Skip to content

McKesson Cardiology ECG Management recallZ-0910-2015 · McKesson Israel Ltd. · initiated Dec 19, 2014

Class II Terminated All recalls posted in January 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from McKesson Israel Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An ECG procedure time stamp is incorrect, due to a software error in the McKesson Cardiology ECG Management with software versions 13.1 and 13.1.1.

FDA root cause category: Software design

What the company is doing

McKesson sent an Urgent Field Safety Notice dated December 16, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed how to identify the potentially affected procedures and update the procedure date (if relevant) until a software update is made advailable for installation by McKesson. McKesson will proactively identify prior procedures affected by this issue and will notify customers of any issues identified. In addition to the Field Safety Notice, McKesson Israel will provide correction to the affected SW versions and will contact customers for deployment of a software patch that prevents this scenario from occurring. Customers with questions should contact McKesson Support at 1-866-777-0202. For questions regarding this recall call 404-338-3556.

Recall details

Recall number
Z-0910-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 19, 2014
Posted by the FDA
Jan 5, 2015
Recalling firm
McKesson Israel Ltd.
Quantity
9
Product code
DQK
FDA submissions
K113515
Distribution
Worldwide Distribution - USA including MA, MS, NH, NC, TX, WA and Internationally to the United Kingdom.

Product description

McKesson Cardiology ECG Management It is a software application designed to import, display, store, analyze, distribute and manage information related to ECG procedures of adult and pediatric patients from external ECG devices.

Code information (models, lots, serial numbers)

MC13.1 and MC13.1.1

Other devices under the same FDA submission

1 device cleared under K113515 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page