ADVIA Centaur Systems Progesterone Kit, (1 -pack) recallZ-1229-2015 · Siemens Healthcare Diagnostics, Inc · initiated Jan 27, 2015
Class II Terminated All recalls posted in March 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L)
FDA root cause category: Process control
What the company is doing
Siemens issued an Urgent Medical Device Correction (UMDC) was sent to affected Siemens Healthcare Diagnostics customers in the United States on January 27, 2015 to be delivered to customers on January 28, 2015. An Urgent Field Safety Notice (U FSN) was sent out for distribution to all customers outside the United States on January 27, 2015. These notices inform the customer of the potential to observe a high bias on patient sample with results greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. To mitigate the bias, Siemens instructs customers to dilute samples with values greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. In addition, Siemens informs customers to continue to follow Urgent Medical Device Correction 10819675, Rev. A and Urgent Field Safety notice 10819674, Rev A (Information Regarding the ADVIA Centaur Systems Calibrator E), dated September 2014. Fax-back forms to be returned to confirm that the customers received the Urgent Medical Device Correction.
Recall details
- Recall number
- Z-1229-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 27, 2015
- Posted by the FDA
- Mar 2, 2015
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 9778 kits
- Product code
- JLS
- FDA submissions
- K932955
- Distribution
- Worldwide Distribution-US (nationwide), Austria, Bahrain, Belgium, Bulgaria, Brazil, Canada, Chile, Colombia, Croatia, Cyprus, Czech, Republic, Denmark, Egypt, Estonia, France, Georgia, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Serbia, Slovakia, Spain, South America, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and Vatican.
Product description
ADVIA Centaur Systems Progesterone Kit, (1 -pack) In Vitro Diagnostic; Catalog number: 02382928; SMN: 10310305.
Code information (models, lots, serial numbers)
Lot Numbers: 42564268 42621268 43834268 Exp. Date: May 22, 2015
Other devices under the same FDA submission
4 devices cleared under K932955 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Atellica IM PRGE 10995659Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PROGESTERONE16 reports
- ADVIA Centaur® PRGE +E 10315522Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PROGESTERONE15 reports
- ADVIA Centaur® PRGE 10310305Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PROGESTERONE3 reports
- Atellica IM PRGE 10995660Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PROGESTERONE3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 70470).
- Z-1230-2015 · TerminatedADVIA Centaur Systems Progesterone Kit, (5 -pack); In Vitro Diagnostic; Catalog number: 01586287; SMN:10315522.
- Z-1231-2015 · TerminatedADVIA Centaur Systems Progesterone Kit, (5-pack Ref); In Vitro Diagnostic Catalog number: 02177364; SMN: 10333111.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page