Brand Name: Persona Anterior Referencing Sizer with Locking… recallZ-1360-2015 · Zimmer, Inc. · initiated Mar 10, 2015

Class II Terminated All recalls posted in April 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Zimmer, Inc is voluntarily recalling 22 lots of Persona Anterior Referencing Sizer with Locking Boom, part number 42-5099-088-10. Devices are being removed from distribution due to nonconforming to print specifications.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Zimmer sent an Urgent Medical device recall letter to all customers on March 10, 2015, via letter and email. The letter identified the product, the problem, and the action to be taken by the customer. Affected units will be inspected. Units that fail inspection will be physically destroyed/scrapped. Units that pass inspection will be re-etched and redistributed. If after reviewing this notification customers have further questions or concerns they were instructed to call the customer call center at 1-877-946-2761. Hours of operation are Monday through Friday, 8:00 a.m. through 5:00 p.m. EST. For questions regarding this recall call 574-372-4487.

Recall details

Recall number
Z-1360-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 10, 2015
Posted by the FDA
Apr 1, 2015
Recalling firm
Zimmer, Inc.
Quantity
688
Product code
JWH
FDA submissions
K113369
Distribution
Worldwide Distribution - US including Domestic: Puerto Rico, AZ, NY. OR, ME, NC, OH, TX, CO, IL, TN, PA, HI, CA, KS, WI, VA, MN, NJ, UT, MA, GA, MI, OK, MS, FL, AL, IN, and Internationally to Germany, India, Dominican Republic, Singapore, Korea, Japan, Austrailia, Canada.

Product description

Brand Name: Persona Anterior Referencing Sizer with Locking Boom The Persona Anterior Referencing Sizer is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. The Persona Surgical Technique (97-5026-001-00, Rev. 9) states that the sizer helps ensure that the desired external rotation is attained and provides a size for the femoral component.

Code information (models, lots, serial numbers)

Part # 42-5099-088-10; Lots: 56573342; 56574505; 56574999; 56574229; 56574818; 56575000; 56574230; 56574819; 56575020; 56574301; 56574820; 56575021; 56574349; 56574861; 56575072; 56574374; 56574862; 56575105; 56574375; 56574863; 56574504; 56574998;