4.0mm Variable Angle, Self-Drilling Hexalobe Screw, 14mm recallZ-1547-2015 · Alphatec Spine, Inc. · initiated Apr 10, 2015
Class II Terminated All recalls posted in April 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alphatec Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During a field audit it was discovered by an employee that a 16mm Trestle Luxe Screw has been both laser marked and labeled as a 14mm Trestle Luxe Screw (4.0mm Variable Angle, Self-Drilling Hexalobe Screw, 14mm). As a result the firm recalling the affected lots codes for this mislabeling.
FDA root cause category: Employee error
What the company is doing
Alphatec Spine sent an Urgent-Medical Device Recall Notification dated April 7, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm first requested that customers notify their consignees of the recall and the risk, and risk mitigation factors. The firm requests that customers take the following field actions: -Upon receipt of the Field Safety Notice to review inventory to determine if affected screw --part number 71240-14, lot 685904 is in their possession. -If the product is in their possession to please abstain from use and contact Alphatec Spine Customer Service ([email protected]) immediately for instructions on how to return the affected product. All shipping instructions will be provided, including replace product and/or credit. -Fill out page 3 of the letter to confirm that they have read the notification and have take all necessary actions stated in the notification. -Return a signed copy using one of the methods below: Mail to: Alphatec Spine, Inc. 5818 El Camino Real Carlsbad, CA 92008 Fax to: 1-800 431-7729 For further questions call ( 760) 431-9286.
Recall details
- Recall number
- Z-1547-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 10, 2015
- Posted by the FDA
- Apr 29, 2015
- Recalling firm
- Alphatec Spine, Inc.
- Quantity
- (244 units in US and 61 units OUS)
- Product code
- KWQ
- FDA submissions
- K102820
- Distribution
- Worldwide Distribution - US Distribution to the states of: KS, LA, TN, NJ, KY, CA, FL, OR, OK, AZ, WI, CO, IN, NH, AR, UT, and AL., and to the countries of : Spain, Japan, Belgium, and Saudi Arabia.
Product description
4.0mm Variable Angle, Self-Drilling Hexalobe Screw, 14mm Alphatec Spine Solutions for the Aging Spine Trestle Luxe Anterior Cervical Plating System REF 71240-14 LOT XXXXXX Alphatec Spine, Inc. EN: 4.0MM Variable Angle, Self--Drilling Hexalobe Screw, 14MM (TI-6AL-4V ELI) NON STERILE Single Use Only
Code information (models, lots, serial numbers)
Part Number: 71240-14 Lot: 685904
Other devices under the same FDA submission
23 devices cleared under K102820 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trestle Luxe 71002-034Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY6 reports
- Trestle Luxe 71004-064Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY3 reports
- Trestle Luxe 71540-16Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY3 reports
- Trestle Luxe 71001-020Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- Trestle Luxe 71002-026Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- Trestle Luxe 71002-032Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- Trestle Luxe 71002-037Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- Trestle Luxe 71240-14Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- TRESTLE LUXE 711-0234Alphatec Spine, Inc. · Appliance, fixation, spinal intervertebral body1 report
- TRESTLE LUXE 722-14014Alphatec Spine, Inc. · Appliance, fixation, spinal intervertebral body1 report
- Trestle Luxe 71001-022Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
- Trestle Luxe 71002-028Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.