Skip to content

ADVIA Centaur Systems Progesterone Kit, (5 -pack) recallZ-1230-2015 · Siemens Healthcare Diagnostics, Inc · initiated Jan 27, 2015

Class II Terminated All recalls posted in March 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L)

FDA root cause category: Process control

What the company is doing

Siemens issued an Urgent Medical Device Correction (UMDC) was sent to affected Siemens Healthcare Diagnostics customers in the United States on January 27, 2015 to be delivered to customers on January 28, 2015. An Urgent Field Safety Notice (U FSN) was sent out for distribution to all customers outside the United States on January 27, 2015. These notices inform the customer of the potential to observe a high bias on patient sample with results greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. To mitigate the bias, Siemens instructs customers to dilute samples with values greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. In addition, Siemens informs customers to continue to follow Urgent Medical Device Correction 10819675, Rev. A and Urgent Field Safety notice 10819674, Rev A (Information Regarding the ADVIA Centaur Systems Calibrator E), dated September 2014. Fax-back forms to be returned to confirm that the customers received the Urgent Medical Device Correction.

Recall details

Recall number
Z-1230-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 27, 2015
Posted by the FDA
Mar 2, 2015
Recalling firm
Siemens Healthcare Diagnostics, Inc
Quantity
11980 kits
Product code
JLS
FDA submissions
K932955
Distribution
Worldwide Distribution-US (nationwide), Austria, Bahrain, Belgium, Bulgaria, Brazil, Canada, Chile, Colombia, Croatia, Cyprus, Czech, Republic, Denmark, Egypt, Estonia, France, Georgia, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Serbia, Slovakia, Spain, South America, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and Vatican.

Product description

ADVIA Centaur Systems Progesterone Kit, (5 -pack); In Vitro Diagnostic; Catalog number: 01586287; SMN:10315522.

Code information (models, lots, serial numbers)

Lot Numbers: 42565268 42624268 43538268 Exp. Date: May 22, 2015

Other devices under the same FDA submission

4 devices cleared under K932955 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 70470).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page