Skip to content

Artis Q, and Artis Q.zen ceiling systems recallZ-0972-2015 · Siemens Medical Solutions USA, Inc · initiated Dec 3, 2014

Class II Terminated All recalls posted in January 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Suboptimal routing of the cable may result in increased wear over time. Without additional measures, damage of a cable may result in limited functionality and in rare cases, failure of a system. An ongoing procedure could be terminated in rare cases.

FDA root cause category: Device Design

What the company is doing

Siemens sent an important customer safety information notice, dated December 3, 2014, to all affected customers. The letter identified the product, problem, and action to be taken. The following hardware modifications will be implemented with the corrective action AX042/14/S: 1. A modified cable outlet to provide additional space for the cabling. 2. A modified plastic cover at the cable outlet providing optimal cable guidance in the area of the arc. In general all systems will be checked for existing damage prior to the implementation of this action. For questions regarding this recall call 610-219-6300.

Recall details

Recall number
Z-0972-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 3, 2014
Posted by the FDA
Jan 14, 2015
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
50
Product code
OWB
FDA submissions
K123529
Distribution
Nationwide Distribution

Product description

Artis Q, and Artis Q.zen ceiling systems; model numbers: 10848281; 10848354: The Artis family is dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.

Code information (models, lots, serial numbers)

model numbers: 10848281; 10848354

Other devices under the same FDA submission

7 devices cleared under K123529 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page