Change Healthcare Cardiology ECG Management 14.1.1Recalls & reported problems
Change Healthcare Israel Ltd · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 17540262110029), received between Feb 20, 2020 and Feb 20, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Computer Software Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K113515); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Computer Software Problem1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2020Computer Software Problem; Computer Software Problem
THE ECG WAVEFORMS GRID SCALE, IN THE SPECIFIC "SEQUENTIAL' CONFIGURATION IS DYNAMICALLY CHANGED BASED ON THE SCREEN RESOLUTION IN USE. THEREFORE, THE ECG WAVEFORMS MAY BE DISPLAYED USING OTHER SCALES RATHER THAN 25MM/SEC. THE ON-SCREEN SCALE BAR CORRECTLY INDICATES THE ACTUAL SCALE USED FOR THE DISPLAY, AND THE HEART RATE VALUE AS SENT FROM THE CART/ CALCULATED BY THE SYSTEM IS CLEARLY DISPLAYED ON THE SCREEN AND IS NOT AFFECTED BY THE ISSUE. IN ADDITION, THE SEQUENTIAL LAYOUT USED IN THE ECG PDF REPORT IS DISPLAYED IN THE 25 MM/SEC SCALE. THE ISSUE WAS REPORTED DURING AN ECG MANAGEMENT IMPLEMENTATION BY THE CUSTOMER, PRIOR TO USING THE DEVICE IN CLINICAL USE. NO PATIENT HARM WAS REPORTED AS A RESULT OF THIS ISSUE.
Patient problems: No Patient Involvement
MAUDE report 9731293 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113515. One submission often covers many models, so check the product and lot details in each record.
- Z-1682-2020 · Feb 10, 2020 · Open, ClassifiedMcKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management
- Z-0910-2015 · Dec 19, 2014 · TerminatedMcKesson Cardiology ECG Management It is a software application designed to import, display, store, analyze, distribute and manage information related to ECG procedures of adult a
Other Computer, Diagnostic, Programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page