 reports
- Persona® The Personalized Knee® 42-5170-003-03Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER135 reports
- Persona® The Personalized Knee® 42-5270-007-07Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER129 reports
- Persona® The Personalized Knee® 42-5170-007-07Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER116 reports
- Persona® The Personalized Knee® 42-5270-003-03Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER112 reports
- PERSONA™ 42-5099-093-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER72 reports
- PERSONA® Natural Tibia® 42-5320-071-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER64 reports
- PERSONA™ Natural Tibia™ 42-5321-083-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- PERSONA™ Natural Tibia™ 42-5321-083-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- PERSONA® 42-5399-083-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page